Effect of applied pressure on bioimpedance measurements of ex-vivo human lung tissue

Abstract The contrast in the electrical properties of healthy and malignant lung tissue could provide a methodology for intraoperative surgical margin assessment to reduce positive margins and improve clinical outcomes. However, the tissue is often compressed and deformed during intraoperative procedures. This study explores the impact applied pressure has on the electrical properties of ex vivo human lung tissue. Impedance spectra spanning 100 Hz to 10 MHz were recorded from freshly resected lung tissue specimens including lesion (n=3), perilesion (n=3), and healthy (n=4) samples, under an applied load and converted into resistivity. A linear regression was performed for each resistivity-pressure curve for all tissue types. A significant difference (p<0.05) was found between regression coefficients of lesion and perilesion tissues and both regression coefficient and constant for lesion and healthy tissue. No significant differences were found between the perilesion and healthy tissue coefficient or constant. These finding suggests that resistivity increases with pressure at different rates for malignant and healthy tissue that may be sufficient for surgical margin assessment.

Authors

Institutions

Publication Details

Journal
Journal of Physics Conference Series
Published
2025-05-01
DOI
https://doi.org/10.1088/1742-6596/3014/1/012010
Primary Topic
Electrical and Bioimpedance Tomography
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Effect of applied pressure on bioimpedance measurements of ex-vivo human lung tissue

David J. Finley, Joseph D. Phillips, Liang Lü, Ethan K. Murphy et al.
2 links
Journal of Physics Conference Series
Electrical and Bioimpedance Tomography
article

Effect of applied pressure on bioimpedance measurements of ex-vivo human lung tissue

David J. Finley, Joseph D. Phillips, Liang Lü, Ethan K. Murphy, Ryan J. Halter, Dana Ferrari-Light, Kayla A. Fay, Zenia Valdiviezo, Safina Suratwala
article en

Abstract

Abstract The contrast in the electrical properties of healthy and malignant lung tissue could provide a methodology for intraoperative surgical margin assessment to reduce positive margins and improve clinical outcomes. However, the tissue is often compressed and deformed during intraoperative procedures. This study explores the impact applied pressure has on the electrical properties of ex vivo human lung tissue. Impedance spectra spanning 100 Hz to 10 MHz were recorded from freshly resected lung tissue specimens including lesion (n=3), perilesion (n=3), and healthy (n=4) samples, under an applied load and converted into resistivity. A linear regression was performed for each resistivity-pressure curve for all tissue types. A significant difference (p<0.05) was found between regression coefficients of lesion and perilesion tissues and both regression coefficient and constant for lesion and healthy tissue. No significant differences were found between the perilesion and healthy tissue coefficient or constant. These finding suggests that resistivity increases with pressure at different rates for malignant and healthy tissue that may be sufficient for surgical margin assessment.

Journal of Physics Conference SeriesVol. 3014(1)
Dartmouth College (US), Dartmouth–Hitchcock Medical Center (US)
Openalex Percentile: Top 90%
Electrical and Bioimpedance Tomography
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.