Accuracy of the HeartMate 3 Left Ventricular Assist Device Flow Estimation

Left ventricular assist devices (LVADs), such as the HeartMate 3 (HM3), are a treatment option for advanced heart failure (HF). Non-invasive monitoring of heart-pump interaction is crucial but relies on estimated parameters only. Despite the widespread clinical utilization of HM3 estimated flow rate (Q<sub>est</sub>) and pulsatility index (PI) their accuracy was not yet quantified. This study assessed the HM3 monitoring by describing underlying physical characteristics, estimation algorithms and deriving accuracy of estimates in clinical practice. Identification measurements were performed to quantify the physical characteristics and mechanisms of flow estimation. Virtual patients coupled to a hybrid hock circulatory loop were employed to evaluate estimator accuracies in realistic operating conditions. The correlation between Q<sub>est</sub> and torque-generating-current was strong (r² > 0.99, p < 0.001), confirming that estimation is based on speed, current, and viscosity. Also, the PI is directly derived from current. However, the underlying current/flow relationship is ambiguous (one current value corresponds to two flow rates). The HM3 flow estimator showed an RMSE of 1.63 L/min and a correlation coefficient of r = 0.86 (p < 0.001). The non-linear relationship obscures the systolic portion of flow as current declines during peak systole at higher flow rates. HM3 monitoring is unable to accurately represent the actual hemodynamics, particularly in the transition between full support and partial support, which is a region highly prevalent and crucial in clinical practice. It is therefore recommended to interpret values and trends of the HM3 monitoring system with caution.

Authors

Institutions

Publication Details

Journal
Scientific Reports
Published
2025-05-30
DOI
https://doi.org/10.1038/s41598-025-03743-9
Citations
1
Primary Topic
Mechanical Circulatory Support Devices
Type
article
Field-Weighted Citation Impact
0.67

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Accuracy of the HeartMate 3 Left Ventricular Assist Device Flow Estimation

Daniel Zimpfer, Theodor Abart, Philipp Aigner, Marcus Granegger et al.
1 citations
Scientific Reports
Mechanical Circulatory Support Devices
0.67
article

Accuracy of the HeartMate 3 Left Ventricular Assist Device Flow Estimation

Daniel Zimpfer, Theodor Abart, Philipp Aigner, Marcus Granegger, Michael Röhrich, Marko Grujic, Stefan Jakubek
article en
1 citations

Abstract

Left ventricular assist devices (LVADs), such as the HeartMate 3 (HM3), are a treatment option for advanced heart failure (HF). Non-invasive monitoring of heart-pump interaction is crucial but relies on estimated parameters only. Despite the widespread clinical utilization of HM3 estimated flow rate (Q<sub>est</sub>) and pulsatility index (PI) their accuracy was not yet quantified. This study assessed the HM3 monitoring by describing underlying physical characteristics, estimation algorithms and deriving accuracy of estimates in clinical practice. Identification measurements were performed to quantify the physical characteristics and mechanisms of flow estimation. Virtual patients coupled to a hybrid hock circulatory loop were employed to evaluate estimator accuracies in realistic operating conditions. The correlation between Q<sub>est</sub> and torque-generating-current was strong (r² > 0.99, p < 0.001), confirming that estimation is based on speed, current, and viscosity. Also, the PI is directly derived from current. However, the underlying current/flow relationship is ambiguous (one current value corresponds to two flow rates). The HM3 flow estimator showed an RMSE of 1.63 L/min and a correlation coefficient of r = 0.86 (p < 0.001). The non-linear relationship obscures the systolic portion of flow as current declines during peak systole at higher flow rates. HM3 monitoring is unable to accurately represent the actual hemodynamics, particularly in the transition between full support and partial support, which is a region highly prevalent and crucial in clinical practice. It is therefore recommended to interpret values and trends of the HM3 monitoring system with caution.

Scientific ReportsVol. 15(1)
TU Wien (AT), Christian Doppler Laboratory for Thermoelectricity (AT), Medical University of Vienna (AT)
Österreichische Nationalstiftung für Forschung, Technologie und Entwicklung, Universität Wien, Medizinische Universität Wien, Christian Doppler Forschungsgesellschaft
Openalex Percentile: Top 31%
Mechanical Circulatory Support Devices
0.67
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.