Can we predict if defibrillation testing of a subcutaneous implantable cardioverter defibrillator can be avoided safely at device replacement?

Abstract Background Current guidelines recommend to perform a defibrillation test (DFT) when a subcutaneous implantable cardioverter defibrillator replacement is performed. For transvenous ICD systems there is evidence that this can be safely omitted but evidence is lacking for patients that have a subcutaneous implantable cardioverter defibrillator implanted. Purpose To evaluate pre-procedural parameters that can predict the outcome of the DFT so this may safely be avoided. Methods A retrospective multicenter analysis of subcutaneous implantable cardioverter defibrillator patients undergoing a device replacement with or without DFT was performed. Results The cohort included 151 patients: 121 male, age 60.9±16.4 year, LVEF 40.5±7.1, primary prevention 65%. In total 176 device replacements where included. At first implant 136 patients ( 92.5%) had a DFT with a single 65J shock success rate of 93.4%. At first implant the average shock impedance was 77(±21). A Praetorian scores of 30-90 was achieved in 71 (60.2%) of the patients. Only 7 (5,9%) had a score ≥ 150 but still successful DFT. During first replacement 55% of the patients (83) underwent DFT testing with a 90,4% single shock success rate. Only 8 patients needed a second DFT. At the second replacement 9 Patients (36%) were tested with a single shock success rate of 100%. X rays were made in 55 (36,4%) and 5 (20,8%) before first or second replacement. X rays showed that 45 patients (84,9%) had a stabile lead position at first replacement and 3 patients (60%) at second replacement. A unstable position might influence the impedance or the Praetorian score in a negative way on shock conversion capability. Average shock impedance was 90 Ohm(± 25) at first replacement and 72 (±22) at second replacement. A high impedance lowers the power of the delivered shock and influence the efficacy of the DFT. More than 60% of the patients had a Praetorian scores of 30-90. Only 3 patients ( 5.9%) at first replacement had a score of ≥ 150. The praetorian score estimates the risk of conversion failure. A praetorian score of ≥ 90 has a positive predictive value of 25% for failed DFT. Conclusion It is feasible to safely avoid the DFT test if certain precautionary measures are performed before the device replacement. Comparing X-ray, Praetorian score and impedance measurements with the values of first implant can assist in making the decision of avoiding the DFT testing.

Authors

Institutions

Publication Details

Journal
EP Europace
Published
2025-05-01
DOI
https://doi.org/10.1093/europace/euaf085.088
Primary Topic
Cardiac pacing and defibrillation studies
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Can we predict if defibrillation testing of a subcutaneous implantable cardioverter defibrillator can be avoided safely at device replacement?

Pier D. Lambiase, Vincent F. van Dijk, Christopher Monkhouse, Lucas V.A. Boersma et al.
EP Europace
Cardiac pacing and defibrillation studies
article

Can we predict if defibrillation testing of a subcutaneous implantable cardioverter defibrillator can be avoided safely at device replacement?

Pier D. Lambiase, Vincent F. van Dijk, Christopher Monkhouse, Lucas V.A. Boersma, J Romers
article en

Abstract

Abstract Background Current guidelines recommend to perform a defibrillation test (DFT) when a subcutaneous implantable cardioverter defibrillator replacement is performed. For transvenous ICD systems there is evidence that this can be safely omitted but evidence is lacking for patients that have a subcutaneous implantable cardioverter defibrillator implanted. Purpose To evaluate pre-procedural parameters that can predict the outcome of the DFT so this may safely be avoided. Methods A retrospective multicenter analysis of subcutaneous implantable cardioverter defibrillator patients undergoing a device replacement with or without DFT was performed. Results The cohort included 151 patients: 121 male, age 60.9±16.4 year, LVEF 40.5±7.1, primary prevention 65%. In total 176 device replacements where included. At first implant 136 patients ( 92.5%) had a DFT with a single 65J shock success rate of 93.4%. At first implant the average shock impedance was 77(±21). A Praetorian scores of 30-90 was achieved in 71 (60.2%) of the patients. Only 7 (5,9%) had a score ≥ 150 but still successful DFT. During first replacement 55% of the patients (83) underwent DFT testing with a 90,4% single shock success rate. Only 8 patients needed a second DFT. At the second replacement 9 Patients (36%) were tested with a single shock success rate of 100%. X rays were made in 55 (36,4%) and 5 (20,8%) before first or second replacement. X rays showed that 45 patients (84,9%) had a stabile lead position at first replacement and 3 patients (60%) at second replacement. A unstable position might influence the impedance or the Praetorian score in a negative way on shock conversion capability. Average shock impedance was 90 Ohm(± 25) at first replacement and 72 (±22) at second replacement. A high impedance lowers the power of the delivered shock and influence the efficacy of the DFT. More than 60% of the patients had a Praetorian scores of 30-90. Only 3 patients ( 5.9%) at first replacement had a score of ≥ 150. The praetorian score estimates the risk of conversion failure. A praetorian score of ≥ 90 has a positive predictive value of 25% for failed DFT. Conclusion It is feasible to safely avoid the DFT test if certain precautionary measures are performed before the device replacement. Comparing X-ray, Praetorian score and impedance measurements with the values of first implant can assist in making the decision of avoiding the DFT testing.

EP EuropaceVol. 27(Supplement_1)
Barts Health NHS Trust (GB), St. Antonius Ziekenhuis (NL)
Boston Scientific Corporation
Openalex Percentile: Top 80%
Cardiac pacing and defibrillation studies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.