Fluoroless catheter ablation of leftsided tachycardias in patients with cardiac implantable electronic devices: a single-centre retrospective study

Abstract Introduction The use of intracardiac echocardiography (ICE) and 3-D electroanatomical mapping (3-D EAM) systems allows safe fluoroless catheter ablations (CAs) in nearly all types of arrhythmias. Nevertheless, for patients with cardiac implantable electronic devices (CIEDs) there exists a notable paucity of data concerning the rate of leads’ dislocation in fluoroless procedures. Purpose The aim of this single-centre retrospective study was to evaluate the rate of leads’ dislocation in fluoroless CA of arrhythmias necessitating left sided approach in patients with CIEDs. Methods We performed a retrospective study in patients with CIED who underwent fluoroless CA with transseptal puncture (TSP) for any arrhythmia between February 2016 and November 2023. All procedures were performed without fluoroscopy, guided by a 3-D EAM system and ICE. The leads were repeatedly visualized with ICE during introduction of catheters, their manipulation, TSP and ablation. The device´s parameters were tested before and after the procedure. Results Our study included 88 patients. A total of 105 procedures (90 [86.5 %] in males, average age at procedure 65.6 ± 10.6 years) was performed for 119 arrhythmias; there were 41 (39.4 %) CAs of atrial fibrillation or atypical flutter and 13 (12.5%) concomitant ablations of typical flutter, 1 (1.0 %) ablation of focal atrial tachycardia, 59 (56.7 %) of ventricular tachycardia and 5 (4.8 %) of ventricular ectopy. 60 patients (56.7 %) had single-chamber intracardiac defibrillator (ICD VR), 11 patients (10.6 %) had dual-chamber intracardiac defibrillator (ICD DR), 1 patient (1.0 %) had permanent single-chamber pacemaker, 15 patients (14.4 %) had permanent dual-chamber pacemaker, and 11 patients (10.6 %) had cardiac resynchronization devices (CRT-D or CRT-P). For 7 patients (6.7 %) the retrospective data about device’s type is missing. In two procedures (2/106, 1.9 %) we observed ventricular lead dislocation while there were no atrial lead dislocations. One ventricular lead dislocation happened in CA of atypical flutter in a patient with ICD DR during manoeuvring of catheters in the right heart chambers. The other ventricular lead dislocated in a patient with ICD VR who experienced coronary air embolism during CA of ischemic ventricular tachycardia necessitating cardiopulmonary resuscitation with chest compressions. According to ICE imaging the lead dislocation happened after the successful resuscitation, therefore it was not related to fluoroless ablation method. Both lead dislocations required subsequent lead repositioning. There were no lead dislocations related to the TSP and the process of obtaining transseptal approach. In other patients, the device interrogation after completion of procedures showed no change in parameters. Conclusion Lead dislocations in patients with CIEDs undergoing leftsided fluoroless CAs are rare and appear unrelated to TSP.

Authors

Institutions

Publication Details

Journal
EP Europace
Published
2025-05-01
DOI
https://doi.org/10.1093/europace/euaf085.135
Primary Topic
Cardiac Arrhythmias and Treatments
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Fluoroless catheter ablation of leftsided tachycardias in patients with cardiac implantable electronic devices: a single-centre retrospective study

Andreja Marn Pernat, Tine Prolič Kalinšek, Bor Antolič, Luka Klemen et al.
EP Europace
Cardiac Arrhythmias and Treatments
article

Fluoroless catheter ablation of leftsided tachycardias in patients with cardiac implantable electronic devices: a single-centre retrospective study

Andreja Marn Pernat, Tine Prolič Kalinšek, Bor Antolič, Luka Klemen, Martin Rauber, V Podlogar, M Jan
article en

Abstract

Abstract Introduction The use of intracardiac echocardiography (ICE) and 3-D electroanatomical mapping (3-D EAM) systems allows safe fluoroless catheter ablations (CAs) in nearly all types of arrhythmias. Nevertheless, for patients with cardiac implantable electronic devices (CIEDs) there exists a notable paucity of data concerning the rate of leads’ dislocation in fluoroless procedures. Purpose The aim of this single-centre retrospective study was to evaluate the rate of leads’ dislocation in fluoroless CA of arrhythmias necessitating left sided approach in patients with CIEDs. Methods We performed a retrospective study in patients with CIED who underwent fluoroless CA with transseptal puncture (TSP) for any arrhythmia between February 2016 and November 2023. All procedures were performed without fluoroscopy, guided by a 3-D EAM system and ICE. The leads were repeatedly visualized with ICE during introduction of catheters, their manipulation, TSP and ablation. The device´s parameters were tested before and after the procedure. Results Our study included 88 patients. A total of 105 procedures (90 [86.5 %] in males, average age at procedure 65.6 ± 10.6 years) was performed for 119 arrhythmias; there were 41 (39.4 %) CAs of atrial fibrillation or atypical flutter and 13 (12.5%) concomitant ablations of typical flutter, 1 (1.0 %) ablation of focal atrial tachycardia, 59 (56.7 %) of ventricular tachycardia and 5 (4.8 %) of ventricular ectopy. 60 patients (56.7 %) had single-chamber intracardiac defibrillator (ICD VR), 11 patients (10.6 %) had dual-chamber intracardiac defibrillator (ICD DR), 1 patient (1.0 %) had permanent single-chamber pacemaker, 15 patients (14.4 %) had permanent dual-chamber pacemaker, and 11 patients (10.6 %) had cardiac resynchronization devices (CRT-D or CRT-P). For 7 patients (6.7 %) the retrospective data about device’s type is missing. In two procedures (2/106, 1.9 %) we observed ventricular lead dislocation while there were no atrial lead dislocations. One ventricular lead dislocation happened in CA of atypical flutter in a patient with ICD DR during manoeuvring of catheters in the right heart chambers. The other ventricular lead dislocated in a patient with ICD VR who experienced coronary air embolism during CA of ischemic ventricular tachycardia necessitating cardiopulmonary resuscitation with chest compressions. According to ICE imaging the lead dislocation happened after the successful resuscitation, therefore it was not related to fluoroless ablation method. Both lead dislocations required subsequent lead repositioning. There were no lead dislocations related to the TSP and the process of obtaining transseptal approach. In other patients, the device interrogation after completion of procedures showed no change in parameters. Conclusion Lead dislocations in patients with CIEDs undergoing leftsided fluoroless CAs are rare and appear unrelated to TSP.

EP EuropaceVol. 27(Supplement_1)
Ljubljana University Medical Centre (SI)
Openalex Percentile: Top 79%
Cardiac Arrhythmias and Treatments
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.