Impact of a new balloon-in-basket pulsed field ablation system on esophageal temperature changes during pulmonary vein isolation

Abstract Background Pulsed field ablation (PFA) for the treatment of atrial fibrillation (AF) offers improved safety compared with thermal ablation. Due to its myocardial tissue selectivity collateral tissue injury, especially esophageal thermal injury, potentially resulting in atrio-esophageal fistula have not been described for PFA. However, several factors such as energy utilized, catheter design and tissue proximity may play a role. Objective To evaluate potential esophageal temperature changes during pulmonary vein isolation (PVI) using a new balloon-in-basket, 3D integrated PFA system. Methods and results Thirty consecutive patients (median age 65 years; 60% male, 50% paroxysmal AF) out of 150 patients enrolled in the VOLT CE Mark Study (NCT06106594), a pre-market prospective, multi-center, single-arm study underwent de-novo AF ablation for symptomatic paroxysmal or persistent AF using a new balloon-in-basket PFA system at a tertiary care center. According to the protocol no additional ablations beyond PVI were allowed. Baseline therapy consisted of two nominal voltage applications with a maximum of eight applications per PV. A pre- and post-ablation voltage map was generated using a multipolar mapping catheter (HD grid) and a 3D-mapping system (Ensite X EP system). In this cohort, all procedures were performed under conscious sedation using Midazolam, Propofol and Fentanyl. Intraluminal esophageal temperature (TESO) was monitored continuously with an S-shaped multi-electrode esophageal temperature probe. A mean of 16.5 ±2.0 PFA applications per subject and 4.2 ± 1.1 PFA applications per PV were delivered. Procedure, fluoroscopy, and LA dwell times were 78.2 ± 10.6 min, 10.0 ± 4.5 min and 27.1 ± 8.5 min, respectively. Mean TESO change was statistically significant and increased by 0.2 ± 0.1 °C, but was not clinically relevant (Table 1). No TESO increase > 1°C was observed. The highest TESO measured was 37.8°C and the largest TESO difference (DTESO) was 1.0°C. All patients remained asymptomatic considering possible esophageal thermal injury (ETI). No esophago-duodenoscopy was necessary. Conclusion PFA delivered via a new balloon-in-basket, 3D integrated PFA system demonstrated a clinically irrelevant TESO change. Thus, these results underline the potential of the PFA technology to possibly eliminate the risk of thermal damage to the esophagus. However, more studies with larger sample sizes are needed to draw robust conclusions and confirm these findings. Table 1: Mean intraluminal esophageal temperature changes during pulmonary vein isolation using a new balloon-in-basket PFA system (n=30).Table 1

Authors

Institutions

Publication Details

Journal
EP Europace
Published
2025-05-01
DOI
https://doi.org/10.1093/europace/euaf085.417
Primary Topic
Cardiac Arrhythmias and Treatments
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Impact of a new balloon-in-basket pulsed field ablation system on esophageal temperature changes during pulmonary vein isolation

Roland Richard Tilz, Julius Nikorowitsch, J P Wenzel, Karl‐Heinz Kück et al.
EP Europace
Cardiac Arrhythmias and Treatments
article

Impact of a new balloon-in-basket pulsed field ablation system on esophageal temperature changes during pulmonary vein isolation

Roland Richard Tilz, Julius Nikorowitsch, J P Wenzel, Karl‐Heinz Kück, Bill Subin, Emel Yaman, Charlotte Eitel, S Hatahet, Julia Vogler, Roman Mamaev, S S Popescu, Z G Demirtakan, S Reincke, M Kuechler, Tugba A. Oezalp
article en

Abstract

Abstract Background Pulsed field ablation (PFA) for the treatment of atrial fibrillation (AF) offers improved safety compared with thermal ablation. Due to its myocardial tissue selectivity collateral tissue injury, especially esophageal thermal injury, potentially resulting in atrio-esophageal fistula have not been described for PFA. However, several factors such as energy utilized, catheter design and tissue proximity may play a role. Objective To evaluate potential esophageal temperature changes during pulmonary vein isolation (PVI) using a new balloon-in-basket, 3D integrated PFA system. Methods and results Thirty consecutive patients (median age 65 years; 60% male, 50% paroxysmal AF) out of 150 patients enrolled in the VOLT CE Mark Study (NCT06106594), a pre-market prospective, multi-center, single-arm study underwent de-novo AF ablation for symptomatic paroxysmal or persistent AF using a new balloon-in-basket PFA system at a tertiary care center. According to the protocol no additional ablations beyond PVI were allowed. Baseline therapy consisted of two nominal voltage applications with a maximum of eight applications per PV. A pre- and post-ablation voltage map was generated using a multipolar mapping catheter (HD grid) and a 3D-mapping system (Ensite X EP system). In this cohort, all procedures were performed under conscious sedation using Midazolam, Propofol and Fentanyl. Intraluminal esophageal temperature (TESO) was monitored continuously with an S-shaped multi-electrode esophageal temperature probe. A mean of 16.5 ±2.0 PFA applications per subject and 4.2 ± 1.1 PFA applications per PV were delivered. Procedure, fluoroscopy, and LA dwell times were 78.2 ± 10.6 min, 10.0 ± 4.5 min and 27.1 ± 8.5 min, respectively. Mean TESO change was statistically significant and increased by 0.2 ± 0.1 °C, but was not clinically relevant (Table 1). No TESO increase > 1°C was observed. The highest TESO measured was 37.8°C and the largest TESO difference (DTESO) was 1.0°C. All patients remained asymptomatic considering possible esophageal thermal injury (ETI). No esophago-duodenoscopy was necessary. Conclusion PFA delivered via a new balloon-in-basket, 3D integrated PFA system demonstrated a clinically irrelevant TESO change. Thus, these results underline the potential of the PFA technology to possibly eliminate the risk of thermal damage to the esophagus. However, more studies with larger sample sizes are needed to draw robust conclusions and confirm these findings. Table 1: Mean intraluminal esophageal temperature changes during pulmonary vein isolation using a new balloon-in-basket PFA system (n=30).Table 1

EP EuropaceVol. 27(Supplement_1)
University of Lübeck (DE)
Openalex Percentile: Top 79%
Cardiac Arrhythmias and Treatments
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.