Post ablation mapping and inducibility of atrial tachycardia in different pulsed field ablation technologies

Abstract Introduction Pulsed field ablation (PFA) is a rapidly developing non-thermal ablation method leading to irreversible electroporation by application of short electrical pulses. Several PFA catheters with different designs, voltage output and energy delivery have been developed (1), therefore lesion durability and size may differ between the systems. This study was designed to compare lesion sizes and to evaluate if large posterior wall lesions leading to small interlesion distances are associated with roof dependent flutter inducibility and occurrence at follow up. Methods Consecutive patients who presented for atrial fibrillation ablation with PFA using three different ablation technologies were included. All patients underwent high density post ablation mapping. Peak-to-peak electrogram bipolar amplitude < 0.5 mV was defined as low-voltage area threshold (Varipulse and PulseSelect; 5); for Affera cases, unipolar low-voltage maps were created with a cutoff of 0.05 mV. Interlesion distance was measured a) at the left atrial roof b) at the posterior wall at the minimal distance between the PV lesions and at the c) lower margin of the PV ablation lesion. In a subset of patients, atrial stimulation until the atrial refractory period was performed after PV ablation. Follow up was scheduled at 3, 6 and 12 months. Results 69 patients were included. Baseline characteristics showed no significant differences between the three groups (see table 1). PV isolation was achieved in all cases. Interlesion distances differed significantly between PFA systems (see table 2). Burst stimulation in 37 patients did not induce sustained atrial tachycardia. After a mean follow-up of three months, one patient experienced sustained atrial tachycardia (perimitral flutter). Conclusion To our knowledge, this study is the first to comparatively assess interlesion distances between three current PFA ablation systems. The study shows considerable variability in lesion size and therefore interlesion distance. So far there is no sign for an increased number of roof dependent flutter episodes during follow up.

Authors

Publication Details

Journal
EP Europace
Published
2025-05-01
DOI
https://doi.org/10.1093/europace/euaf085.443
Primary Topic
Cardiac Arrhythmias and Treatments
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Post ablation mapping and inducibility of atrial tachycardia in different pulsed field ablation technologies

Verena Tscholl, Johannes Lucas, Patrick Nagel, Tobias Schreiber et al.
EP Europace
Cardiac Arrhythmias and Treatments
article

Post ablation mapping and inducibility of atrial tachycardia in different pulsed field ablation technologies

Verena Tscholl, Johannes Lucas, Patrick Nagel, Tobias Schreiber, C Baldauf, Martin Huemer, Ulf Landmesser, Philipp Attanasio, G Hindricks, P Wienke
article en

Abstract

Abstract Introduction Pulsed field ablation (PFA) is a rapidly developing non-thermal ablation method leading to irreversible electroporation by application of short electrical pulses. Several PFA catheters with different designs, voltage output and energy delivery have been developed (1), therefore lesion durability and size may differ between the systems. This study was designed to compare lesion sizes and to evaluate if large posterior wall lesions leading to small interlesion distances are associated with roof dependent flutter inducibility and occurrence at follow up. Methods Consecutive patients who presented for atrial fibrillation ablation with PFA using three different ablation technologies were included. All patients underwent high density post ablation mapping. Peak-to-peak electrogram bipolar amplitude < 0.5 mV was defined as low-voltage area threshold (Varipulse and PulseSelect; 5); for Affera cases, unipolar low-voltage maps were created with a cutoff of 0.05 mV. Interlesion distance was measured a) at the left atrial roof b) at the posterior wall at the minimal distance between the PV lesions and at the c) lower margin of the PV ablation lesion. In a subset of patients, atrial stimulation until the atrial refractory period was performed after PV ablation. Follow up was scheduled at 3, 6 and 12 months. Results 69 patients were included. Baseline characteristics showed no significant differences between the three groups (see table 1). PV isolation was achieved in all cases. Interlesion distances differed significantly between PFA systems (see table 2). Burst stimulation in 37 patients did not induce sustained atrial tachycardia. After a mean follow-up of three months, one patient experienced sustained atrial tachycardia (perimitral flutter). Conclusion To our knowledge, this study is the first to comparatively assess interlesion distances between three current PFA ablation systems. The study shows considerable variability in lesion size and therefore interlesion distance. So far there is no sign for an increased number of roof dependent flutter episodes during follow up.

EP EuropaceVol. 27(Supplement_1)
Openalex Percentile: Top 80%
Cardiac Arrhythmias and Treatments
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.