Detecting Apnoea in Premature Neonates: Exploring Frequency-Domain and Power Analysis of Electrical Impedance Tomography Lung Impedance Signals

Abstract Neonatal apnoea is often defined as a cessation of breathing, a critical challenge in premature infant care, demanding early and accurate detection. This study integrates Electrical Impedance Tomography (EIT) with frequency-domain analysis to identify apnoea-related changes in lung impedance signals. We processed lung impedance waveforms from fifteen premature infants (median gestational age: 31 weeks) and assessed frequency, amplitude, and power as potential biomarkers. Statistical analysis using two-way ANOVA (p < 0.05) demonstrated significant changes in impedance features before, during, and after apnoea episodes. We further evaluated each feature’s classification performance via Receiver Operating Characteristic (ROC) curves, reporting Area Under the ROC Curve (AUC), sensitivity, specificity, Positive Predictive Value (PPV), and Negative Predictive Value (NPV). Frequency alone achieved a high AUC (0.98), with 85.96% sensitivity and 96.67% specificity, outperforming amplitude (AUC = 0.81) and power (AUC = 0.92). Notably, amplitude and power combined (AUC = 0.98) performed similarly to frequency on AUC but better on other measures (98.33% specificity). Similarly, combining power and frequency improved sensitivity, 89.47% at 97.50% specificity, surpassing frequency alone on both metrics. These results demonstrate the value of utilising a variety of attributes. While amplitude changes may be expected during apnoea, adding frequency and power metrics offers a more robust, multidimensional view of lung impedance behaviour. Moreover, residual signals during apnoea likely stem from cardiac activity and may also reflect slight changes in heart rate. This underscores the novelty of isolating lung-specific impedance from overlapping physiological signals. In conclusion, combining EIT and frequency-domain analysis offers a promising non-invasive method for real-time apnoea detection. In neonatal critical care units, it emphasises using multi-feature techniques to improve diagnostic accuracy.

Authors

Institutions

Publication Details

Journal
Journal of Physics Conference Series
Published
2025-05-01
DOI
https://doi.org/10.1088/1742-6596/3014/1/012016
Primary Topic
Wireless Power Transfer Systems
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Detecting Apnoea in Premature Neonates: Exploring Frequency-Domain and Power Analysis of Electrical Impedance Tomography Lung Impedance Signals

Andreas Demosthenous, Rebecca Yerworth, Richard Bayford, Xiaotao Wen et al.
Journal of Physics Conference Series
Wireless Power Transfer Systems
article

Detecting Apnoea in Premature Neonates: Exploring Frequency-Domain and Power Analysis of Electrical Impedance Tomography Lung Impedance Signals

Andreas Demosthenous, Rebecca Yerworth, Richard Bayford, Xiaotao Wen, J Omisanya, M Meidema
article en

Abstract

Abstract Neonatal apnoea is often defined as a cessation of breathing, a critical challenge in premature infant care, demanding early and accurate detection. This study integrates Electrical Impedance Tomography (EIT) with frequency-domain analysis to identify apnoea-related changes in lung impedance signals. We processed lung impedance waveforms from fifteen premature infants (median gestational age: 31 weeks) and assessed frequency, amplitude, and power as potential biomarkers. Statistical analysis using two-way ANOVA (p < 0.05) demonstrated significant changes in impedance features before, during, and after apnoea episodes. We further evaluated each feature’s classification performance via Receiver Operating Characteristic (ROC) curves, reporting Area Under the ROC Curve (AUC), sensitivity, specificity, Positive Predictive Value (PPV), and Negative Predictive Value (NPV). Frequency alone achieved a high AUC (0.98), with 85.96% sensitivity and 96.67% specificity, outperforming amplitude (AUC = 0.81) and power (AUC = 0.92). Notably, amplitude and power combined (AUC = 0.98) performed similarly to frequency on AUC but better on other measures (98.33% specificity). Similarly, combining power and frequency improved sensitivity, 89.47% at 97.50% specificity, surpassing frequency alone on both metrics. These results demonstrate the value of utilising a variety of attributes. While amplitude changes may be expected during apnoea, adding frequency and power metrics offers a more robust, multidimensional view of lung impedance behaviour. Moreover, residual signals during apnoea likely stem from cardiac activity and may also reflect slight changes in heart rate. This underscores the novelty of isolating lung-specific impedance from overlapping physiological signals. In conclusion, combining EIT and frequency-domain analysis offers a promising non-invasive method for real-time apnoea detection. In neonatal critical care units, it emphasises using multi-feature techniques to improve diagnostic accuracy.

Journal of Physics Conference SeriesVol. 3014(1)
Amsterdam UMC Location University of Amsterdam (NL), University Hospital Schleswig-Holstein (DE), Amsterdam University Medical Centers (NL), University College London (GB), Middlesex University London (GB), University of Amsterdam (NL), University of Lübeck (DE)
Openalex Percentile: Top 90%
Wireless Power Transfer Systems
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.