Sustained electrical synchrony with left bundle branch area pacing: results at long-term follow-up

Abstract Background Left bundle branch area pacing (LBBAP) is an effective method to provide physiologic pacing. Several electrical parameters derived from the paced QRS complex are used to confirm LBB capture during implantation. Whether electrical synchrony persists at follow-up in patients undergoing left bundle branch area pacing (LBBAP) remains unknown. Purpose To determine the long-term performance of electrical synchrony in a cohort of patients undergoing LBBAP by comparing electrocardiographic parameters related to LBB capture at implant and follow-up Methods Consecutive patients with successful LBBAP for bradycardia pacing indication and preserved left ventricular ejection fraction were selected. At follow-up, a 12-lead ECG was recorded in unipolar pacing configuration. V6-R wave peak time (RWPT), V1-RWTP and QRS duration were measured and compared to measurements at implantation. Results A total of 149 patients completed the study protocol and were analyzed (mean age 77.3±9.9years, 52.3% male). Non-diseased LBB was present in 73.2% of patients. Average baseline QRS complex duration was of 116.6±28.2. Majority of implants were performed with lumenless fixed helix leads (94%). Confirmed LBB capture was achieved in 114 patients (76.5%) and LVSP in 35 patients. Mean QRS duration (from onset) after LBBAP was 115.7±12.9 ms. After 18.2 ± 7.3 months of follow-up visit, V6-RWPT of the LBBA captures was reduced from 77.8±11.8 ms at implant to 74.8±11.9 ms (p<0.001) (average reduction: 3.5±9.1 ms). V6-RWTP significantly decreased both among LBBP and LVSP captures (table 1). Average reduction in V6-RWPT did not differed among LBBP and LVSP captures (figure ). V1-RWPT was reduced from 118.7±12.1 ms at implant to 112.4±11.9 ms at follow-up among the entire LBBA captures (p<0.001) (mean reduction: 6.7±10.0 ms). At follow-up V1-RWPT significantly decreased among LBBP and LVSP captures (table 1). Mean decrease in V1-RWTP tended to be higher among LBBP compared to LVSP (figure 1). Among LBBP captures, V6-V1 interpeak interval significantly decreased, however, in LVS captures, V6-V1 interpeak interval significantly increased (table 1 and figure 1). Paced QRS complex duration (p-QRS) significantly decreased at follow-up, among the whole study population. P-QRS, measured from pacing spike (p-QRS-stim), was reduced from 146.4±13.9 ms at implant to 138.9±13.8 ms at follow-up (p<0.001); and p-QRS, measured from onset (p-QRS-onset) decreased from 115.7±12.9 ms to 109.3±13.3 ms at follow-up (p<0.001). Among LBBP captures, p-QRS significantly decreased, while among LVS capture patients, duration of the paced QRS complex remained stable at follow-up (table 1). Average decrease in p-QRS-stim and p-QRS-onset was higher among LBBP compared to LVS captures (figure 1). Conclusions Electrical synchrony in patients undergoing LBBAP for bradycardia pacing indications remained stable over time. LBBAP is a reliable and durable method for providing physiologic pacing.Table 1. Figure 1

Authors

Institutions

Publication Details

Journal
EP Europace
Published
2025-05-01
DOI
https://doi.org/10.1093/europace/euaf085.534
Primary Topic
EEG and Brain-Computer Interfaces
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Sustained electrical synchrony with left bundle branch area pacing: results at long-term follow-up

A Estevez Paniagua, Silvia Jiménez Loeches, S Briongos Figuero, D Heredero et al.
EP Europace
EEG and Brain-Computer Interfaces
article

Sustained electrical synchrony with left bundle branch area pacing: results at long-term follow-up

A Estevez Paniagua, Silvia Jiménez Loeches, S Briongos Figuero, D Heredero, E Sanchez Lopez, M Tapia Martinez, A Luna Cabadas, A Sanchez Hernandez
article en

Abstract

Abstract Background Left bundle branch area pacing (LBBAP) is an effective method to provide physiologic pacing. Several electrical parameters derived from the paced QRS complex are used to confirm LBB capture during implantation. Whether electrical synchrony persists at follow-up in patients undergoing left bundle branch area pacing (LBBAP) remains unknown. Purpose To determine the long-term performance of electrical synchrony in a cohort of patients undergoing LBBAP by comparing electrocardiographic parameters related to LBB capture at implant and follow-up Methods Consecutive patients with successful LBBAP for bradycardia pacing indication and preserved left ventricular ejection fraction were selected. At follow-up, a 12-lead ECG was recorded in unipolar pacing configuration. V6-R wave peak time (RWPT), V1-RWTP and QRS duration were measured and compared to measurements at implantation. Results A total of 149 patients completed the study protocol and were analyzed (mean age 77.3±9.9years, 52.3% male). Non-diseased LBB was present in 73.2% of patients. Average baseline QRS complex duration was of 116.6±28.2. Majority of implants were performed with lumenless fixed helix leads (94%). Confirmed LBB capture was achieved in 114 patients (76.5%) and LVSP in 35 patients. Mean QRS duration (from onset) after LBBAP was 115.7±12.9 ms. After 18.2 ± 7.3 months of follow-up visit, V6-RWPT of the LBBA captures was reduced from 77.8±11.8 ms at implant to 74.8±11.9 ms (p<0.001) (average reduction: 3.5±9.1 ms). V6-RWTP significantly decreased both among LBBP and LVSP captures (table 1). Average reduction in V6-RWPT did not differed among LBBP and LVSP captures (figure ). V1-RWPT was reduced from 118.7±12.1 ms at implant to 112.4±11.9 ms at follow-up among the entire LBBA captures (p<0.001) (mean reduction: 6.7±10.0 ms). At follow-up V1-RWPT significantly decreased among LBBP and LVSP captures (table 1). Mean decrease in V1-RWTP tended to be higher among LBBP compared to LVSP (figure 1). Among LBBP captures, V6-V1 interpeak interval significantly decreased, however, in LVS captures, V6-V1 interpeak interval significantly increased (table 1 and figure 1). Paced QRS complex duration (p-QRS) significantly decreased at follow-up, among the whole study population. P-QRS, measured from pacing spike (p-QRS-stim), was reduced from 146.4±13.9 ms at implant to 138.9±13.8 ms at follow-up (p<0.001); and p-QRS, measured from onset (p-QRS-onset) decreased from 115.7±12.9 ms to 109.3±13.3 ms at follow-up (p<0.001). Among LBBP captures, p-QRS significantly decreased, while among LVS capture patients, duration of the paced QRS complex remained stable at follow-up (table 1). Average decrease in p-QRS-stim and p-QRS-onset was higher among LBBP compared to LVS captures (figure 1). Conclusions Electrical synchrony in patients undergoing LBBAP for bradycardia pacing indications remained stable over time. LBBAP is a reliable and durable method for providing physiologic pacing.Table 1. Figure 1

EP EuropaceVol. 27(Supplement_1)
Hospital Central de la Defensa Gómez Ulla (ES), Hospital Universitario Infanta Leonor (ES)
Openalex Percentile: Top 89%
EEG and Brain-Computer Interfaces
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.