Abstract Background While strong evidence supports omitting defibrillation testing (DT) in transvenous implantable cardioverter defibrillator (ICD), the same cannot be assumed for subcutaneous ICD (S-ICD), where DT is still recommended due to limited evidence. Purpose This study aims to compare the long-term outcomes of S-ICD recipients with and without DT, focusing on cause-specific mortality, especially sudden cardiac death (SCD), and the incidence of shocks. Methods We analyzed data from the HONEST cohort, which includes 4,924 patients who received an S-ICD between 2012 and 2019 across150 French centers. Clinical follow-up and remote monitoring were used to track outcomes, including overall death, cardiovascular (CV) death, sudden cardiac death (SCD), a composite of SCD and deaths from unknown causes, as well as the incidence of appropriate and inappropriate shocks. Vital status was confirmed via the National Institute of Statistics and Economic Studies database, which tracks all deaths in France. Causes of death were adjudicated and classified following ESC guidelines. Propensity score weighting was used to adjust for confounding variables in comparisons between patients with DT (S-ICD Testing (+)) and those without DT (S-ICD Testing (-)). database Results In the HONEST cohort, 82.6% of the 4,924 patients underwent DT at S-ICD implantation, with a high success rate of 97.3%. Patients without DT were generally older (51.2 vs. 49.6 years, SMD=0.107), had a lower LVEF (37.6 vs. 43.3%, SMD=0.343), and were more frequently receiving S-ICD for primary prevention (68.0 vs. 62.4%, SMD=0.118). Crude analysis suggested lower overall mortality with DT (10.6 vs. 13.6%, HR=0.68, 95% CI 0.55–0.84, p<0.001), though this association did not remain significant after propensity score weighting (HR=0.89, 95% CI 0.73–1.10, p=0.285). No significant differences were observed in cardiovascular mortality (5.34 vs. 8.15%, HR=0.80, 95% CI 0.61–1.05, p=0.112), SCD (0.12 vs. 0.59%, HR=0.82, 95% CI 0.44–1.50, p= .978), or the composite of SCD and unknown causes (2.56 vs. 2.71%, HR=0.74, 95% CI 0.54–1.02, p=0.066). Rates of appropriate (12.10 vs. 12.34%, HR=0.92, 95% CI 0.74–1.14, p=0.442) and inappropriate shocks (12.20 vs. 10.10%, HR=1.07, 95% CI 0.85–1.35, p=0.572) were also similar. Conclusions Routine DFT does not significantly impact long-term outcomes, including overall mortality, cardiovascular death, SCD, composite outcomes of SCD and unknown death, or the incidence of appropriate and inappropriate shocks in the general population of S-ICD recipients.

Authors

Institutions

Publication Details

Journal
EP Europace
Published
2025-05-01
DOI
https://doi.org/10.1093/europace/euaf085.089
Primary Topic
Cardiac pacing and defibrillation studies
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Defibrillation testing at implant of subcutaneous implantable cardioverter defibrillator: weighting propensity score analysis and results from nationwide HONEST cohort

Éloi Marijon, R Eschalier, Vincent Probst, Nathalie Béhar et al.
EP Europace
Cardiac pacing and defibrillation studies
article

Defibrillation testing at implant of subcutaneous implantable cardioverter defibrillator: weighting propensity score analysis and results from nationwide HONEST cohort

Éloi Marijon, R Eschalier, Vincent Probst, Nathalie Béhar, Jacques Mansourati, Pierre Mondoly, Peggy Jacon, Serge Bovéda, Jérôme Hourdain, Nicolas Badenco, C Marquie, Vincent Mansourati, Fawzi Kerkouri, Laurence Jesel, R. Ortiz García
article en

Abstract

Abstract Background While strong evidence supports omitting defibrillation testing (DT) in transvenous implantable cardioverter defibrillator (ICD), the same cannot be assumed for subcutaneous ICD (S-ICD), where DT is still recommended due to limited evidence. Purpose This study aims to compare the long-term outcomes of S-ICD recipients with and without DT, focusing on cause-specific mortality, especially sudden cardiac death (SCD), and the incidence of shocks. Methods We analyzed data from the HONEST cohort, which includes 4,924 patients who received an S-ICD between 2012 and 2019 across150 French centers. Clinical follow-up and remote monitoring were used to track outcomes, including overall death, cardiovascular (CV) death, sudden cardiac death (SCD), a composite of SCD and deaths from unknown causes, as well as the incidence of appropriate and inappropriate shocks. Vital status was confirmed via the National Institute of Statistics and Economic Studies database, which tracks all deaths in France. Causes of death were adjudicated and classified following ESC guidelines. Propensity score weighting was used to adjust for confounding variables in comparisons between patients with DT (S-ICD Testing (+)) and those without DT (S-ICD Testing (-)). database Results In the HONEST cohort, 82.6% of the 4,924 patients underwent DT at S-ICD implantation, with a high success rate of 97.3%. Patients without DT were generally older (51.2 vs. 49.6 years, SMD=0.107), had a lower LVEF (37.6 vs. 43.3%, SMD=0.343), and were more frequently receiving S-ICD for primary prevention (68.0 vs. 62.4%, SMD=0.118). Crude analysis suggested lower overall mortality with DT (10.6 vs. 13.6%, HR=0.68, 95% CI 0.55–0.84, p<0.001), though this association did not remain significant after propensity score weighting (HR=0.89, 95% CI 0.73–1.10, p=0.285). No significant differences were observed in cardiovascular mortality (5.34 vs. 8.15%, HR=0.80, 95% CI 0.61–1.05, p=0.112), SCD (0.12 vs. 0.59%, HR=0.82, 95% CI 0.44–1.50, p= .978), or the composite of SCD and unknown causes (2.56 vs. 2.71%, HR=0.74, 95% CI 0.54–1.02, p=0.066). Rates of appropriate (12.10 vs. 12.34%, HR=0.92, 95% CI 0.74–1.14, p=0.442) and inappropriate shocks (12.20 vs. 10.10%, HR=1.07, 95% CI 0.85–1.35, p=0.572) were also similar. Conclusions Routine DFT does not significantly impact long-term outcomes, including overall mortality, cardiovascular death, SCD, composite outcomes of SCD and unknown death, or the incidence of appropriate and inappropriate shocks in the general population of S-ICD recipients.

EP EuropaceVol. 27(Supplement_1)
Centre Hospitalier Universitaire de Grenoble (FR), Centre Hospitalier Universitaire de Nantes (FR), Université de Poitiers (FR), Sorbonne Université (FR), Lille’s Cardiology Hospital (FR), Pitié-Salpêtrière Hospital (FR), Hôpital Pontchaillou (FR), Centre Hospitalier Universitaire de Clermont-Ferrand (FR), Centre Hospitalier Régional Universitaire de Brest (FR), Clinique Pasteur (FR), European Hospital (IT), Hôpital de la Timone (FR), Université de Strasbourg (FR), Université Grenoble Alpes (FR)
Good health and well-being
Openalex Percentile: Top 79%
Cardiac pacing and defibrillation studies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.