Afterloader integrated EMT enables improved dwell position model definition and quality assurance in Venezia gynaecological brachytherapy applicators

<i>Objective.</i>In brachytherapy for gynecological cancers using intracavitary applicators, implant reconstruction is commonly performed using applicator libraries. These libraries contain applicator geometry models as well as dwell position (DP) models defined in respect to the applicator geometry. In this study, we investigate whether an afterloader integrated electromagnetic tracking (EMT) system can be utilized for DP model definition and quality assurance in such applicators.<i>Approach.</i>DPs in four sets of two configurations of the Elekta Venezia Advanced Gynaecological Applicator (22 mm ovoids/40 mm intrauterine (IU) and 26 mm ovoids/70 mm IU) were measured using an afterloader integrated EMT system. Measurements were evaluated for reproducibility and compared against manufacturer-specified (MS) DPs and a computed tomography (CT)-corrected DP model.<i>Main Results.</i>Excellent EMT measurement reproducibility was observed, with values of ⩽0.2 mm for both configurations. The overall reproducibility, including applicator geometry reproducibility, was ⩽0.4 mm for both configurations. Significant discrepancies from the manufacturer's DP model were observed, with a mean ± sd deviation of 1.13 ± 0.66 mm (22/40) and 1.37 ± 0.63 (26/70), particularly in the IU channel, where MS DPs were not experimentally defined. Discrepancies were reduced to 0.89 ± 0.41 mm (22/40) and 0.81 ± 0.33 mm (26/70) when the CT-corrected DP model was used as baseline, highlighting the need for experimentally defined DP models. The overall uncertainty of single measurements was below the clinically acceptable 2 mm limit.<i>Significance.</i>This study confirms that afterloader integrated EMT can accurately reconstruct source paths in gynecological brachytherapy applicators and supports its incorporation into clinical workflows for improved quality assurance and treatment precision. The importance of EMT for quality assurance was highlighted by measured deviations from manufacturer's DP model in a clinical relevant part of the IU channel.

Authors

Institutions

Publication Details

Journal
Physics in Medicine and Biology
Published
2025-06-02
DOI
https://doi.org/10.1088/1361-6560/addfa7
Citations
2
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
3.77
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Afterloader integrated EMT enables improved dwell position model definition and quality assurance in Venezia gynaecological brachytherapy applicators

Jérémy Godart, Remi A. Nout, Robin Straathof, Mischa S. Hoogeman et al.
2 citations
Physics in Medicine and Biology
Advanced Radiotherapy Techniques
3.77
article

Afterloader integrated EMT enables improved dwell position model definition and quality assurance in Venezia gynaecological brachytherapy applicators

Jérémy Godart, Remi A. Nout, Robin Straathof, Mischa S. Hoogeman, Ioannis Androulakis, Henrike Westerveld, Inger‐Karine K. Kolkman‐Deurloo, Myra van Laar
article en
2 citations

Abstract

<i>Objective.</i>In brachytherapy for gynecological cancers using intracavitary applicators, implant reconstruction is commonly performed using applicator libraries. These libraries contain applicator geometry models as well as dwell position (DP) models defined in respect to the applicator geometry. In this study, we investigate whether an afterloader integrated electromagnetic tracking (EMT) system can be utilized for DP model definition and quality assurance in such applicators.<i>Approach.</i>DPs in four sets of two configurations of the Elekta Venezia Advanced Gynaecological Applicator (22 mm ovoids/40 mm intrauterine (IU) and 26 mm ovoids/70 mm IU) were measured using an afterloader integrated EMT system. Measurements were evaluated for reproducibility and compared against manufacturer-specified (MS) DPs and a computed tomography (CT)-corrected DP model.<i>Main Results.</i>Excellent EMT measurement reproducibility was observed, with values of ⩽0.2 mm for both configurations. The overall reproducibility, including applicator geometry reproducibility, was ⩽0.4 mm for both configurations. Significant discrepancies from the manufacturer's DP model were observed, with a mean ± sd deviation of 1.13 ± 0.66 mm (22/40) and 1.37 ± 0.63 (26/70), particularly in the IU channel, where MS DPs were not experimentally defined. Discrepancies were reduced to 0.89 ± 0.41 mm (22/40) and 0.81 ± 0.33 mm (26/70) when the CT-corrected DP model was used as baseline, highlighting the need for experimentally defined DP models. The overall uncertainty of single measurements was below the clinically acceptable 2 mm limit.<i>Significance.</i>This study confirms that afterloader integrated EMT can accurately reconstruct source paths in gynecological brachytherapy applicators and supports its incorporation into clinical workflows for improved quality assurance and treatment precision. The importance of EMT for quality assurance was highlighted by measured deviations from manufacturer's DP model in a clinical relevant part of the IU channel.

Physics in Medicine and BiologyVol. 70(12)
Erasmus MC Cancer Institute (NL), Erasmus University Rotterdam (NL), Delft University of Technology (NL)
Openalex Percentile: Top 7%
Advanced Radiotherapy Techniques
3.77
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.