Ion-exchange high-performance liquid chromatography (HPLC) is commonly used to measure hemoglobin A1c (HbA1c) by distinguishing it from other hemoglobin (Hb) fractions based on net charge. Hb variants can interfere with this analysis, leading to spurious HbA1c results, particularly in HPLC-based methods. This study investigated blood samples showing an HV3 peak - which indicates an Hb variant - on the Tosoh G11 chromatogram during routine HbA1c analysis in a Danish laboratory. Over 30 workdays, 53/33,006 samples displayed an HV3 peak. Sanger sequencing identified Hb Riccarton as the most common variant associated with these peaks (n = 27), consistent with its prevalence in the Danish Blood Donor Study, suggesting it is common in the Danish Caucasian population. Hb Riccarton posed a particular analytical challenge, as over half of the cases showed fluctuating HV3 peaks, initially separated from the HbA1c fraction but subsequently integrated into it upon re-analysis. In regions where Hb Riccarton is prevalent, clinical laboratories using the Tosoh G11 must be aware of this phenomenon to avoid bias and inconsistency in HbA1c reporting. Following Hb Riccarton, HV3 peaks most often indicated HbE heterozygosity (n = 16). Compared to Tosoh G11, HbA1c from the DCA Vantage showed a mean bias of -9.7% for HbE, versus -2.1% in samples without Hb variants. For Hb Riccarton, the bias was -7.2% when the HV3-peak was integrated, and +8.6% when separated from the HbA1c fraction. However, if clinicians are aware of these variants being present, both methods may be used for diabetes monitoring if applied consistently.

Authors

Institutions

Publication Details

Journal
Hemoglobin
Published
2025-05-04
DOI
https://doi.org/10.1080/03630269.2025.2509004
Primary Topic
Diabetes Management and Research
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

High Prevalence of Hb Riccarton Challenges HbA1c Analysis in a Danish Clinical Laboratory Using the Tosoh G11

Thore Hillig, Ida Stangerup, Nanna Brøns, Ole Birger Pedersen et al.
Hemoglobin
Diabetes Management and Research
article

High Prevalence of Hb Riccarton Challenges HbA1c Analysis in a Danish Clinical Laboratory Using the Tosoh G11

Thore Hillig, Ida Stangerup, Nanna Brøns, Ole Birger Pedersen, Sisse Rye Ostrowski, Jesper Petersen, Christian Erikstrup, Majbritt Lund Witte, Andreas Glenthøj
article en

Abstract

Ion-exchange high-performance liquid chromatography (HPLC) is commonly used to measure hemoglobin A1c (HbA1c) by distinguishing it from other hemoglobin (Hb) fractions based on net charge. Hb variants can interfere with this analysis, leading to spurious HbA1c results, particularly in HPLC-based methods. This study investigated blood samples showing an HV3 peak - which indicates an Hb variant - on the Tosoh G11 chromatogram during routine HbA1c analysis in a Danish laboratory. Over 30 workdays, 53/33,006 samples displayed an HV3 peak. Sanger sequencing identified Hb Riccarton as the most common variant associated with these peaks (n = 27), consistent with its prevalence in the Danish Blood Donor Study, suggesting it is common in the Danish Caucasian population. Hb Riccarton posed a particular analytical challenge, as over half of the cases showed fluctuating HV3 peaks, initially separated from the HbA1c fraction but subsequently integrated into it upon re-analysis. In regions where Hb Riccarton is prevalent, clinical laboratories using the Tosoh G11 must be aware of this phenomenon to avoid bias and inconsistency in HbA1c reporting. Following Hb Riccarton, HV3 peaks most often indicated HbE heterozygosity (n = 16). Compared to Tosoh G11, HbA1c from the DCA Vantage showed a mean bias of -9.7% for HbE, versus -2.1% in samples without Hb variants. For Hb Riccarton, the bias was -7.2% when the HV3-peak was integrated, and +8.6% when separated from the HbA1c fraction. However, if clinicians are aware of these variants being present, both methods may be used for diabetes monitoring if applied consistently.

HemoglobinVol. 49(3)
University of Copenhagen (DK), Aarhus University (DK), Frederiksberg Hospital (DK), Aarhus University Hospital (DK), Copenhagen University Hospital (DK), Rigshospitalet (DK), Zealand University Hospital (DK)
Gentofte Hospital
Openalex Percentile: Top 83%
Diabetes Management and Research
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.