Safety and effectiveness of adapted definitive pacemakers to perform temporary cardiac stimulation - The TEMPACE multicenter study

Abstract Background Temporary Cardiac Pacing (TCP) is commonly used not only in urgent cases but also in elective procedures, providing essential support during interventional cardiology. Non-fixation systems are linked to a high rate of complications, whereas definitive pacemakers, used off-label as temporary devices, show better clinical outcomes. The lack of certified devices for TCP poses risks to patients. Purpose To evaluate the safety and efficacy of TCP using active fixation leads and a novel holding system. (Picture 1) Methods This was a multicenter clinical study of a medical device approved by the Clinical Research Ethics Committee and the Spanish Regulatory Agency. Patients requiring temporary cardiac pacing for more than 48 hours, from January 2023 to October 2024, were included. A 58-60 cm active fixation lead was implanted under ultrasound guidance via the left or right internal jugular vein, anchored in the right ventricle. In the end, all the parts of the pacemaker were held in place by a specifically designed holdyng system. Effectiveness measures, including electrode impedance, R-wave sensing, and pacing thresholds, were collected every other day. Complications like capture or sensing failure, electrode displacement or fracture, generator detachment, and cardiac perforation were monitored daily. Additional data included pacing hours in the Intensive Care / Acute Cardiology Unit and the general ward. Results A total of 29 patients from two centers were included: 9 with transcatheter aortic valve implantation, 9 with complete atrioventricular block, 5 with septal alcohol ablation, 2 with bradycardia-tachycardia syndrome, 2 with slow atrial fibrillation, 1 with device infection, and 1 with myocarditis. All R-wave amplitudes, impedance values, and pacing thresholds remained within normal limits (Picture 2), except for one case of electrode displacement due to patient agitation on day 25, with no clinical consequences. The mean pacing duration was 7.5 days, with 79.4% of pacing time spent in the ward without mobility restrictions. Conclusions Temporary cardiac pacing with active fixation leads is effective and safe, allowing patient management in general hospital wards. Agitation and disorientation may increase the risk of complications. A holding system that adapts definitive pacemakers provides the necessary safety for temporary pacingTemporary pacemaker holding system Electrical pacemaker parameters

Authors

Institutions

Publication Details

Journal
EP Europace
Published
2025-05-01
DOI
https://doi.org/10.1093/europace/euaf085.529
Primary Topic
Cardiac pacing and defibrillation studies
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Safety and effectiveness of adapted definitive pacemakers to perform temporary cardiac stimulation - The TEMPACE multicenter study

R Adelino Recasens, Vicente Bertomeu‐González, Manuel Ramírez, Laura Fuertes-Kenneally et al.
EP Europace
Cardiac pacing and defibrillation studies
article

Safety and effectiveness of adapted definitive pacemakers to perform temporary cardiac stimulation - The TEMPACE multicenter study

R Adelino Recasens, Vicente Bertomeu‐González, Manuel Ramírez, Laura Fuertes-Kenneally, Raquel M. Miralles, P Jordan Marchite, Alicia Ibáñez-Críado, Celia María Andreu Concha, J.C. Pascual, Juan Gabriel Martı́nez, Fernando Sánchez, M. Barranco, M. A. Galvany Ferrer
article en

Abstract

Abstract Background Temporary Cardiac Pacing (TCP) is commonly used not only in urgent cases but also in elective procedures, providing essential support during interventional cardiology. Non-fixation systems are linked to a high rate of complications, whereas definitive pacemakers, used off-label as temporary devices, show better clinical outcomes. The lack of certified devices for TCP poses risks to patients. Purpose To evaluate the safety and efficacy of TCP using active fixation leads and a novel holding system. (Picture 1) Methods This was a multicenter clinical study of a medical device approved by the Clinical Research Ethics Committee and the Spanish Regulatory Agency. Patients requiring temporary cardiac pacing for more than 48 hours, from January 2023 to October 2024, were included. A 58-60 cm active fixation lead was implanted under ultrasound guidance via the left or right internal jugular vein, anchored in the right ventricle. In the end, all the parts of the pacemaker were held in place by a specifically designed holdyng system. Effectiveness measures, including electrode impedance, R-wave sensing, and pacing thresholds, were collected every other day. Complications like capture or sensing failure, electrode displacement or fracture, generator detachment, and cardiac perforation were monitored daily. Additional data included pacing hours in the Intensive Care / Acute Cardiology Unit and the general ward. Results A total of 29 patients from two centers were included: 9 with transcatheter aortic valve implantation, 9 with complete atrioventricular block, 5 with septal alcohol ablation, 2 with bradycardia-tachycardia syndrome, 2 with slow atrial fibrillation, 1 with device infection, and 1 with myocarditis. All R-wave amplitudes, impedance values, and pacing thresholds remained within normal limits (Picture 2), except for one case of electrode displacement due to patient agitation on day 25, with no clinical consequences. The mean pacing duration was 7.5 days, with 79.4% of pacing time spent in the ward without mobility restrictions. Conclusions Temporary cardiac pacing with active fixation leads is effective and safe, allowing patient management in general hospital wards. Agitation and disorientation may increase the risk of complications. A holding system that adapts definitive pacemakers provides the necessary safety for temporary pacingTemporary pacemaker holding system Electrical pacemaker parameters

EP EuropaceVol. 27(Supplement_1)
Hebron University (PS), Hospital Clínica Benidorm (ES), Hospital General Universitario de Alicante Doctor Balmis (ES)
Sociedad Española de Cardiología, EIT Health
Openalex Percentile: Top 79%
Cardiac pacing and defibrillation studies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.