Reassessment of peripheral nerve stimulation thresholds for the Impulse model‐optimized asymmetric head gradient coil
The PNS threshold of the x-axis measured in the scanner was much lower than predicted by the optimization model and as measured at the factory. Our measurements verified that PNS thresholds of asymmetric head gradient coils were sensitive to head position, subject variability, and age. The discrepancy of the PNS prediction model remains to be elucidated.
Authors
- Alexander J. S. Beckett (ORCID: https://orcid.org/0000-0002-7123-9559)
- Nicolas Boulant (ORCID: https://orcid.org/0000-0003-2144-2484)
- Erica Walker (ORCID: https://orcid.org/0000-0001-5066-1403)
- Mathias Davids (ORCID: https://orcid.org/0000-0003-0217-9123)
- Elmar Rummert
- J. Samantha (ORCID: https://orcid.org/0000-0003-3127-455X)
- Peter Dietz (ORCID: https://orcid.org/0009-0002-1803-0892)
- David A. Feinberg
- Dominik Rattenbacher
Institutions
- Centre National de la Recherche Scientifique (FR)
- Allen Institute for Brain Science (US)
- Harvard University (US)
- Neurosciences Institute (US)
- Commissariat à l'Énergie Atomique et aux Énergies Alternatives (FR)
- Université Paris-Saclay (FR)
- Massachusetts General Hospital (US)
- Advanced MRI Technologies (United States) (US)
- Athinoula A. Martinos Center for Biomedical Imaging (US)
- CEA Paris-Saclay (FR)
- Siemens Healthineers (Germany) (DE)
- Imaging Center (US)
- University of California, Berkeley (US)
Publication Details
- Journal
- Magnetic Resonance in Medicine
- Published
- 2025-05-23
- DOI
- https://doi.org/10.1002/mrm.30523
- Citations
- 2
- Primary Topic
- Transcranial Magnetic Stimulation Studies
- Type
- article
- Field-Weighted Citation Impact
- 1.68
Funders
- Horizon 2020 Framework Programme
- National Institute of Mental Health
- National Institute of Biomedical Imaging and Bioengineering