Conduction system pacing: how to reduce procedural failures utilizing different sheaths

Abstract Background Conduction system pacing (CSP) ensures physiological ventricular activation and avoids the dyssynchrony associated with traditional myocardial pacing. New tools (sheaths and leads) can now better assist the implant thus increasing procedural successes. Objective Retrospective analysis of CSP implants from January 2019 to April 2024, analyzing procedural success and failures based on tools utilized. Methods From January 2019 to April 2024, 1033 CSP implant were performed, being 72% of the overall implants (excluding single and dual chamber ICDs). 89% were PMs and 11% were ICDs. 31.6% were styled driven leads while 68.4% were lumenless leads. Pacing indications were standard with prevalence of AV block (50%); AF with slow ventricular rate 12%; sinus node disease 12%; ablate&pace 9%; heart failure or cardiomyopathy 14%, re-interventions after malfunction of a previous implant 1%. Mean age was 78±8 years; 62% were males. Results CSP was successfully achieved in 95.2% of cases considering all indications. In 4.8% of cases, CSP lead implant failed and the lead was then implanted in the RV septum or coronary sinus. Reasons for failure were: 2.7% due to suboptimal electrical parameters (mainly high threshold); 1.1% uncorrected BBB; 0.9% intraprocedural lead dislodgement; 1.4% impossibility to penetrate the septum; 1.4 % severe right atrial enlargement. Heart failure and left ventricular dysfunction (EF<45%) were found more frequently in the failed implant subgroup. Procedural failures were more frequently associated with intraprocedural need for changing the sheath (utilizing a different curve) or changing the lead. Interestingly, in 13.4 % of cases the intraprocedural sheath’s change allowed to successfully complete the procedure; and in the 5.4% of cases the intraprocedural lead’s change (i.e. shifting from lumenless to styled-driven or viceversa) turned a failure into a success. No predictors of procedural failure were found in the logistic regression analysis. Failed implants have not decreased over the years, likely being related to increasingly complex cardiac conditions where CSP has been attempted. Conclusion conduction system pacing is feasible in daily clinical practice. Both lumenless and stylet-driven lead can be utilized. Different sheaths allow to customize the procedure based on the anatomy. During a 5-year experience, the overall procedural success rate was 95.2% while 4.8% of the procedures failed. Intraprocedural sheath’s change turned 13.4% of initial failure into successes while intraprocedural lead’s change turned 5.4% of initial failure into successes. Interestingly the failure rate has not decreased over the years.

Authors

Institutions

Publication Details

Journal
EP Europace
Published
2025-05-01
DOI
https://doi.org/10.1093/europace/euaf085.556
Primary Topic
Cardiac pacing and defibrillation studies
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Conduction system pacing: how to reduce procedural failures utilizing different sheaths

G. Pastore, Lina Marcantoni, S Valenza, Francesco Zanon et al.
4 links
EP Europace
Cardiac pacing and defibrillation studies
article

Conduction system pacing: how to reduce procedural failures utilizing different sheaths

G. Pastore, Lina Marcantoni, S Valenza, Francesco Zanon, Frank G. DeLuca, G Porcelli
article en

Abstract

Abstract Background Conduction system pacing (CSP) ensures physiological ventricular activation and avoids the dyssynchrony associated with traditional myocardial pacing. New tools (sheaths and leads) can now better assist the implant thus increasing procedural successes. Objective Retrospective analysis of CSP implants from January 2019 to April 2024, analyzing procedural success and failures based on tools utilized. Methods From January 2019 to April 2024, 1033 CSP implant were performed, being 72% of the overall implants (excluding single and dual chamber ICDs). 89% were PMs and 11% were ICDs. 31.6% were styled driven leads while 68.4% were lumenless leads. Pacing indications were standard with prevalence of AV block (50%); AF with slow ventricular rate 12%; sinus node disease 12%; ablate&pace 9%; heart failure or cardiomyopathy 14%, re-interventions after malfunction of a previous implant 1%. Mean age was 78±8 years; 62% were males. Results CSP was successfully achieved in 95.2% of cases considering all indications. In 4.8% of cases, CSP lead implant failed and the lead was then implanted in the RV septum or coronary sinus. Reasons for failure were: 2.7% due to suboptimal electrical parameters (mainly high threshold); 1.1% uncorrected BBB; 0.9% intraprocedural lead dislodgement; 1.4% impossibility to penetrate the septum; 1.4 % severe right atrial enlargement. Heart failure and left ventricular dysfunction (EF<45%) were found more frequently in the failed implant subgroup. Procedural failures were more frequently associated with intraprocedural need for changing the sheath (utilizing a different curve) or changing the lead. Interestingly, in 13.4 % of cases the intraprocedural sheath’s change allowed to successfully complete the procedure; and in the 5.4% of cases the intraprocedural lead’s change (i.e. shifting from lumenless to styled-driven or viceversa) turned a failure into a success. No predictors of procedural failure were found in the logistic regression analysis. Failed implants have not decreased over the years, likely being related to increasingly complex cardiac conditions where CSP has been attempted. Conclusion conduction system pacing is feasible in daily clinical practice. Both lumenless and stylet-driven lead can be utilized. Different sheaths allow to customize the procedure based on the anatomy. During a 5-year experience, the overall procedural success rate was 95.2% while 4.8% of the procedures failed. Intraprocedural sheath’s change turned 13.4% of initial failure into successes while intraprocedural lead’s change turned 5.4% of initial failure into successes. Interestingly the failure rate has not decreased over the years.

EP EuropaceVol. 27(Supplement_1)
Ospedale Santa Maria della Misericordia di Rovigo (IT)
Openalex Percentile: Top 79%
Cardiac pacing and defibrillation studies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.