The continous grid model: a novel approach to improve spatial resolution and omnipolar signal reliability for electroanatomical mapping of arrhythmogenic substrates

Abstract Introduction Multipolar catheters are widely used for clinical electro-anatomical mapping to identify ablation targets. Different catheter configurations have been proposed. Grid-like catheter designs with fixed electrode arrangement allow to explore wavefront characteristics and estimate omnipolar electrograms. Yet, particularly in diseased tissue their interelectrode spacing of 4 mm often does not meet critical assumptions regarding uniformity of the unipolar signal between electrodes or the presence of a planar wavefront that are essential for accurate omnipolar signal calculations. This failure to meet the assumptions can be attributed to the spatial resolution, as the electrodes used in the omnipolar calculation may be placed in areas with tissue heterogeneity or poor electrode contact. Purpose This study aimed to assess omnipolar signal reliability of clinically employed grid-catheters by evaluating the catheters ability to meet the assumptions of uniformity of unipolar signals between electrodes and presence of planar wavefront in each clique by analyzing three developed parameters: amplitude variability (σa), morphology variability (σm), and Non-Planarity (θ). Methods The Continuous Grid (C-Grid) model was developed to address issues with interelectrode spacing by spatially interpolating unipolar signals based on the originally recorded signals. This model uses amplitude, lag, and morphology as variables in an equation that allows for the computation of the unipolar signal at any point within the catheter's area. The interpolation relies on spline equations for amplitude and lag, while morphology is handled using inverse Euclidean distance. Through resampling within this model, the interelectrode distance can be virtually reduced, enhancing spatial resolution (Figure 1.A). Results Two ventricular substrate maps with a total of 4,311 mapping points were studied. Results indicate that with a 4 mm interelectrode spacing of clinically employed grid-catheters, the assumptions are weakly met, showing 17.71±9.73% σa, 4.38±4.56% σm, and 22.17±25.13% θ. As the inter-electrode spacing decreases, adherence to the underlying assumptions improves, thereby enabling the computation of more reliable omnipolar signals (Figure 1.B). However, based on the C-Grid results, achieving a tolerance below 5% for the three parameters would require less than 0.5 mm spacing, which is currently a challenge to develop from the hardware perspective. Conclusion The novel C-Grid model approach allows for more accurate omnipolar signal calculations and thus, derived biomarkers. Improvement in mapping accuracy with the C-Grid model method may support more precise arrhythmogenic substrate characterization and could improve identification of ablation targets in clinical procedures.

Authors

Institutions

Publication Details

Journal
EP Europace
Published
2025-05-01
DOI
https://doi.org/10.1093/europace/euaf085.075
Citations
1
Primary Topic
Electrochemical Analysis and Applications
Type
article
Field-Weighted Citation Impact
0.52

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

The continous grid model: a novel approach to improve spatial resolution and omnipolar signal reliability for electroanatomical mapping of arrhythmogenic substrates

Pier D. Lambiase, José Millet, Elisa Ramírez, Francisco Castells et al.
1 citations
EP Europace
Electrochemical Analysis and Applications
0.52
article

The continous grid model: a novel approach to improve spatial resolution and omnipolar signal reliability for electroanatomical mapping of arrhythmogenic substrates

Pier D. Lambiase, José Millet, Elisa Ramírez, Francisco Castells, Samuel Ruipérez-Campillo, Johanna Tonko, R Alós
article en
1 citations

Abstract

Abstract Introduction Multipolar catheters are widely used for clinical electro-anatomical mapping to identify ablation targets. Different catheter configurations have been proposed. Grid-like catheter designs with fixed electrode arrangement allow to explore wavefront characteristics and estimate omnipolar electrograms. Yet, particularly in diseased tissue their interelectrode spacing of 4 mm often does not meet critical assumptions regarding uniformity of the unipolar signal between electrodes or the presence of a planar wavefront that are essential for accurate omnipolar signal calculations. This failure to meet the assumptions can be attributed to the spatial resolution, as the electrodes used in the omnipolar calculation may be placed in areas with tissue heterogeneity or poor electrode contact. Purpose This study aimed to assess omnipolar signal reliability of clinically employed grid-catheters by evaluating the catheters ability to meet the assumptions of uniformity of unipolar signals between electrodes and presence of planar wavefront in each clique by analyzing three developed parameters: amplitude variability (σa), morphology variability (σm), and Non-Planarity (θ). Methods The Continuous Grid (C-Grid) model was developed to address issues with interelectrode spacing by spatially interpolating unipolar signals based on the originally recorded signals. This model uses amplitude, lag, and morphology as variables in an equation that allows for the computation of the unipolar signal at any point within the catheter's area. The interpolation relies on spline equations for amplitude and lag, while morphology is handled using inverse Euclidean distance. Through resampling within this model, the interelectrode distance can be virtually reduced, enhancing spatial resolution (Figure 1.A). Results Two ventricular substrate maps with a total of 4,311 mapping points were studied. Results indicate that with a 4 mm interelectrode spacing of clinically employed grid-catheters, the assumptions are weakly met, showing 17.71±9.73% σa, 4.38±4.56% σm, and 22.17±25.13% θ. As the inter-electrode spacing decreases, adherence to the underlying assumptions improves, thereby enabling the computation of more reliable omnipolar signals (Figure 1.B). However, based on the C-Grid results, achieving a tolerance below 5% for the three parameters would require less than 0.5 mm spacing, which is currently a challenge to develop from the hardware perspective. Conclusion The novel C-Grid model approach allows for more accurate omnipolar signal calculations and thus, derived biomarkers. Improvement in mapping accuracy with the C-Grid model method may support more precise arrhythmogenic substrate characterization and could improve identification of ablation targets in clinical procedures.

EP EuropaceVol. 27(Supplement_1)
St Bartholomew's Hospital (GB), ETH Zurich (CH), University College London (GB), Universitat Politècnica de València (ES)
Ministerio de Ciencia e Innovación
Openalex Percentile: Top 43%
Electrochemical Analysis and Applications
0.52
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.