Design and Evaluation of a Portable Pinhole SPECT System for 177Lu Imaging: Monte Carlo Simulations and Experimental Study

<b>Background</b>/<b>Objectives:</b> Lutetium-177 is a widely used radioisotope in targeted radionuclide therapy, particularly for treating certain types of cancers relying on beta and low-energy gamma emissions, making it suitable for both therapeutic and post-therapy monitoring purposes. The purpose of this study was to evaluate the technical parameters for developing a prototype portable gamma camera dedicated to <sup>177</sup>Lu imaging applications. <b>Methods</b>: The well-validated GATE Monte Carlo toolkit was used to study the characteristics of the system and evaluate its performance in terms of spatial resolution, sensitivity, and image quality. For this purpose, a series of Monte Carlo simulations were executed, modeling a channel-edge aperture pinhole collimator incorporating a variety of computational phantoms. The final configuration of the prototype was standardized, incorporating the crystal size, collimator design, shielding, and the optimal FOV. After the development of the actual prototype camera, the system was also validated experimentally on the same setups as the simulations. <b>Results</b>: The final configuration of the prototype imaging system was standardized based on simulation results and then experimentally validated using physical phantoms under equivalent conditions. A minification of 1:5, spatial resolution of 1.0 cm, and sensitivity of 5.2 Cps/MBq at 10 cm distance source-to-collimator distance were assessed and confirmed. The experimental results agreed within 5% of simulated values. <b>Conclusions</b>: This study establishes the technical feasibility and foundational performance of a portable pinhole imaging system for potential clinical use in <sup>177</sup>Lu imaging workflows and thereby improving therapeutic effectiveness.

Authors

Institutions

Publication Details

Journal
Diagnostics
Published
2025-05-30
DOI
https://doi.org/10.3390/diagnostics15111387
Primary Topic
Medical Imaging Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Design and Evaluation of a Portable Pinhole SPECT System for 177Lu Imaging: Monte Carlo Simulations and Experimental Study

V. Paneta, George C. Kagadis, George Loudos, Paraskevi Katsakiori et al.
Diagnostics
Medical Imaging Techniques and Applications
article

Design and Evaluation of a Portable Pinhole SPECT System for 177Lu Imaging: Monte Carlo Simulations and Experimental Study

V. Paneta, George C. Kagadis, George Loudos, Paraskevi Katsakiori, Georgios Savvidis, Vasileios Eleftheriadis, Μαρία Γεωργίου, Panagiotis Papadimitroulas, Eleftherios Fysikopoulos, Efthimis Lamprou, Sofia Lagoumtzi
article en

Abstract

<b>Background</b>/<b>Objectives:</b> Lutetium-177 is a widely used radioisotope in targeted radionuclide therapy, particularly for treating certain types of cancers relying on beta and low-energy gamma emissions, making it suitable for both therapeutic and post-therapy monitoring purposes. The purpose of this study was to evaluate the technical parameters for developing a prototype portable gamma camera dedicated to <sup>177</sup>Lu imaging applications. <b>Methods</b>: The well-validated GATE Monte Carlo toolkit was used to study the characteristics of the system and evaluate its performance in terms of spatial resolution, sensitivity, and image quality. For this purpose, a series of Monte Carlo simulations were executed, modeling a channel-edge aperture pinhole collimator incorporating a variety of computational phantoms. The final configuration of the prototype was standardized, incorporating the crystal size, collimator design, shielding, and the optimal FOV. After the development of the actual prototype camera, the system was also validated experimentally on the same setups as the simulations. <b>Results</b>: The final configuration of the prototype imaging system was standardized based on simulation results and then experimentally validated using physical phantoms under equivalent conditions. A minification of 1:5, spatial resolution of 1.0 cm, and sensitivity of 5.2 Cps/MBq at 10 cm distance source-to-collimator distance were assessed and confirmed. The experimental results agreed within 5% of simulated values. <b>Conclusions</b>: This study establishes the technical feasibility and foundational performance of a portable pinhole imaging system for potential clinical use in <sup>177</sup>Lu imaging workflows and thereby improving therapeutic effectiveness.

DiagnosticsVol. 15(11)
University of Thessaly (GR), University of Patras (GR), Ubitech (Greece) (GR)
European Regional Development Fund
Affordable and clean energy
Openalex Percentile: Top 81%
Medical Imaging Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.