Influence of Thermal Treatment on Fracture Due to Cyclic Fatigue of Rotary Files

The objective of this study aimed to assess the cyclic fatigue fracture of ZenFlexTM, ProTaper GoldTM, and HyFlexTM. EDM rotary files in simulated curved canals, along with evaluating their physicochemical properties and thermal behavior. A total of 69 instruments from three different NiTi rotary file systems were utilized: ProTaper GoldTM (08/25), HyflexTM EDM (08/25), and ZenflexTM (06/25). The analysis consisted of four phases: Phase 1 involved SEM observation to detect manufacturing defects; Phase 2 focused on cyclic fatigue-induced fracture in a curved canal with a 60° radius of curvature of 2 mm and a diameter of 1.5 mm; Phase 3 encompassed SEM observation of the fractured fragment and obtaining a fractographic study; Phase 4 included physicochemical characterization, such as equiatomic relationships by Inductively Coupled Plasma Optical Emission Spectroscopy (ICP-OES), phase transformation by DSC, crystalline structure by X-ray diffraction, and morphological analysis by SEM. Microcracks and defects were observed on cutting edges of ZenFlexTM and ProTaper GoldTM files. The mean cycles to fracture were 2814.50±161.58, 2649.94±120.93, and 1362.89±88.33 for HyFlexTM, ZenFlexTM, and ProTaper GoldTM, respectively. Moreover, different phase transition temperatures were noted, with ZenFlexTM in the austenite phase, ProTaper GoldTM in martensite, and HyFlexTM in the R phase at room temperature, as corroborated by X-ray diffraction. Additionally, a quasi-equimolar relationship was observed for the different systems, with ZenFlexTM exhibiting a reduced helical angle, followed by HyFlexTM and ProTaper GoldTM. ZenFlexTM files demonstrated greater resistance to cyclic fatigue, which appeared to be attributed to their physicochemical properties, heat treatment, and design.

Authors

Institutions

Publication Details

Journal
Odovtos - International Journal of Dental Sciences
Published
2025-05-29
DOI
https://doi.org/10.15517/ijds.2025.65254
Primary Topic
Mechanical Failure Analysis and Simulation
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Influence of Thermal Treatment on Fracture Due to Cyclic Fatigue of Rotary Files

Vladimir Alonso Escobar‐Barrios, Amaury Pozos‐Guillén, Verónica Méndez-González, César Gaitán‐Fonseca et al.
1 links
Odovtos - International Journal of Dental Sciences
Mechanical Failure Analysis and Simulation
article

Influence of Thermal Treatment on Fracture Due to Cyclic Fatigue of Rotary Files

Vladimir Alonso Escobar‐Barrios, Amaury Pozos‐Guillén, Verónica Méndez-González, César Gaitán‐Fonseca, Mariana Gutiérrez-Sánchez, María Concepción Betanzos-Juárez, Alejandra Izquierdo-Orozco
article en

Abstract

The objective of this study aimed to assess the cyclic fatigue fracture of ZenFlexTM, ProTaper GoldTM, and HyFlexTM. EDM rotary files in simulated curved canals, along with evaluating their physicochemical properties and thermal behavior. A total of 69 instruments from three different NiTi rotary file systems were utilized: ProTaper GoldTM (08/25), HyflexTM EDM (08/25), and ZenflexTM (06/25). The analysis consisted of four phases: Phase 1 involved SEM observation to detect manufacturing defects; Phase 2 focused on cyclic fatigue-induced fracture in a curved canal with a 60° radius of curvature of 2 mm and a diameter of 1.5 mm; Phase 3 encompassed SEM observation of the fractured fragment and obtaining a fractographic study; Phase 4 included physicochemical characterization, such as equiatomic relationships by Inductively Coupled Plasma Optical Emission Spectroscopy (ICP-OES), phase transformation by DSC, crystalline structure by X-ray diffraction, and morphological analysis by SEM. Microcracks and defects were observed on cutting edges of ZenFlexTM and ProTaper GoldTM files. The mean cycles to fracture were 2814.50±161.58, 2649.94±120.93, and 1362.89±88.33 for HyFlexTM, ZenFlexTM, and ProTaper GoldTM, respectively. Moreover, different phase transition temperatures were noted, with ZenFlexTM in the austenite phase, ProTaper GoldTM in martensite, and HyFlexTM in the R phase at room temperature, as corroborated by X-ray diffraction. Additionally, a quasi-equimolar relationship was observed for the different systems, with ZenFlexTM exhibiting a reduced helical angle, followed by HyFlexTM and ProTaper GoldTM. ZenFlexTM files demonstrated greater resistance to cyclic fatigue, which appeared to be attributed to their physicochemical properties, heat treatment, and design.

Odovtos - International Journal of Dental SciencesVol. 27(3)
Autonomous University of San Luis Potosí (MX), Polytechnic University of San Luis Potosí (MX), Institute for Scientific and Technological Research (MX), Universidad Autónoma de Zacatecas "Francisco García Salinas" (MX)
Universidad Autónoma de San Luis Potosí
Openalex Percentile: Top 89%
Mechanical Failure Analysis and Simulation
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.