COMPARISON OF AUTOMATED COMPLETE BLOOD COUNT PARAMETERS AND HEMOLYSIS OCCURRENCE FROM VACUUM AND NON-VACUUM K2EDTA TUBES

Background: The preanalytical factors, including tube type and storage duration, may influence CBC results and hemolysis. Nowadays, blood collection using K₂EDTA, both vacuum and non-vacuum tubes, is widely utilized in clinical laboratories for complete blood count (CBC) analysis. However, the impact of vacuum-induced negative pressure and the anticoagulant-to-blood ratio in under-filled tubes on sample integrity remains a concern. Objective: This study aimed to compare the performance of K₂EDTA vacuum and non-vacuum blood collection tubes, including under-filled non-vacuum tubes, in automated CBC parameters and to assess hemolysis occurrence at different time points. Methods: Blood samples from 30 healthy subjects were collected into three different K₂EDTA blood collection tubes: vacuum, non-vacuum, and non-vacuum tubes with under-filled volumes. CBC parameters, including RBC, Hb, Hct, MCV, MCH, MCHC, RDW, WBC, PLT, and MPV, were analyzed at 0 hours, 3 hours, and 24 hours post-collection. Hemolysis was assessed via spectrophotometry at 540 nm. Results: No statistically significant differences were observed in CBC parameters among the three groups at 0 hours. However, significant differences occurred at 3 hours post-collection, including WBC and PLT in vacuum tubes, MCV in non-vacuum tubes, RDW in under-filled tubes, and MPV in all tube types. At 24 hours, all tube types showed significant changes in MCV, MCHC, RDW, and MPV, while RBC showed only changes in vacuum tubes. The highest hemolysis levels occurred in vacuum tubes at 0 hours, whereas under-filled tubes exhibited increased hemolysis at 3 and 24 hours. Conclusion: The results indicated that all three tube types are suitable for hematological analysis if processed within 3 hours. However, the risk of hemolysis in vacuum tubes should be carefully considered, particularly in individuals with suspected hemolysis.

Authors

Institutions

Publication Details

Journal
Journal of Southeast Asian Medical Research
Published
2025-05-26
DOI
https://doi.org/10.55374/jseamed.v9.235
Primary Topic
Clinical Laboratory Practices and Quality Control
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

COMPARISON OF AUTOMATED COMPLETE BLOOD COUNT PARAMETERS AND HEMOLYSIS OCCURRENCE FROM VACUUM AND NON-VACUUM K2EDTA TUBES

Punnee Butthep, Thirayost Nimmanon, Dollapak Apipongrat, Pasra Arnutti et al.
Journal of Southeast Asian Medical Research
Clinical Laboratory Practices and Quality Control
article

COMPARISON OF AUTOMATED COMPLETE BLOOD COUNT PARAMETERS AND HEMOLYSIS OCCURRENCE FROM VACUUM AND NON-VACUUM K2EDTA TUBES

Punnee Butthep, Thirayost Nimmanon, Dollapak Apipongrat, Pasra Arnutti, Montalee Theeraapisakkun
article en

Abstract

Background: The preanalytical factors, including tube type and storage duration, may influence CBC results and hemolysis. Nowadays, blood collection using K₂EDTA, both vacuum and non-vacuum tubes, is widely utilized in clinical laboratories for complete blood count (CBC) analysis. However, the impact of vacuum-induced negative pressure and the anticoagulant-to-blood ratio in under-filled tubes on sample integrity remains a concern. Objective: This study aimed to compare the performance of K₂EDTA vacuum and non-vacuum blood collection tubes, including under-filled non-vacuum tubes, in automated CBC parameters and to assess hemolysis occurrence at different time points. Methods: Blood samples from 30 healthy subjects were collected into three different K₂EDTA blood collection tubes: vacuum, non-vacuum, and non-vacuum tubes with under-filled volumes. CBC parameters, including RBC, Hb, Hct, MCV, MCH, MCHC, RDW, WBC, PLT, and MPV, were analyzed at 0 hours, 3 hours, and 24 hours post-collection. Hemolysis was assessed via spectrophotometry at 540 nm. Results: No statistically significant differences were observed in CBC parameters among the three groups at 0 hours. However, significant differences occurred at 3 hours post-collection, including WBC and PLT in vacuum tubes, MCV in non-vacuum tubes, RDW in under-filled tubes, and MPV in all tube types. At 24 hours, all tube types showed significant changes in MCV, MCHC, RDW, and MPV, while RBC showed only changes in vacuum tubes. The highest hemolysis levels occurred in vacuum tubes at 0 hours, whereas under-filled tubes exhibited increased hemolysis at 3 and 24 hours. Conclusion: The results indicated that all three tube types are suitable for hematological analysis if processed within 3 hours. However, the risk of hemolysis in vacuum tubes should be carefully considered, particularly in individuals with suspected hemolysis.

Journal of Southeast Asian Medical ResearchVol. 9
University of Health and Allied Sciences (GH), Phramongkutklao Hospital (TH), Phramongkutklao College of Medicine
Office of Research and Development
Openalex Percentile: Top 85%
Clinical Laboratory Practices and Quality Control
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.