Swift and accurate detection of tacrolimus concentration post-organ transplantation using a photonic crystal chip

Tacrolimus (FK506) is a potent immunosuppressant widely used to prevent rejection following organ transplantation. In this study, we employed a photonic crystal microarray-based competitive assay to measure Tacrolimus blood concentrations, facilitating bedside and at-home monitoring for transplant recipients. The photonic crystal microarray technology offers flexibility, cost-effectiveness, and high efficiency. When integrated with a portable fluorescence intensity detector, it enables rapid and quantitative analysis. By fabricating the photonic crystal microarray and leveraging the photonic crystal bandgap enhancement effect to amplify the fluorescence intensity of probe molecules, the detection sensitivity is significantly improved. The photonic crystal microarray demonstrated a remarkable fluorescence enhancement factor of 17.2-fold compared to conventional substrates, significantly improving the sensitivity of Tacrolimus detection. The optimized system achieved a detection limit of 0.4 ng·mL<sup>-1</sup>, enabling rapid and accurate quantification of Tacrolimus concentration within 20 min.

Authors

Publication Details

Journal
Nano Research
Published
2025-05-26
DOI
https://doi.org/10.26599/nr.2025.94907619
Primary Topic
Spectroscopy Techniques in Biomedical and Chemical Research
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Swift and accurate detection of tacrolimus concentration post-organ transplantation using a photonic crystal chip

Zewei Lian, Meng Su, Yingli Li, Weiping Liu et al.
1 links
Nano Research
Spectroscopy Techniques in Biomedical and Chemical Research
article

Swift and accurate detection of tacrolimus concentration post-organ transplantation using a photonic crystal chip

Zewei Lian, Meng Su, Yingli Li, Weiping Liu, Cui Zhang, Bin Shi, Fangwei Wang, Wenjing Lü, Yicong Geng, Xueping Ma
article en

Abstract

Tacrolimus (FK506) is a potent immunosuppressant widely used to prevent rejection following organ transplantation. In this study, we employed a photonic crystal microarray-based competitive assay to measure Tacrolimus blood concentrations, facilitating bedside and at-home monitoring for transplant recipients. The photonic crystal microarray technology offers flexibility, cost-effectiveness, and high efficiency. When integrated with a portable fluorescence intensity detector, it enables rapid and quantitative analysis. By fabricating the photonic crystal microarray and leveraging the photonic crystal bandgap enhancement effect to amplify the fluorescence intensity of probe molecules, the detection sensitivity is significantly improved. The photonic crystal microarray demonstrated a remarkable fluorescence enhancement factor of 17.2-fold compared to conventional substrates, significantly improving the sensitivity of Tacrolimus detection. The optimized system achieved a detection limit of 0.4 ng·mL<sup>-1</sup>, enabling rapid and accurate quantification of Tacrolimus concentration within 20 min.

Nano ResearchVol. 18(8)
National Natural Science Foundation of China, National Key Research and Development Program of China, Beijing National Laboratory for Molecular Sciences
Zero hunger
Openalex Percentile: Top 91%
Spectroscopy Techniques in Biomedical and Chemical Research
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.