Comparison of the Isometric Position of the External Elbow Fixator: Self-Centering Versus Traditional Techniques, Postoperative CT Evaluation, and 3D Motion Analysis

<b>Background/Objectives</b>: The external hinged elbow fixator is a surgical choice both in the case of simple dislocations and elbow dislocation fractures. The correct positioning with respect to the elbow's center of rotation is demanding. Authors developed a self-centering external fixator that does not require a pin in the elbow's center of rotation. The aim of this study was to analyze the margin of error in its positioning. <b>Methods:</b> We subjected 16 patients to a CT-3D study reconstruction using 3D motion software to analyze the divergence angle and offset between the elbow's center of rotation and that of the external fixator. The results were compared to those published on traditional implants. <b>Results:</b> All elbows were correctly reduced without re-dislocation. The average distance was 2° in relation to the center of rotation in the antero-posterior view, 3° in the cranio-caudal, and 2° in the medio-lateral. The divergence angle was 3.5° (min 0.4°; max 9.3°) and the offset 6.8 mm (min 0.06; max 17.5). The average range of motion was 10-145 (range 0-155). <b>Discussion:</b> The traditional hinged elbow external fixator creates severe complexity for surgeons in the necessary positioning of the elbow axial rod to correctly align the implant. The self-centering device avoids this step, making the procedure faster and easier. Although the alignment is still not perfect, the results are still comparable with traditional devices. <b>Conclusions:</b> The self-centering external fixator allows for correct alignment with the elbow's center of rotation. It is less invasive and simpler, with a shorter learning curve, faster operating time, and less radiographic exposure.

Authors

Institutions

Publication Details

Journal
Journal of Clinical Medicine
Published
2025-05-23
DOI
https://doi.org/10.3390/jcm14113653
Primary Topic
Elbow and Forearm Trauma Treatment
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Comparison of the Isometric Position of the External Elbow Fixator: Self-Centering Versus Traditional Techniques, Postoperative CT Evaluation, and 3D Motion Analysis

Andrea Poggetti, Sandra Pfanner, Irene Felici, Chiara Suardi et al.
Journal of Clinical Medicine
Elbow and Forearm Trauma Treatment
article

Comparison of the Isometric Position of the External Elbow Fixator: Self-Centering Versus Traditional Techniques, Postoperative CT Evaluation, and 3D Motion Analysis

Andrea Poggetti, Sandra Pfanner, Irene Felici, Chiara Suardi, Marco Biondi, Prospero Bigazzi, Anna Rosa Rizzo
article en

Abstract

<b>Background/Objectives</b>: The external hinged elbow fixator is a surgical choice both in the case of simple dislocations and elbow dislocation fractures. The correct positioning with respect to the elbow's center of rotation is demanding. Authors developed a self-centering external fixator that does not require a pin in the elbow's center of rotation. The aim of this study was to analyze the margin of error in its positioning. <b>Methods:</b> We subjected 16 patients to a CT-3D study reconstruction using 3D motion software to analyze the divergence angle and offset between the elbow's center of rotation and that of the external fixator. The results were compared to those published on traditional implants. <b>Results:</b> All elbows were correctly reduced without re-dislocation. The average distance was 2° in relation to the center of rotation in the antero-posterior view, 3° in the cranio-caudal, and 2° in the medio-lateral. The divergence angle was 3.5° (min 0.4°; max 9.3°) and the offset 6.8 mm (min 0.06; max 17.5). The average range of motion was 10-145 (range 0-155). <b>Discussion:</b> The traditional hinged elbow external fixator creates severe complexity for surgeons in the necessary positioning of the elbow axial rod to correctly align the implant. The self-centering device avoids this step, making the procedure faster and easier. Although the alignment is still not perfect, the results are still comparable with traditional devices. <b>Conclusions:</b> The self-centering external fixator allows for correct alignment with the elbow's center of rotation. It is less invasive and simpler, with a shorter learning curve, faster operating time, and less radiographic exposure.

Journal of Clinical MedicineVol. 14(11)
Azienda Ospedaliero-Universitaria Careggi (IT)
Openalex Percentile: Top 85%
Elbow and Forearm Trauma Treatment
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.