Entrapment of the <scp>PulseSelect™</scp> catheter in the pulmonary vein: A rare complication due to soft tissue obstruction
We encountered a case in which a guidewire and PulseSelect™ catheter became entrapped within a pulmonary vein during atrial fibrillation ablation. Although careful manual traction allowed successful retrieval without complications, the maneuver carried a risk of serious adverse events. Operators should be aware of this rare but important complication and avoid excessive catheter advancement into a distal pulmonary vein to minimize the risk of entrapment.
Authors
- Makoto Mizutani (ORCID: https://orcid.org/0000-0001-8762-1106)
- Nobuaki Tanaka (ORCID: https://orcid.org/0000-0002-8324-1728)
- Kohei Iwasa (ORCID: https://orcid.org/0000-0001-7766-6896)
- Yoshihiko Ikeda (ORCID: https://orcid.org/0000-0001-6395-5072)
- Masato Okada (ORCID: https://orcid.org/0000-0002-3190-7313)
Institutions
- National Cerebral and Cardiovascular Center (JP)
- Sakurabashi Watanabe Hospital (JP)
Publication Details
- Journal
- Journal of Arrhythmia
- Published
- 2025-05-28
- DOI
- https://doi.org/10.1002/joa3.70098
- Citations
- 1
- Primary Topic
- Cardiac Arrhythmias and Treatments
- Type
- article
- Field-Weighted Citation Impact
- 1.39