Method Comparison and Reference Intervals of β-Hydroxybutyric Acid Measurements by Using a Veterinary Point-of-Care Ketone Meter and a Reference Laboratory Analyzer in Central Bearded Dragons (Pogona vitticeps)

Abstract Hepatic lipid accumulation is highly prevalent in managed central bearded dragons ( Pogona vitticeps ), and standard blood biomarkers have proven ineffective for its diagnostic management. Recently, the ketone β-hydroxybutyrate (BHBA) has demonstrated potential in association with the severity of hepatic lipid accumulation in this species. Because BHBA measurement is not readily available to veterinary practices, the purpose of this study was to perform a method comparison assessment between a veterinary handheld point-of-care (POC) ketone meter (NovaVet Ketone/Glucose Meter, Nova Biomedical, Waltham, MA, USA) and a reference laboratory wet-chemistry analyzer (Ortho-Vitros 5600 Analyzer, QuidelOrtho, San Diego, CA, USA). Forty-four adult bearded dragons of both sexes were sampled, and BHBA concentrations were obtained using both analyzers. The veterinary POC ketone meter was found to underestimate BHBA values, with a significant proportional bias. The total observed error was higher than the combined inherent imprecision of the analyzers, but lower than the total allowable error for clinical agreement. A simple corrective formula (× 0.9) was found to improve agreement with the reference analyzer. No interference was observed with packed cell volume, lipemia, hemolysis, and blood glucose. Reference intervals for BHBA on the two analyzers were also provided on 30 healthy bearded dragons free of moderate-to-severe hepatic lipid accumulation. In conclusion, the veterinary POC ketone meter was found to not be analytically equivalent to the reference analyzer, but to have acceptable agreement for clinical use. This supports the use of this POC ketone meter for measuring BHBA in the diagnostic management of hepatic lipid accumulation and assessment of fasting status and peripheral lipolysis in central bearded dragons.

Authors

Institutions

Publication Details

Journal
Journal of Herpetological Medicine and Surgery
Published
2025-01-01
DOI
https://doi.org/10.5818/jhms-d-24-00026
Primary Topic
Metabolomics and Mass Spectrometry Studies
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Method Comparison and Reference Intervals of β-Hydroxybutyric Acid Measurements by Using a Veterinary Point-of-Care Ketone Meter and a Reference Laboratory Analyzer in Central Bearded Dragons (Pogona vitticeps)

Lisa Pacumio, Hugues Beaufrère, Danielle K. Tarbert, Mélanie Ammersbach
Journal of Herpetological Medicine and Surgery
Metabolomics and Mass Spectrometry Studies
article

Method Comparison and Reference Intervals of β-Hydroxybutyric Acid Measurements by Using a Veterinary Point-of-Care Ketone Meter and a Reference Laboratory Analyzer in Central Bearded Dragons (Pogona vitticeps)

Lisa Pacumio, Hugues Beaufrère, Danielle K. Tarbert, Mélanie Ammersbach
article en

Abstract

Abstract Hepatic lipid accumulation is highly prevalent in managed central bearded dragons ( Pogona vitticeps ), and standard blood biomarkers have proven ineffective for its diagnostic management. Recently, the ketone β-hydroxybutyrate (BHBA) has demonstrated potential in association with the severity of hepatic lipid accumulation in this species. Because BHBA measurement is not readily available to veterinary practices, the purpose of this study was to perform a method comparison assessment between a veterinary handheld point-of-care (POC) ketone meter (NovaVet Ketone/Glucose Meter, Nova Biomedical, Waltham, MA, USA) and a reference laboratory wet-chemistry analyzer (Ortho-Vitros 5600 Analyzer, QuidelOrtho, San Diego, CA, USA). Forty-four adult bearded dragons of both sexes were sampled, and BHBA concentrations were obtained using both analyzers. The veterinary POC ketone meter was found to underestimate BHBA values, with a significant proportional bias. The total observed error was higher than the combined inherent imprecision of the analyzers, but lower than the total allowable error for clinical agreement. A simple corrective formula (× 0.9) was found to improve agreement with the reference analyzer. No interference was observed with packed cell volume, lipemia, hemolysis, and blood glucose. Reference intervals for BHBA on the two analyzers were also provided on 30 healthy bearded dragons free of moderate-to-severe hepatic lipid accumulation. In conclusion, the veterinary POC ketone meter was found to not be analytically equivalent to the reference analyzer, but to have acceptable agreement for clinical use. This supports the use of this POC ketone meter for measuring BHBA in the diagnostic management of hepatic lipid accumulation and assessment of fasting status and peripheral lipolysis in central bearded dragons.

Journal of Herpetological Medicine and SurgeryVol. 35(3)
University of Tennessee at Knoxville (US), University of California, Davis (US)
Openalex Percentile: Top 89%
Metabolomics and Mass Spectrometry Studies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.