Intraoperative impedance and electrogram features can predict chronic capture threshold in active fixation ventricular leadless pacemakers

Abstract Background Leadless pacemakers (LP) with the capability to obtain electrical measurements such as impedance and electrograms (EGM) can provide early feedback on implant site selection, before committing to fixation and therefore minimizing the need to reposition the LP. Purpose The objective was to utilize intraoperative features of the electrogram (EGM) and paced impedance measurements to predict pacing capture thresholds (PCT) at the 3-month follow-up. Methods This is a retrospective study of a leadless pacemaker clinical trial (NCT#:05252702), including patients with complete sets of impedance measurements and intracardiac EGMs collected during the mapping phase and while in tether mode, and a capture threshold obtained at the 3-month follow-up. A computerized algorithm was developed to quantify features of the EGM signal: amplitudes of the R-wave, S-wave, and COI, the slope of the upstroke and downstroke, and the sharpness of the R-wave peak (calculated as average slope of points within 1 sample of the peak). Linear regression was performed to identify significant predictors of the chronic PCT. Binary logistic regression models were constructed by converting the 3-month PCT into a binary outcome using a cutoff of 1.5V and analyzed using receiver operating characteristic (ROC) curves. Results 88 patients were included. PCT at 3-months was 0.73±0.84 V. 8 patients had PCT >1.5V at 3 months. In univariate linear regression, impedance during mapping and tether, the sharpness of the R-wave during mapping, and the R-wave amplitude during tether were significant predictors of 3-month PCT (p=0.04, <0.01, 0.05, 0.03, respectively). Two logistic regression models were identified: 1) using only mapping variables (COI and impedance), 2) including both mapping COI and tether impedance. The mapping logistic regression model included COI (p=0.01) and impedance (p=0.1) during mapping and produced an area under the curve (AUC) of 0.88 with sensitivity and specificity of 100% and 70%, respectively. A logistic regression model including COI (mapping, p=0.04) and impedance (tether, p=0.03) produced an AUC of 0.92 with sensitivity and specificity of 100% and 81%, respectively. Test of the Χ2 statistic vs. constant model had p<0.01 in both models. Conclusion We developed a computerized prediction model using intraoperative EGM and impedance to predict 3-month PCT of a leadless pacemaker. This may be useful in enhancing procedural efficacy and efficiency.Linear Regression Results Binary Logistic Regression ROC Curves

Authors

Institutions

Publication Details

Journal
EP Europace
Published
2025-05-01
DOI
https://doi.org/10.1093/europace/euaf085.573
Primary Topic
Cardiac pacing and defibrillation studies
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Intraoperative impedance and electrogram features can predict chronic capture threshold in active fixation ventricular leadless pacemakers

Derek V. Exner, Leyla Sabet, Morio Shoda, Gerhard Hindricks et al.
EP Europace
Cardiac pacing and defibrillation studies
article

Intraoperative impedance and electrogram features can predict chronic capture threshold in active fixation ventricular leadless pacemakers

Derek V. Exner, Leyla Sabet, Morio Shoda, Gerhard Hindricks, J Guthrie, James E. Ip, Rahul N. Doshi, Chad D. Huff, Reinoud E. Knops, Edward A. Johnson, Pascal Defaye, T K Tam, Ashok Reddy, Robert C. Canby, M. G. Bongiorni
article en

Abstract

Abstract Background Leadless pacemakers (LP) with the capability to obtain electrical measurements such as impedance and electrograms (EGM) can provide early feedback on implant site selection, before committing to fixation and therefore minimizing the need to reposition the LP. Purpose The objective was to utilize intraoperative features of the electrogram (EGM) and paced impedance measurements to predict pacing capture thresholds (PCT) at the 3-month follow-up. Methods This is a retrospective study of a leadless pacemaker clinical trial (NCT#:05252702), including patients with complete sets of impedance measurements and intracardiac EGMs collected during the mapping phase and while in tether mode, and a capture threshold obtained at the 3-month follow-up. A computerized algorithm was developed to quantify features of the EGM signal: amplitudes of the R-wave, S-wave, and COI, the slope of the upstroke and downstroke, and the sharpness of the R-wave peak (calculated as average slope of points within 1 sample of the peak). Linear regression was performed to identify significant predictors of the chronic PCT. Binary logistic regression models were constructed by converting the 3-month PCT into a binary outcome using a cutoff of 1.5V and analyzed using receiver operating characteristic (ROC) curves. Results 88 patients were included. PCT at 3-months was 0.73±0.84 V. 8 patients had PCT >1.5V at 3 months. In univariate linear regression, impedance during mapping and tether, the sharpness of the R-wave during mapping, and the R-wave amplitude during tether were significant predictors of 3-month PCT (p=0.04, <0.01, 0.05, 0.03, respectively). Two logistic regression models were identified: 1) using only mapping variables (COI and impedance), 2) including both mapping COI and tether impedance. The mapping logistic regression model included COI (p=0.01) and impedance (p=0.1) during mapping and produced an area under the curve (AUC) of 0.88 with sensitivity and specificity of 100% and 70%, respectively. A logistic regression model including COI (mapping, p=0.04) and impedance (tether, p=0.03) produced an AUC of 0.92 with sensitivity and specificity of 100% and 81%, respectively. Test of the Χ2 statistic vs. constant model had p<0.01 in both models. Conclusion We developed a computerized prediction model using intraoperative EGM and impedance to predict 3-month PCT of a leadless pacemaker. This may be useful in enhancing procedural efficacy and efficiency.Linear Regression Results Binary Logistic Regression ROC Curves

EP EuropaceVol. 27(Supplement_1)
Chinese University of Hong Kong (HK), Cornell University (US), Abbott (Sweden) (SE), Centre Hospitalier Universitaire de Grenoble (FR), Foothills Medical Centre (CA), HonorHealth (US), Tokyo Women's Medical University (JP), Texas Cardiac Arrhythmia (US), Amsterdam University Medical Centers (NL), Weill Cornell Medicine (US), Charité - Universitätsmedizin Berlin (DE), Icahn School of Medicine at Mount Sinai (US)
Openalex Percentile: Top 80%
Cardiac pacing and defibrillation studies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.