Hanita Intensity, a New 5-Focal Intraocular Lens for Cataract and Refractive Lens Exchange Patients: Eighteen-Months Results

Ninety-one implantations of the Hanita Intensity, intraocular lens, including 34 patients in both eyes were performed over 18 months.In the majority (55 eyes), implantation was performed during refractive lens exchange, and in the remaining 36 cases, it was performed because of cataract.The qualifying examination, apart from the evaluation of standard parameters, included the evaluation of kappa and alpha angles, value of higher-order aberrations generated by the cornea, and pupil size.High postoperative satisfaction due to very good vision parameters (more than 80% of operated patients achieved vision to far distances 0.1 logMAR in the cataract group and 87% in the refractive lens exchange group, and almost 100% vision to near distances at the level of D-0.5 (Snellen chart) in the refractive lens exchange group) allows us to recommend this lens model in the most demanding group of patients qualified for refractive lens exchange.Statistical analysis showed that patients with the Hanita Intensity lens had significantly better visual acuity to far distances after surgery than those with the PanOptix (Alcon) lens.Conflict of Interest declaration.

Authors

Publication Details

Journal
Archives of Ophthalmology and Optometry
Published
2025-05-23
DOI
https://doi.org/10.22259/2638-5120.0601004
Primary Topic
Ophthalmology and Visual Impairment Studies
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Hanita Intensity, a New 5-Focal Intraocular Lens for Cataract and Refractive Lens Exchange Patients: Eighteen-Months Results

Adam Cywiński, Paulina Bazgier
Archives of Ophthalmology and Optometry
Ophthalmology and Visual Impairment Studies
article

Hanita Intensity, a New 5-Focal Intraocular Lens for Cataract and Refractive Lens Exchange Patients: Eighteen-Months Results

Adam Cywiński, Paulina Bazgier
article en

Abstract

Ninety-one implantations of the Hanita Intensity, intraocular lens, including 34 patients in both eyes were performed over 18 months.In the majority (55 eyes), implantation was performed during refractive lens exchange, and in the remaining 36 cases, it was performed because of cataract.The qualifying examination, apart from the evaluation of standard parameters, included the evaluation of kappa and alpha angles, value of higher-order aberrations generated by the cornea, and pupil size.High postoperative satisfaction due to very good vision parameters (more than 80% of operated patients achieved vision to far distances 0.1 logMAR in the cataract group and 87% in the refractive lens exchange group, and almost 100% vision to near distances at the level of D-0.5 (Snellen chart) in the refractive lens exchange group) allows us to recommend this lens model in the most demanding group of patients qualified for refractive lens exchange.Statistical analysis showed that patients with the Hanita Intensity lens had significantly better visual acuity to far distances after surgery than those with the PanOptix (Alcon) lens.Conflict of Interest declaration.

Archives of Ophthalmology and OptometryVol. 6(1)
Openalex Percentile: Top 83%
Ophthalmology and Visual Impairment Studies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Hanita Intensity, a New 5-Focal Intraocular Lens for Cataract and Refractive Lens Exchange Patients: Eighteen-Months Results — Adam Cywiński, Paulina Bazgier · Archives of Ophthalmology and Optometry (2025) | TGRS Research Map | TGRS