Evaluation of hemolysis index thresholds for 18 biochemistry assays: implications for laboratory-developed tests in the era of the IVDR

Abstract Objectives Hemolysis in laboratory samples is a frequent error that may lead to clinical misinterpretation and incorrect patient treatment. Thus, modern laboratory analyzers are able to assess the hemolysis index (HI). To prevent reporting of erroneous results, manufacturers are obliged to provide reliable HI threshold values. The aim of this study was to proof the quality of manufacturer’s given HI thresholds. Methods Dilution series with defined degrees of hemolysis were prepared by using blood samples from voluntary participants (n=77). In addition to the HI in each dilution, 18 measurands, namely sodium, potassium, chloride, aspartate aminotransferase (ASAT), alanine aminotransferase (ALAT), gamma-glutamyl transferase (GGT), lactate dehydrogenase (LDH), haptoglobin, total bilirubin, direct bilirubin, glutamate dehydrogenase (GLDH), creatine kinase (CK), iron, amylase, phosphate, total protein, enzymatic creatinine and high-sensitive troponin T, were measured on a cobas ® pro analyzer. Thresholds for maximum deviation of measurement results of 5 , 10 and 20 % were calculated. Then we determined cut-offs as the 5th percentile and median. Results Using potassium as an example we found HI thresholds at 57 (5th percentile) and 70 (median) with a corresponding result deviation of 5 % while the manufacturer HI threshold is given at 20. Thus, a higher HI threshold might be tolerated. Conclusions We established HI thresholds for 18 biochemistry assays. Eight assays showed considerable deviations of which six may have potential clinical relevance like potassium and high-sensitive troponin T. Optimizing thresholds can help to reduce the risk of unnecessary blocking of results and preventing considerably impaired results from being released.

Authors

Institutions

Publication Details

Journal
Journal of Laboratory Medicine
Published
2025-05-26
DOI
https://doi.org/10.1515/labmed-2024-0206
Citations
1
Primary Topic
Clinical Laboratory Practices and Quality Control
Type
article
Field-Weighted Citation Impact
1.16
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Evaluation of hemolysis index thresholds for 18 biochemistry assays: implications for laboratory-developed tests in the era of the IVDR

Michael J. Muehlbauer, Gunnar Brandhorst, Karen Friederike Gauß, Daniel Rosenkranz et al.
1 citations
Journal of Laboratory Medicine
Clinical Laboratory Practices and Quality Control
1.16
article

Evaluation of hemolysis index thresholds for 18 biochemistry assays: implications for laboratory-developed tests in the era of the IVDR

Michael J. Muehlbauer, Gunnar Brandhorst, Karen Friederike Gauß, Daniel Rosenkranz, Astrid Petersmann
article en
1 citations

Abstract

Abstract Objectives Hemolysis in laboratory samples is a frequent error that may lead to clinical misinterpretation and incorrect patient treatment. Thus, modern laboratory analyzers are able to assess the hemolysis index (HI). To prevent reporting of erroneous results, manufacturers are obliged to provide reliable HI threshold values. The aim of this study was to proof the quality of manufacturer’s given HI thresholds. Methods Dilution series with defined degrees of hemolysis were prepared by using blood samples from voluntary participants (n=77). In addition to the HI in each dilution, 18 measurands, namely sodium, potassium, chloride, aspartate aminotransferase (ASAT), alanine aminotransferase (ALAT), gamma-glutamyl transferase (GGT), lactate dehydrogenase (LDH), haptoglobin, total bilirubin, direct bilirubin, glutamate dehydrogenase (GLDH), creatine kinase (CK), iron, amylase, phosphate, total protein, enzymatic creatinine and high-sensitive troponin T, were measured on a cobas ® pro analyzer. Thresholds for maximum deviation of measurement results of 5 , 10 and 20 % were calculated. Then we determined cut-offs as the 5th percentile and median. Results Using potassium as an example we found HI thresholds at 57 (5th percentile) and 70 (median) with a corresponding result deviation of 5 % while the manufacturer HI threshold is given at 20. Thus, a higher HI threshold might be tolerated. Conclusions We established HI thresholds for 18 biochemistry assays. Eight assays showed considerable deviations of which six may have potential clinical relevance like potassium and high-sensitive troponin T. Optimizing thresholds can help to reduce the risk of unnecessary blocking of results and preventing considerably impaired results from being released.

Journal of Laboratory MedicineVol. 49(4)
Carl von Ossietzky Universität Oldenburg (DE), Universitätsmedizin Greifswald (DE)
Openalex Percentile: Top 22%
Clinical Laboratory Practices and Quality Control
1.16
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.