Agreement among glass mercury thermometers, infrared thermometers and digital thermometer temperature recordings in febrile under-five children

Background: This study aimed to assess the degree of agreement between the temperature readings of a glass mercury thermometer (GMT), a digital thermometer (DT) and an infrared thermometer (IR) in paediatric patients. This was a comparative study conducted at a tertiary institution in Nnewi, Anambra State, Southeast Nigeria. One hundred febrile children <5 years of age were recruited via convenience sampling. Methods: Temperatures were measured via an axillary GMT, an axillary DT and a noncontact forehead IR thermometer. Data were compared among the three methods via a 2×2 contingency table, receiver operating curve (ROC) and Bland Altman plot with GMT as the standard. Results: There was a greater significant correlation between GMT and DT (r=0.901, p=0.001) than between GMT and IR (r=0.695, p=0.001). Bland Altman plots revealed that IR and DT could be used interchangeably with GMT with an arithmetic mean of 1.1°C, confidence interval of 0.9359 to 1.1581, limits of agreement of -0.1–2.1 (37.4°C–39.6°C), p<0.001, outliers 3% at 1.96 SD and arithmetic mean of -0.1°C, confidence interval of -0.1337-0.00427, limits of agreement of -0.7–0.5 (36.8°C–38.0°C), p=0.037 and outliers 4% at 1.96 SD with GMT as the standard against IR and DT, respectively. The area under the receiver operator curve values were 0.811 and 0.923 for GMT versus IR and GMT versus DT, respectively. Conclusions: Digital and infrared temperature readings are in reasonable agreement with a glass mercury thermometer and both can be used interchangeably with a glass mercury thermometer. Compared with infrared thermometers, digital thermometers are more closely related to glass mercury thermometers.

Authors

Institutions

Publication Details

Journal
International Journal of Contemporary Pediatrics
Published
2025-05-26
DOI
https://doi.org/10.18203/2349-3291.ijcp20251465
Primary Topic
Thermal Regulation in Medicine
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Agreement among glass mercury thermometers, infrared thermometers and digital thermometer temperature recordings in febrile under-five children

Ernest Ndukaife Anyabolu, Esther Ngozi Umeadi, Sylvia Tochukwu Echendu, Njideka C. Uchefuna et al.
International Journal of Contemporary Pediatrics
Thermal Regulation in Medicine
article

Agreement among glass mercury thermometers, infrared thermometers and digital thermometer temperature recordings in febrile under-five children

Ernest Ndukaife Anyabolu, Esther Ngozi Umeadi, Sylvia Tochukwu Echendu, Njideka C. Uchefuna, Kenneth Nchekwube Okeke, Boniface E. Chukwujekwu, Ann Oluchi Ikeabbah, Nkechinyere G. Obichukwu, Chigozie K. Onyeje
article en

Abstract

Background: This study aimed to assess the degree of agreement between the temperature readings of a glass mercury thermometer (GMT), a digital thermometer (DT) and an infrared thermometer (IR) in paediatric patients. This was a comparative study conducted at a tertiary institution in Nnewi, Anambra State, Southeast Nigeria. One hundred febrile children <5 years of age were recruited via convenience sampling. Methods: Temperatures were measured via an axillary GMT, an axillary DT and a noncontact forehead IR thermometer. Data were compared among the three methods via a 2×2 contingency table, receiver operating curve (ROC) and Bland Altman plot with GMT as the standard. Results: There was a greater significant correlation between GMT and DT (r=0.901, p=0.001) than between GMT and IR (r=0.695, p=0.001). Bland Altman plots revealed that IR and DT could be used interchangeably with GMT with an arithmetic mean of 1.1°C, confidence interval of 0.9359 to 1.1581, limits of agreement of -0.1–2.1 (37.4°C–39.6°C), p<0.001, outliers 3% at 1.96 SD and arithmetic mean of -0.1°C, confidence interval of -0.1337-0.00427, limits of agreement of -0.7–0.5 (36.8°C–38.0°C), p=0.037 and outliers 4% at 1.96 SD with GMT as the standard against IR and DT, respectively. The area under the receiver operator curve values were 0.811 and 0.923 for GMT versus IR and GMT versus DT, respectively. Conclusions: Digital and infrared temperature readings are in reasonable agreement with a glass mercury thermometer and both can be used interchangeably with a glass mercury thermometer. Compared with infrared thermometers, digital thermometers are more closely related to glass mercury thermometers.

International Journal of Contemporary PediatricsVol. 12(6)
Chukwuemeka Odumegwu Ojukwu University (NG), Nnamdi Azikiwe University (NG), Nnamdi Azikiwe University Teaching Hospital (NG)
Openalex Percentile: Top 85%
Thermal Regulation in Medicine
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.