A Bioresorbable Backscatter Sensor
We present the design, simulation, fabrication and measurement results of a biodegradable sensor for post operative monitoring. A backscattering measurement technique is demonstrated to detect frequency shifts resulting from the change in the thickness of a sensing film. Our sensor is composed of a modified split-ring resonator loaded with interdigitated capacitors. The sensor operates at around 3.2 GHz in free space and around 2 GHz in liquid solution. We demonstrated that our backscattering measurement data, quantified as resonance frequency in a laboratory environment, matches well with the simulation results.
Authors
- Luke A. Beardslee (ORCID: https://orcid.org/0000-0002-5713-5059)
- Sree Adinarayana Dasari (ORCID: https://orcid.org/0000-0002-2418-3003)
- Yuheng He (ORCID: https://orcid.org/0009-0008-9337-5034)
- Nima Ghalichechian (ORCID: https://orcid.org/0000-0002-2070-9864)
- Chinaza Ogbonna (ORCID: https://orcid.org/0000-0002-9623-3882)
Institutions
- Georgia Institute of Technology (US)
Publication Details
- Published
- 2025-04-22
- DOI
- https://doi.org/10.1109/rfid64926.2025.11015510
- Citations
- 1
- Primary Topic
- Water Quality Monitoring Technologies
- Type
- article
- Field-Weighted Citation Impact
- 0.82