A contamination-free electrolyte-gated organic transistors platform for high-accuracy tumor biomarker detection

• A contamination free microfluidics for increasing the signal-to-noise ratio has been successfully developed and tested. • P3CPT inkjet deposition method, patterned gate electrodes and 3D printed package improve the stability and repeatability. • Ang2 tumor biomarker has been detected at concentrations as low as 10 pM in a dilute phosphate buffer. • Measurements with a different tumor biomarker, VEGF, were also performed to demonstrate the selectivity. A novel biosensor platform for high-accuracy tumor biomarker detection exploiting contamination free microfluidics for increasing the signal-to-noise ratio has been successfully developed and tested. Electrolyte-gated organic Transistors (EGOT) has been employed to detect an important tumor marker, Angiopoietin-2 (Ang2). Although organic semiconductors have become popular in the last years in biosensing applications due to their many advantages, there is still a main concern about stability and selectivity. This work presents major improvement in terms of the stability and selective detection of Ang2 in the range of interest for biomedical applications. The semiconducting polymer poly[3-(5-carboxypentyl)thiophene (P3CPT) is deposited by picoliter volume control and micrometer diameter of the droplet to allow for high uniformity and repeatability from sample to sample. The optimized gold electrodes improve the detection of the minimal concentration of the target and microfluidic interfacing by a specific pattern with the desired dimensions is obtained by UV-lithography and wet etching. A microfluidics with multiple flow control allows for maintain a constant fresh solution without analytes on reference gate electrode, while another inlet and functionalized gate is used for sensing, thus reaching high stability and reproducibility. All these (four) optimizations lead to new measuring protocol and new 3D printed top cover that ensure better stabilization and repeatability of the results. The device has successfully detected Ang2 concentrations as low as 10 pM in saline, therefore demonstrating the ability of the device to detect clinically relevant concentrations.

Authors

Institutions

Publication Details

Journal
Sensors and Actuators Reports
Published
2025-05-25
DOI
https://doi.org/10.1016/j.snr.2025.100341
Citations
1
Primary Topic
Analytical Chemistry and Sensors
Type
article
Field-Weighted Citation Impact
0.74
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A contamination-free electrolyte-gated organic transistors platform for high-accuracy tumor biomarker detection

Matteo Parmeggiani, Alberto Ballesio, Lucia Napione, Francesca Frascella et al.
1 citations
Sensors and Actuators Reports
Analytical Chemistry and Sensors
0.74
article

A contamination-free electrolyte-gated organic transistors platform for high-accuracy tumor biomarker detection

Matteo Parmeggiani, Alberto Ballesio, Lucia Napione, Francesca Frascella, Simone Luigi Marasso, Jovana Babic, Mattia Pagani
article en
1 citations

Abstract

• A contamination free microfluidics for increasing the signal-to-noise ratio has been successfully developed and tested. • P3CPT inkjet deposition method, patterned gate electrodes and 3D printed package improve the stability and repeatability. • Ang2 tumor biomarker has been detected at concentrations as low as 10 pM in a dilute phosphate buffer. • Measurements with a different tumor biomarker, VEGF, were also performed to demonstrate the selectivity. A novel biosensor platform for high-accuracy tumor biomarker detection exploiting contamination free microfluidics for increasing the signal-to-noise ratio has been successfully developed and tested. Electrolyte-gated organic Transistors (EGOT) has been employed to detect an important tumor marker, Angiopoietin-2 (Ang2). Although organic semiconductors have become popular in the last years in biosensing applications due to their many advantages, there is still a main concern about stability and selectivity. This work presents major improvement in terms of the stability and selective detection of Ang2 in the range of interest for biomedical applications. The semiconducting polymer poly[3-(5-carboxypentyl)thiophene (P3CPT) is deposited by picoliter volume control and micrometer diameter of the droplet to allow for high uniformity and repeatability from sample to sample. The optimized gold electrodes improve the detection of the minimal concentration of the target and microfluidic interfacing by a specific pattern with the desired dimensions is obtained by UV-lithography and wet etching. A microfluidics with multiple flow control allows for maintain a constant fresh solution without analytes on reference gate electrode, while another inlet and functionalized gate is used for sensing, thus reaching high stability and reproducibility. All these (four) optimizations lead to new measuring protocol and new 3D printed top cover that ensure better stabilization and repeatability of the results. The device has successfully detected Ang2 concentrations as low as 10 pM in saline, therefore demonstrating the ability of the device to detect clinically relevant concentrations.

Sensors and Actuators ReportsVol. 9
Politecnico di Torino (IT), Institute of Materials for Electronics and Magnetism (IT)
Openalex Percentile: Top 33%
Analytical Chemistry and Sensors
0.74
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.