Long-term low-voltage impedance measurements in subcutaneous implantable cardioverter-defibrillators

Abstract Background High-voltage impedance (HVI) measurement is commonly performed during subcutaneous implantable cardioverter-defibrillator (S-ICD) implantation, either through 65 J/80 J defibrillation testing or with a synchronized 10 J shock. HVI has been associated with defibrillation efficacy. Recently, S-ICD systems have been upgraded to provide automatic long-term measurements of low-voltage impedance (LVI) using a subthreshold 1 V pulse. This study aimed to evaluate the performance of LVI as a surrogate for HVI and to describe the long-term trend of LVI in S-ICD recipients. Methods Consecutive patients undergoing de novo S-ICD implantation were enrolled across 13 Italian centers. Data were collected from the LATITUDE remote monitoring database, where the S-ICD recorded LVI measurements every three days. To assess agreement between LVI and HVI, HVI values recorded at the time of each 80 J shock delivered during follow-up were compared with the nearest LVI values stored by the device. Results Data from 729 patients (mean age 49±15 years, 80% male, BMI 26±5 kg/m²) were analyzed between 2015 and 2023. The median follow-up duration was 39 months [IQR: 21–56 months]. During follow-up, 130 patients (18%) received shocks. A total of 225 paired HVI and LVI measurements were available for analysis. LVI values were closely correlated with HVI values (r = 0.91; 95% CI: 0.88–0.93; P < 0.001). Bland-Altman analysis, accounting for repeated measurements per patient, showed a bias of -4 Ohm between LVI and HVI, with limits of agreement (bias ± 2SD) from -24 to 15 Ohm. Figure 1 illustrates the long-term trend of average LVI in the cohort: mean LVI rose significantly within the first 10 weeks post-implantation (reaching 90% of the long-term value), stabilizing at a long-term mean of 79±1 Ohm (values measured after 52 weeks) without further variation over time. Conclusions LVI measurements by S-ICD demonstrated strong agreement with HVI values measured during 80 J shock delivery, suggesting that LVI may serve as a surrogate marker for defibrillation efficacy at implantation, without requiring shock delivery. Long-term LVI monitoring could provide continuous insight into system integrity. The observed progressive increase in LVI during the initial weeks post-implantation likely reflects inflammatory and healing processes, with stable values thereafter.

Authors

Institutions

Publication Details

Journal
EP Europace
Published
2025-05-01
DOI
https://doi.org/10.1093/europace/euaf085.766
Primary Topic
Muscle activation and electromyography studies
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Long-term low-voltage impedance measurements in subcutaneous implantable cardioverter-defibrillators

Paolo De Filippo, Luca Tomasi, Giacomo Mugnai, Luca Ottaviano et al.
EP Europace
Muscle activation and electromyography studies
article

Long-term low-voltage impedance measurements in subcutaneous implantable cardioverter-defibrillators

Paolo De Filippo, Luca Tomasi, Giacomo Mugnai, Luca Ottaviano, S Valsecchi, Gianfranco Tola, G Ricciardi, Carmelo La Greca, Gerardo Nigro, Stefano Viani, Roberto Rordorf, Valter Bianchi, Giovanni Luca Botto, Giovanni Bisignani, Matteo Ziacchi
article en

Abstract

Abstract Background High-voltage impedance (HVI) measurement is commonly performed during subcutaneous implantable cardioverter-defibrillator (S-ICD) implantation, either through 65 J/80 J defibrillation testing or with a synchronized 10 J shock. HVI has been associated with defibrillation efficacy. Recently, S-ICD systems have been upgraded to provide automatic long-term measurements of low-voltage impedance (LVI) using a subthreshold 1 V pulse. This study aimed to evaluate the performance of LVI as a surrogate for HVI and to describe the long-term trend of LVI in S-ICD recipients. Methods Consecutive patients undergoing de novo S-ICD implantation were enrolled across 13 Italian centers. Data were collected from the LATITUDE remote monitoring database, where the S-ICD recorded LVI measurements every three days. To assess agreement between LVI and HVI, HVI values recorded at the time of each 80 J shock delivered during follow-up were compared with the nearest LVI values stored by the device. Results Data from 729 patients (mean age 49±15 years, 80% male, BMI 26±5 kg/m²) were analyzed between 2015 and 2023. The median follow-up duration was 39 months [IQR: 21–56 months]. During follow-up, 130 patients (18%) received shocks. A total of 225 paired HVI and LVI measurements were available for analysis. LVI values were closely correlated with HVI values (r = 0.91; 95% CI: 0.88–0.93; P < 0.001). Bland-Altman analysis, accounting for repeated measurements per patient, showed a bias of -4 Ohm between LVI and HVI, with limits of agreement (bias ± 2SD) from -24 to 15 Ohm. Figure 1 illustrates the long-term trend of average LVI in the cohort: mean LVI rose significantly within the first 10 weeks post-implantation (reaching 90% of the long-term value), stabilizing at a long-term mean of 79±1 Ohm (values measured after 52 weeks) without further variation over time. Conclusions LVI measurements by S-ICD demonstrated strong agreement with HVI values measured during 80 J shock delivery, suggesting that LVI may serve as a surrogate marker for defibrillation efficacy at implantation, without requiring shock delivery. Long-term LVI monitoring could provide continuous insight into system integrity. The observed progressive increase in LVI during the initial weeks post-implantation likely reflects inflammatory and healing processes, with stable values thereafter.

EP EuropaceVol. 27(Supplement_1)
Boston Scientific (United States) (US), IRCCS Azienda Ospedliero-Universitaria di Bologna Policlinico di Sant'Orsola (IT), Istituto Clinico Sant'Ambrogio (IT), Ospedale Papa Giovanni XXIII (IT), Azienda Ospedaliero-Universitaria Careggi (IT), Ospedale Monaldi (IT), Mylan (Switzerland) (CH), Fondazione Poliambulanza Istituto Ospedaliero (IT), Policlinico San Matteo Fondazione (IT), Ferrari (Italy) (IT), Azienda Ospedaliera Universitaria Pisana (IT), Azienda Ospedaliera G. Brotzu (IT)
Openalex Percentile: Top 89%
Muscle activation and electromyography studies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.