Abstract A Compton camera prototype was developed and characterized, utilizing a novel epitaxial-quenching-resistor Silicon Photomultiplier (EQR SiPM). The SiPM, with dimensions of 3.75 × 3.55 mm 2 and an effective area of 3.00 × 3.00 mm 2 , was configured into a 6 × 6 array. The array, with horizontal and vertical spacings of 0.58 mm and 0.82 mm respectively, formed a total size of 25.4 × 25.4 mm 2 . Both the scatterer and absorber were constructed from identical Lutetium-yttrium oxyorthosilicate (LYSO) crystal arrays, each comprising 24 × 24 crystals with dimensions of 0.98 × 0.98 × 10 mm 3 in 1.0 mm pitch. Coincidence events were acquired using an 8-channel data acquisition card (DAQ). Image reconstruction was performed using a filtered back-projection (FBP) algorithm with a Shepp-Logan filter, optimized for multi-graphics processing units (GPUs) processing. Calibration results demonstrated an energy resolution (Full Width at Half Maximum, FWHM) of 11.89% at 662 keV. Imaging experiments with a point source yielded an angular resolution of 8.95° (FWHM) at a scatterer-to-absorber distance of 50 mm. Additionally, successful imaging of a V-shaped tubing phantom confirmed the system's ability to accurately image distributed sources. The study also investigated how angular resolution and detection efficiency change with varying distances between the two detector layers. The results from Monte Carlo simulations demonstrated the system's ability to simultaneously image two distinct sources. Furthermore, the findings suggest that the current system's energy resolution is the main factor constraining the overall angular resolution. Future work will focus on optimizing the system design, including studying the effects of temperature and bias on SiPM gain, to enhance the camera's performance for various applications.

Authors

Institutions

Publication Details

Journal
Journal of Instrumentation
Published
2025-05-01
DOI
https://doi.org/10.1088/1748-0221/20/05/p05048
Primary Topic
Radiation Detection and Scintillator Technologies
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Performance characterization of a Compton camera prototype based on a novel epitaxial-quenching-resistor SiPM

Jianlang Hua, Jianyong Jiang, Haihao Wang, Ziquan Yuan et al.
Journal of Instrumentation
Radiation Detection and Scintillator Technologies
article

Performance characterization of a Compton camera prototype based on a novel epitaxial-quenching-resistor SiPM

Jianlang Hua, Jianyong Jiang, Haihao Wang, Ziquan Yuan, Ruoxi Liao, Chenxi Li, Yuexian Hou, Dejun Han, Wei Hu
article en

Abstract

Abstract A Compton camera prototype was developed and characterized, utilizing a novel epitaxial-quenching-resistor Silicon Photomultiplier (EQR SiPM). The SiPM, with dimensions of 3.75 × 3.55 mm 2 and an effective area of 3.00 × 3.00 mm 2 , was configured into a 6 × 6 array. The array, with horizontal and vertical spacings of 0.58 mm and 0.82 mm respectively, formed a total size of 25.4 × 25.4 mm 2 . Both the scatterer and absorber were constructed from identical Lutetium-yttrium oxyorthosilicate (LYSO) crystal arrays, each comprising 24 × 24 crystals with dimensions of 0.98 × 0.98 × 10 mm 3 in 1.0 mm pitch. Coincidence events were acquired using an 8-channel data acquisition card (DAQ). Image reconstruction was performed using a filtered back-projection (FBP) algorithm with a Shepp-Logan filter, optimized for multi-graphics processing units (GPUs) processing. Calibration results demonstrated an energy resolution (Full Width at Half Maximum, FWHM) of 11.89% at 662 keV. Imaging experiments with a point source yielded an angular resolution of 8.95° (FWHM) at a scatterer-to-absorber distance of 50 mm. Additionally, successful imaging of a V-shaped tubing phantom confirmed the system's ability to accurately image distributed sources. The study also investigated how angular resolution and detection efficiency change with varying distances between the two detector layers. The results from Monte Carlo simulations demonstrated the system's ability to simultaneously image two distinct sources. Furthermore, the findings suggest that the current system's energy resolution is the main factor constraining the overall angular resolution. Future work will focus on optimizing the system design, including studying the effects of temperature and bias on SiPM gain, to enhance the camera's performance for various applications.

Journal of InstrumentationVol. 20(05)
Beijing Normal University (CN), WuXi AppTec (China) (CN)
Openalex Percentile: Top 82%
Radiation Detection and Scintillator Technologies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.