In silico modeling of cardiac defibrillation with novel deep septal ICD leads

Abstract Background Recent studies have demonstrated that implantable cardioverter defibrillator (ICD) leads placed higher and deeper into the right ventricular (RV) septum, compared to traditional RV apical placement, were able to achieve conduction system pacing (CSP). A higher and deeper septal position may impact defibrillation thresholds (DFT) compared to the traditional ICD lead implant position. Purpose To compare DFTs associated with ICD leads implanted within the deep septum (CSP group) vs. an apical position (traditional). Methods In silico torso/heart models were generated from 5 human subjects (Figure). A commercially available ICD lead and a prototype 7Fr lead with a 5cm defibrillation coil were modeled in software, for apical and CSP positions, respectively, with a perpendicular insertion angle and nominal lead slack in the RV. For each torso model, a stable lead pathway was generated following simulated dynamic cardiac motion, for multiple positions: apical (control), deep high (1.5cm from tricuspid valve), deep mid (≈3.5cm from tricuspid valve), and deep low (midpoint of mid and apical) septal. The shock vector was between RV coil and ICD can placed at the 2nd rib. DFT energy was recorded at >5V/cm in 95% of ventricular tissue. Results are reported as median (range). Results The 5 torso models demonstrated a median left ventricular volume of 200cm³ (163 to 266cm³), 66Ω (56 to 77Ω) shock vector impedance, and 12cm (10 to 13cm) coil-to-can distance. For the apical position, the DFT was 18.4J (6.9 to 28.9J, n=5). As compared to apical, the CSP-positioned DFTs overall increased 1.4J (-0.6 to 4.4J, n=5), across all subjects. DFTs for deep high and mid septal nominally increased by 1.1J (-2.4 to 4.4J, n=5) and 1.8J (0.2 to 2.2J, n=3), respectively. The DFT for deep low septal slightly decreased by 0.3J (0.1 to 0.6J, n=2). All DFTs were substantially below defibrillation energies current ICDs offer. Conclusion Based on human torso models, ICD leads implanted in CSP positions resulted in a slight increase in DFTs compared to RV apical placement. With the maximum energies current ICDs offer, the slight DFT increase is not expected to substantially affect defibrillation efficacy. Larger clinical studies are needed to replicate these findings.In silico defibrillation modeling

Authors

Institutions

Publication Details

Journal
EP Europace
Published
2025-05-01
DOI
https://doi.org/10.1093/europace/euaf085.587
Citations
1
Primary Topic
Cardiac pacing and defibrillation studies
Type
article
Field-Weighted Citation Impact
1.39
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

In silico modeling of cardiac defibrillation with novel deep septal ICD leads

Aurel Neic, David R. Booth, Shunmuga Sundaram Ponnusamy, Guram Imnadze et al.
1 citations
EP Europace
Cardiac pacing and defibrillation studies
1.39
article

In silico modeling of cardiac defibrillation with novel deep septal ICD leads

Aurel Neic, David R. Booth, Shunmuga Sundaram Ponnusamy, Guram Imnadze, W Li, Jacqueline E. Joza, Haixia Liu, Robert Schaller
article en
1 citations

Abstract

Abstract Background Recent studies have demonstrated that implantable cardioverter defibrillator (ICD) leads placed higher and deeper into the right ventricular (RV) septum, compared to traditional RV apical placement, were able to achieve conduction system pacing (CSP). A higher and deeper septal position may impact defibrillation thresholds (DFT) compared to the traditional ICD lead implant position. Purpose To compare DFTs associated with ICD leads implanted within the deep septum (CSP group) vs. an apical position (traditional). Methods In silico torso/heart models were generated from 5 human subjects (Figure). A commercially available ICD lead and a prototype 7Fr lead with a 5cm defibrillation coil were modeled in software, for apical and CSP positions, respectively, with a perpendicular insertion angle and nominal lead slack in the RV. For each torso model, a stable lead pathway was generated following simulated dynamic cardiac motion, for multiple positions: apical (control), deep high (1.5cm from tricuspid valve), deep mid (≈3.5cm from tricuspid valve), and deep low (midpoint of mid and apical) septal. The shock vector was between RV coil and ICD can placed at the 2nd rib. DFT energy was recorded at >5V/cm in 95% of ventricular tissue. Results are reported as median (range). Results The 5 torso models demonstrated a median left ventricular volume of 200cm³ (163 to 266cm³), 66Ω (56 to 77Ω) shock vector impedance, and 12cm (10 to 13cm) coil-to-can distance. For the apical position, the DFT was 18.4J (6.9 to 28.9J, n=5). As compared to apical, the CSP-positioned DFTs overall increased 1.4J (-0.6 to 4.4J, n=5), across all subjects. DFTs for deep high and mid septal nominally increased by 1.1J (-2.4 to 4.4J, n=5) and 1.8J (0.2 to 2.2J, n=3), respectively. The DFT for deep low septal slightly decreased by 0.3J (0.1 to 0.6J, n=2). All DFTs were substantially below defibrillation energies current ICDs offer. Conclusion Based on human torso models, ICD leads implanted in CSP positions resulted in a slight increase in DFTs compared to RV apical placement. With the maximum energies current ICDs offer, the slight DFT increase is not expected to substantially affect defibrillation efficacy. Larger clinical studies are needed to replicate these findings.In silico defibrillation modeling

EP EuropaceVol. 27(Supplement_1)
Danaher (United States) (US), Abbott (Sweden) (SE), Hospital of the University of Pennsylvania (US), Heart and Diabetes Center North Rhine-Westphalia (DE), McGill University Health Centre (CA), Velammal Medical College Hospital and Research Institute (IN), Numerica Corporation (United States) (US)
Openalex Percentile: Top 19%
Cardiac pacing and defibrillation studies
1.39
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.