Hexagonal multi-electrode design for recording local electrograms with superior characteristics

Abstract Background The morphology and amplitude of bipolar electrograms depend on the direction of the activation front, and the local activation cannot be attributed to a single electrode location. Coaxial electrograms that use an evenly distributed local reference should address these shortcomings but require more electrodes. Purpose To determine the feasibility and efficiency of recoding coaxial electrograms using a triangular local reference that can scale up to hexagonal multi-electrode designs. Methods We performed in silico and experimental studies using a custom multi-electrode to validate the activation moment, and to directly compare coaxial electrograms recorded using square and triangular local references. Finally, we compared the efficiency of multi-electrode configurations for the recording of coaxial electrograms that may form the basic design of future catheter designs for clinical use. Results The computed coaxial electrograms retained morphological features of unipolar electrograms and the activation moment determined by –dV/dtmax were almost identical. The triangular and cross shaped local reference of the custom multi-electrode matched closely (r = 0.9999, p<0.0001). In contrast to bipolar electrograms, the morphology of the coaxial electrograms was insensitive to electrode orientation relative to the activation front. The local activation time in unipolar electrograms differed less from coaxial than bipolar electrograms (0.56±0.13ms vs 1.31±0.31ms, p<0.001). The recording of coaxial electrograms was more efficient using hexagonal rather than a square multi-electrode design requiring a minimum of 38 electrodes vs 49 to record 25 coaxial electrograms. Conclusions Coaxial electrograms recorded using a triangular references have superior characteristics over bipolar electrograms. The use of efficient hexagonal multi-electrode designs for their recording may improve the precision of activation mapping.Custom Multi-Electrode Comparison of Activation Times

Authors

Institutions

Publication Details

Journal
EP Europace
Published
2025-05-01
DOI
https://doi.org/10.1093/europace/euaf085.062
Primary Topic
Analytical Chemistry and Sensors
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Hexagonal multi-electrode design for recording local electrograms with superior characteristics

Bas J. Boukens, Marion Kuiper, Rudolf A. de Boer, Natasja M.S. de Groot et al.
3 links
EP Europace
Analytical Chemistry and Sensors
article

Hexagonal multi-electrode design for recording local electrograms with superior characteristics

Bas J. Boukens, Marion Kuiper, Rudolf A. de Boer, Natasja M.S. de Groot, Ruben Coronel, Rishi Arora, Edward J. Vigmond, Mark G. Hoogendijk
article en

Abstract

Abstract Background The morphology and amplitude of bipolar electrograms depend on the direction of the activation front, and the local activation cannot be attributed to a single electrode location. Coaxial electrograms that use an evenly distributed local reference should address these shortcomings but require more electrodes. Purpose To determine the feasibility and efficiency of recoding coaxial electrograms using a triangular local reference that can scale up to hexagonal multi-electrode designs. Methods We performed in silico and experimental studies using a custom multi-electrode to validate the activation moment, and to directly compare coaxial electrograms recorded using square and triangular local references. Finally, we compared the efficiency of multi-electrode configurations for the recording of coaxial electrograms that may form the basic design of future catheter designs for clinical use. Results The computed coaxial electrograms retained morphological features of unipolar electrograms and the activation moment determined by –dV/dtmax were almost identical. The triangular and cross shaped local reference of the custom multi-electrode matched closely (r = 0.9999, p<0.0001). In contrast to bipolar electrograms, the morphology of the coaxial electrograms was insensitive to electrode orientation relative to the activation front. The local activation time in unipolar electrograms differed less from coaxial than bipolar electrograms (0.56±0.13ms vs 1.31±0.31ms, p<0.001). The recording of coaxial electrograms was more efficient using hexagonal rather than a square multi-electrode design requiring a minimum of 38 electrodes vs 49 to record 25 coaxial electrograms. Conclusions Coaxial electrograms recorded using a triangular references have superior characteristics over bipolar electrograms. The use of efficient hexagonal multi-electrode designs for their recording may improve the precision of activation mapping.Custom Multi-Electrode Comparison of Activation Times

EP EuropaceVol. 27(Supplement_1)
Maastricht University Medical Centre (NL), Electrophysiology and Heart Modeling Institute (FR), University of Chicago Medical Center (US), Amsterdam University Medical Centers (NL), Erasmus University Rotterdam (NL)
Openalex Percentile: Top 85%
Analytical Chemistry and Sensors
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.