Validity of a Single Inertial Measurement Unit to Measure Hip Range of Motion During Gait in Patients Undergoing Total Hip Arthroplasty

Hip flexion range of motion (ROM) during gait is an important surgery outcome for patients undergoing total hip arthroplasty (THA) that could help patient monitoring and rehabilitation. To allow systematic measurements during patients' clinical pathways, hip ROM measurement should be as simple and cheap as possible to ensure patient and clinician acceptance. Single IMU options can match these requirements and offer measurements both during daily living conditions and standardized clinical tests (e.g., 10 m walk, timed up-and-go). However, single-IMU approaches to measure hip ROM have been limited. Thus, the objective of this study was to explore the accuracy of one IMU in measuring hip ROM during gait and to determine whether a single-IMU approach can provide results comparable to those of multi-IMU systems. To assess this, machine learning models were employed, ranging from the simplest (linear regression) to more complex approaches (artificial neural networks). Eighteen patients undergoing THA and seven controls were measured using a 3D opto-electronic motion capture system and one thigh-mounted IMU. Hip ROM was predicted from thigh ROM using regression and classification models and was compared to the reference hip ROM. Multiple regression was the best-performing model, with limits of agreement (LoA) of ±13° and a systematic bias of 0. Random forest, RNN, GRU and LSTM models yielded LoA ranges > 27.8°, exceeding the threshold of acceptable error. These results showed that one IMU can measure hip ROM with errors comparable to those of two-IMU methods, with potential for improvement. Using multiple linear regression was sufficient and more appropriate than employing complex ANN models. This approach offers simplicity and acceptance to users in clinical settings.

Authors

Institutions

Publication Details

Journal
Sensors
Published
2025-05-27
DOI
https://doi.org/10.3390/s25113363
Citations
2
Primary Topic
Total Knee Arthroplasty Outcomes
Type
article
Field-Weighted Citation Impact
4.01
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Validity of a Single Inertial Measurement Unit to Measure Hip Range of Motion During Gait in Patients Undergoing Total Hip Arthroplasty

Kévin Rose-Dulcina, Stéphane Armand, Xavier Gasparutto, Didier Hannouche et al.
2 citations
Sensors
Total Knee Arthroplasty Outcomes
4.01
article

Validity of a Single Inertial Measurement Unit to Measure Hip Range of Motion During Gait in Patients Undergoing Total Hip Arthroplasty

Kévin Rose-Dulcina, Stéphane Armand, Xavier Gasparutto, Didier Hannouche, Peter Luca DiGiovanni, Noor Alalem
article en
2 citations

Abstract

Hip flexion range of motion (ROM) during gait is an important surgery outcome for patients undergoing total hip arthroplasty (THA) that could help patient monitoring and rehabilitation. To allow systematic measurements during patients' clinical pathways, hip ROM measurement should be as simple and cheap as possible to ensure patient and clinician acceptance. Single IMU options can match these requirements and offer measurements both during daily living conditions and standardized clinical tests (e.g., 10 m walk, timed up-and-go). However, single-IMU approaches to measure hip ROM have been limited. Thus, the objective of this study was to explore the accuracy of one IMU in measuring hip ROM during gait and to determine whether a single-IMU approach can provide results comparable to those of multi-IMU systems. To assess this, machine learning models were employed, ranging from the simplest (linear regression) to more complex approaches (artificial neural networks). Eighteen patients undergoing THA and seven controls were measured using a 3D opto-electronic motion capture system and one thigh-mounted IMU. Hip ROM was predicted from thigh ROM using regression and classification models and was compared to the reference hip ROM. Multiple regression was the best-performing model, with limits of agreement (LoA) of ±13° and a systematic bias of 0. Random forest, RNN, GRU and LSTM models yielded LoA ranges > 27.8°, exceeding the threshold of acceptable error. These results showed that one IMU can measure hip ROM with errors comparable to those of two-IMU methods, with potential for improvement. Using multiple linear regression was sufficient and more appropriate than employing complex ANN models. This approach offers simplicity and acceptance to users in clinical settings.

SensorsVol. 25(11)
University of Geneva (CH), University Hospital of Geneva (CH)
Openalex Percentile: Top 7%
Total Knee Arthroplasty Outcomes
4.01
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.