Extraction of novel, small diameter OmniaSecure defibrillation lead from the deep septum in sheep

Abstract Introduction: The OmniaSecure lead is a novel, small-diameter (4.7Fr), lumenless, catheter-delivered defibrillation lead based on the established SelectSecure SureScan MRI Model 3830 pacing lead. The ongoing LEADR and LEADR LBBAP clinical studies will assess the safety and efficacy of implanting OmniaSecure in the standard RV and deep septal locations, respectively. Purpose The aim of this study is to preclinically evaluate the extractability of the OmniaSecure lead. Methods Forty sheep were implanted with 80 leads (2 leads/sheep) for a 5 year extraction study, with 16 leads extracted yearly. Thus far, 32 leads from 16 sheep were extracted at year 1 and 2 using simple traction or market available extraction tools. Among these were 25 OmniaSecure leads, 14 were implanted in a standard RV location and 11 in deep septal locations. As controls, 7 Sprint Quattro leads (6935M) were implanted in a standard RV location. The force required to remove the lead was recorded. Pre- and post-extraction ECG were visually compared for morphology changes. Scar burden from coil to tip was assessed on the extracted lead and assigned a value (1-low, 5-high burden). Animals were monitored for one hour post-extraction. Images of the suspected lead implant location were taken at necropsy and evaluated for tissue damage. Extraction was deemed successful if the lead was fully extracted and there were no complications. Results Sixteen leads were extracted from 8 sheep each year (Table 1). Extraction of all OmniaSecure and Sprint Quattro leads was successful at 1 and 2 years. At year 1, extraction was performed for all leads with traction alone (7 OmniaSecure leads from the RV location, 6 from the deep septum, and 3 Sprint Quattro leads from the RV location). At year 2, extraction required the use of a Cook Evolution tool in 2/4 Sprint Quattro leads and 1/8 OmniaSecure leads in the standard RV location. All other leads were removed with simple traction, including all 4 OmniaSecure leads from the deep septum. Minimal force was required to remove OmniaSecure and Sprint Quattro leads at Year 1 (Table 1). At year 2, extractions trended towards less force for OmniaSecure compared to Sprint Quattro leads. When visually comparing ECGs, a persistent change in the T wave direction was observed in 3 of the 8 animals at year 2, all three of which were extracted from the deep septum. Sprint Quattro had a higher scar burden than OmniaSecure (4.3 vs 2.8) (Figure 1). Upon visual inspection of the implant locations in the dissected RV, leads extracted from the deep septum created more localized tissue damage and bruising than leads extracted from the standard RV (Figure 1); however, all leads from the deep septum were extracted successfully. Conclusion These data demonstrate the extractability of OmniaSecure leads through 2 years in large animal models. OmniaSecure leads may offer advantages in clinical lead extraction, including lower lead removal forces and reduced scar burden.

Authors

Institutions

Publication Details

Journal
EP Europace
Published
2025-05-01
DOI
https://doi.org/10.1093/europace/euaf085.651
Citations
1
Primary Topic
Muscle activation and electromyography studies
Type
article
Field-Weighted Citation Impact
0.67
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Extraction of novel, small diameter OmniaSecure defibrillation lead from the deep septum in sheep

Héctor Mazzetti, Carlos Perona, Matthew M. Marshall, M Upchurch
1 citations
EP Europace
Muscle activation and electromyography studies
0.67
article

Extraction of novel, small diameter OmniaSecure defibrillation lead from the deep septum in sheep

Héctor Mazzetti, Carlos Perona, Matthew M. Marshall, M Upchurch
article en
1 citations

Abstract

Abstract Introduction: The OmniaSecure lead is a novel, small-diameter (4.7Fr), lumenless, catheter-delivered defibrillation lead based on the established SelectSecure SureScan MRI Model 3830 pacing lead. The ongoing LEADR and LEADR LBBAP clinical studies will assess the safety and efficacy of implanting OmniaSecure in the standard RV and deep septal locations, respectively. Purpose The aim of this study is to preclinically evaluate the extractability of the OmniaSecure lead. Methods Forty sheep were implanted with 80 leads (2 leads/sheep) for a 5 year extraction study, with 16 leads extracted yearly. Thus far, 32 leads from 16 sheep were extracted at year 1 and 2 using simple traction or market available extraction tools. Among these were 25 OmniaSecure leads, 14 were implanted in a standard RV location and 11 in deep septal locations. As controls, 7 Sprint Quattro leads (6935M) were implanted in a standard RV location. The force required to remove the lead was recorded. Pre- and post-extraction ECG were visually compared for morphology changes. Scar burden from coil to tip was assessed on the extracted lead and assigned a value (1-low, 5-high burden). Animals were monitored for one hour post-extraction. Images of the suspected lead implant location were taken at necropsy and evaluated for tissue damage. Extraction was deemed successful if the lead was fully extracted and there were no complications. Results Sixteen leads were extracted from 8 sheep each year (Table 1). Extraction of all OmniaSecure and Sprint Quattro leads was successful at 1 and 2 years. At year 1, extraction was performed for all leads with traction alone (7 OmniaSecure leads from the RV location, 6 from the deep septum, and 3 Sprint Quattro leads from the RV location). At year 2, extraction required the use of a Cook Evolution tool in 2/4 Sprint Quattro leads and 1/8 OmniaSecure leads in the standard RV location. All other leads were removed with simple traction, including all 4 OmniaSecure leads from the deep septum. Minimal force was required to remove OmniaSecure and Sprint Quattro leads at Year 1 (Table 1). At year 2, extractions trended towards less force for OmniaSecure compared to Sprint Quattro leads. When visually comparing ECGs, a persistent change in the T wave direction was observed in 3 of the 8 animals at year 2, all three of which were extracted from the deep septum. Sprint Quattro had a higher scar burden than OmniaSecure (4.3 vs 2.8) (Figure 1). Upon visual inspection of the implant locations in the dissected RV, leads extracted from the deep septum created more localized tissue damage and bruising than leads extracted from the standard RV (Figure 1); however, all leads from the deep septum were extracted successfully. Conclusion These data demonstrate the extractability of OmniaSecure leads through 2 years in large animal models. OmniaSecure leads may offer advantages in clinical lead extraction, including lower lead removal forces and reduced scar burden.

EP EuropaceVol. 27(Supplement_1)
Medtronic (Ireland) (IE), Hospital Donación Francisco Santojanni (AR), Hospital de Niños de la Santísima Trinidad (AR)
Openalex Percentile: Top 31%
Muscle activation and electromyography studies
0.67
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.