Bilateral comparison of 1.018 V and 10 V standards between the SMD (Belgium) and the BIPM, November to December 2024 (part of the ongoing BIPM key comparison BIPM.EM-K11.a and b)

Main text As part of the ongoing BIPM key comparison BIPM.EM-K11.a and b, a comparison of the 1.018 V and 10 V voltage reference standards of the BIPM and the Service de Métrologie (SMD), Brussel, Belgium, was carried out from November to December 2024. Two BIPM Zener diode-based travelling standards (Fluke 732A), BIPM_1 (Z1) and BIPM_H (ZH), were transported to SMD and back to BIPM by road. Since the laboratories are close to each other, Fluke 732A standards that have a long metrological history were selected to serve as transfer standards. At the SMD and the BIPM, the reference standard for DC voltage is a Programmable Josephson Voltage Standard (PJVS). The output electromotive force (EMF) of each travelling standard was measured by direct comparison with the primary standard. The final result of the comparison is presented as the difference between the values assigned to DC voltage standards by SMD, at the level of 1.018 V and 10 V, at SMD, USMD, and those assigned by the BIPM, at the BIPM, UBIPM, at the reference date of the 6th of December 2024. USMD - UBIPM = -0.047 μV; u c = 0.017 μV, at 1.018 V USMD - UBIPM = 0.06 μV; u c = 0.09 μV, at 10 V where u c is the combined standard uncertainty associated with the measured difference, including the uncertainty of the realization of the volt at the BIPM and at SMD, and the uncertainty related to the comparison. To reach the main text of this paper, click on Final Report . Note that this text is that which appears in Appendix B of the BIPM key comparison database https://www.bipm.org/kcdb/ . The final report has been peer-reviewed and approved for publication by the CCEM, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA).

Authors

Institutions

Publication Details

Journal
Metrologia
Published
2025-01-01
DOI
https://doi.org/10.1088/0026-1394/62/1a/01001
Primary Topic
Radioactive Decay and Measurement Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Bilateral comparison of 1.018 V and 10 V standards between the SMD (Belgium) and the BIPM, November to December 2024 (part of the ongoing BIPM key comparison BIPM.EM-K11.a and b)

R Chayramy, S Solve, Michael J. Stock, Q Baire
Metrologia
Radioactive Decay and Measurement Techniques
article

Bilateral comparison of 1.018 V and 10 V standards between the SMD (Belgium) and the BIPM, November to December 2024 (part of the ongoing BIPM key comparison BIPM.EM-K11.a and b)

R Chayramy, S Solve, Michael J. Stock, Q Baire
article en

Abstract

Main text As part of the ongoing BIPM key comparison BIPM.EM-K11.a and b, a comparison of the 1.018 V and 10 V voltage reference standards of the BIPM and the Service de Métrologie (SMD), Brussel, Belgium, was carried out from November to December 2024. Two BIPM Zener diode-based travelling standards (Fluke 732A), BIPM_1 (Z1) and BIPM_H (ZH), were transported to SMD and back to BIPM by road. Since the laboratories are close to each other, Fluke 732A standards that have a long metrological history were selected to serve as transfer standards. At the SMD and the BIPM, the reference standard for DC voltage is a Programmable Josephson Voltage Standard (PJVS). The output electromotive force (EMF) of each travelling standard was measured by direct comparison with the primary standard. The final result of the comparison is presented as the difference between the values assigned to DC voltage standards by SMD, at the level of 1.018 V and 10 V, at SMD, USMD, and those assigned by the BIPM, at the BIPM, UBIPM, at the reference date of the 6th of December 2024. USMD - UBIPM = -0.047 μV; u c = 0.017 μV, at 1.018 V USMD - UBIPM = 0.06 μV; u c = 0.09 μV, at 10 V where u c is the combined standard uncertainty associated with the measured difference, including the uncertainty of the realization of the volt at the BIPM and at SMD, and the uncertainty related to the comparison. To reach the main text of this paper, click on Final Report . Note that this text is that which appears in Appendix B of the BIPM key comparison database https://www.bipm.org/kcdb/ . The final report has been peer-reviewed and approved for publication by the CCEM, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA).

MetrologiaVol. 62(1A)
Service Public Fédéral Économie (BE), Bureau international des poids et mesures (FR)
Openalex Percentile: Top 83%
Radioactive Decay and Measurement Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Bilateral comparison of 1.018 V and 10 V standards between the SMD (Belgium) and the BIPM, November to December 2024 (part of the ongoing BIPM key comparison BIPM.EM-K11.a and b) — R Chayramy, S Solve, et al. · Metrologia (2025) | TGRS Research Map | TGRS