Clinical Performance and Optical Quality of a Non-Diffractive Extended-Depth-of-Focus Intraocular Lens in Patients Undergoing Cataract Surgery or Refractive Lensectomy

<b>Objectives:</b> To evaluate the clinical performance and optical quality of a non-diffractive extended-depth-of-focus (EDOF) intraocular lens (IOL), Asqelio™ EDOF (models ETLIO130C/ETPIO130C), in patients undergoing cataract surgery or refractive lensectomy. <b>Methods:</b> This prospective observational, case-control study included patients bilaterally implanted with either the Asqelio™ EDOF IOL (Study Group) or the spherical monofocal TECNIS<sup>®</sup> 1-Piece ZCB00 IOL (Control Group). The postoperative outcomes-at 3 months after surgery-included visual acuities at multiple distances, refraction, contrast sensitivity, the optical scatter index (OSI), wavefront aberrations, and patient-reported outcomes (Catquest-9SF and a quality-of-vision questionnaire). <b>Results:</b> Twenty-three patients (46 eyes) in the Asqelio™ EDOF group and 17 patients (34 eyes) in the monofocal control group were enrolled. Postoperatively, 91% of eyes in the EDOF group were within ±0.50 D of the intended spherical equivalent. The binocular uncorrected distance, intermediate, and near visual acuities were 0.00 ± 0.09, 0.13 ± 0.12, and 0.32 ± 0.15 logMAR, respectively. Contrast sensitivity and OSI values were similar between the study and control groups (<i>p</i> > 0.05). Higher-order aberrations were significantly lower in the EDOF group (<i>p</i> < 0.001), but values in both groups were clinically low. No adverse events were reported. Most patients expressed high satisfaction and reported few visual disturbances. <b>Conclusions:</b> The Asqelio™ EDOF IOL provided good refractive predictability, effective uncorrected vision across distance and intermediate ranges, and high patient satisfaction. Contrast sensitivity and optical scatter were comparable to monofocal implants. This lens can be considered a valuable option for patients seeking an extended range of functional vision with minimal side effects.

Authors

Institutions

Publication Details

Journal
Journal of Clinical Medicine
Published
2025-05-26
DOI
https://doi.org/10.3390/jcm14113717
Citations
1
Primary Topic
Ophthalmology and Visual Impairment Studies
Type
article
Field-Weighted Citation Impact
1.35
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Clinical Performance and Optical Quality of a Non-Diffractive Extended-Depth-of-Focus Intraocular Lens in Patients Undergoing Cataract Surgery or Refractive Lensectomy

Álvaro Sánchez-Ventosa, Alberto Villarrubia, Juan J. Prados-Carmona, Antonio Cano-Ortiz et al.
1 citations
Journal of Clinical Medicine
Ophthalmology and Visual Impairment Studies
1.35
article

Clinical Performance and Optical Quality of a Non-Diffractive Extended-Depth-of-Focus Intraocular Lens in Patients Undergoing Cataract Surgery or Refractive Lensectomy

Álvaro Sánchez-Ventosa, Alberto Villarrubia, Juan J. Prados-Carmona, Antonio Cano-Ortiz, Marta Villalba-González, Timoteo González-Cruces, Vanesa Díaz-Mesa, Francisco Javier Aguilar-Salazar
article en
1 citations

Abstract

<b>Objectives:</b> To evaluate the clinical performance and optical quality of a non-diffractive extended-depth-of-focus (EDOF) intraocular lens (IOL), Asqelio™ EDOF (models ETLIO130C/ETPIO130C), in patients undergoing cataract surgery or refractive lensectomy. <b>Methods:</b> This prospective observational, case-control study included patients bilaterally implanted with either the Asqelio™ EDOF IOL (Study Group) or the spherical monofocal TECNIS<sup>®</sup> 1-Piece ZCB00 IOL (Control Group). The postoperative outcomes-at 3 months after surgery-included visual acuities at multiple distances, refraction, contrast sensitivity, the optical scatter index (OSI), wavefront aberrations, and patient-reported outcomes (Catquest-9SF and a quality-of-vision questionnaire). <b>Results:</b> Twenty-three patients (46 eyes) in the Asqelio™ EDOF group and 17 patients (34 eyes) in the monofocal control group were enrolled. Postoperatively, 91% of eyes in the EDOF group were within ±0.50 D of the intended spherical equivalent. The binocular uncorrected distance, intermediate, and near visual acuities were 0.00 ± 0.09, 0.13 ± 0.12, and 0.32 ± 0.15 logMAR, respectively. Contrast sensitivity and OSI values were similar between the study and control groups (<i>p</i> > 0.05). Higher-order aberrations were significantly lower in the EDOF group (<i>p</i> < 0.001), but values in both groups were clinically low. No adverse events were reported. Most patients expressed high satisfaction and reported few visual disturbances. <b>Conclusions:</b> The Asqelio™ EDOF IOL provided good refractive predictability, effective uncorrected vision across distance and intermediate ranges, and high patient satisfaction. Contrast sensitivity and optical scatter were comparable to monofocal implants. This lens can be considered a valuable option for patients seeking an extended range of functional vision with minimal side effects.

Journal of Clinical MedicineVol. 14(11)
Universidad Loyola Andalucía (ES), Hospital Universitario Reina Sofía (ES)
Openalex Percentile: Top 19%
Ophthalmology and Visual Impairment Studies
1.35
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.