Abstract Background Recent studies suggest that atrioventricular delay (AVD) optimization can play a key role in QRSd narrowing in patients treated with cardiac resynchronization therapy (CRT) , thus improving outcomes. For this purpose, device-based algorithms have been developed. These algorithms continuously monitor the patient’s intrinsic AVD and dynamically adjust the pacing AVD to allow fusion of the left ventricle pacing with intrinsic right ventricle depolarization. AVD optimization is usually performed in a single patient condition, i.e. at rest, and in a single operating scenario of the device, i.e. during atrial sensing. Purpose This study aims to compare QRS narrowing after AVD optimization in patients treated with CRT, during moderate exertion, and during atrial pacing. Methods This is a prospective, non-randomized, non-blinded, multicenter study. The study population consisted of patients in sinus rhythm, PR interval < 350 ms, and rest rate < 100 bpm. Patients were assessed for AVD optimization with SyncAV, selecting the offset (i.e. 10 to 40% AVD) that resulted in the shortest QRS duration (QRSd). Subsequently, patients performed moderate physical activity using a bicycle ergo-meter while a surface ECG was collected. QRSd was assessed once patients achieved a heart rate over 20 bpm to baseline heart rate and during a programmed increase in atrial pacing. Results 20 patients (mean age 70 ± 9.6 years; 50% male) underwent CRT-D implantation, the mean baseline PR interval was 178 ± 26 ms, QRSd 162 ± 21 ms, and left ventricular ejection fraction (LVEF) 29 ± 4%. AVD optimization was conducted at rest using SyncAV, yielding a median optimal offset of 20% (IQR: 15%–22.5%). Optimized SyncAV resulted in a 20% QRSd reduction compared to intrinsic (157.5 ± 24.1 ms to 125.9 ± 15.8 ms, p < 0.05) and an 8% reduction compared to biventricular (BiV) pacing with fixed AVD (p = 0.13). During exercise, heart rate increased by 23% (65.5 ± 9.1 bpm to 84.9 ± 14.4 bpm, p < 0.05), with a mean QRSd reduction of 19% using SyncAV compared to intrinsic (157.8 ± 26.1 ms to 127.7 ± 17.4 ms, p < 0.05). Also, during atrial pacing, SyncAV showed a 13% QRSd reduction compared to intrinsic (150.0 ± 23.0 ms to 130.1 ± 18.8 ms, p < 0.05). No significant differences were found in QRSd across rest, exercise, and atrial pacing conditions with optimized SyncAV, suggesting consistent benefit in QRS narrowing regardless patients heart rate and device operating scenario. Conclusions AVD optimization using SyncAV reduced QRS duration compared to intrinsic conduction and BiV pacing with fixed AVD, both at rest and during exercise and atrial pacing. These findings suggest that device-based algorithms effectively enhance ventricular synchrony and boosts CRT-D efficacy across various physiological conditions. This optimization may contribute to improved clinical outcomes, supporting its use as a valuable tool to personalize CRT-D therapy.

Authors

Institutions

Publication Details

Journal
EP Europace
Published
2025-05-01
DOI
https://doi.org/10.1093/europace/euaf085.620
Primary Topic
Advanced MRI Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Effectiveness of dynamic atrioventricular optimization in CRT patients across different physiological conditions

G Caroli, S Leccisotti, G Covino, D Vittoria et al.
EP Europace
Advanced MRI Techniques and Applications
article

Effectiveness of dynamic atrioventricular optimization in CRT patients across different physiological conditions

G Caroli, S Leccisotti, G Covino, D Vittoria, R Calvanese, Assunta Iuliano, A Medico, Michelangelo Canciello, Alessio Simonetti, Vincenzo Sollazzo, Valter Bianchi, Giorgio Spiniello, Sara Poggi
article en

Abstract

Abstract Background Recent studies suggest that atrioventricular delay (AVD) optimization can play a key role in QRSd narrowing in patients treated with cardiac resynchronization therapy (CRT) , thus improving outcomes. For this purpose, device-based algorithms have been developed. These algorithms continuously monitor the patient’s intrinsic AVD and dynamically adjust the pacing AVD to allow fusion of the left ventricle pacing with intrinsic right ventricle depolarization. AVD optimization is usually performed in a single patient condition, i.e. at rest, and in a single operating scenario of the device, i.e. during atrial sensing. Purpose This study aims to compare QRS narrowing after AVD optimization in patients treated with CRT, during moderate exertion, and during atrial pacing. Methods This is a prospective, non-randomized, non-blinded, multicenter study. The study population consisted of patients in sinus rhythm, PR interval < 350 ms, and rest rate < 100 bpm. Patients were assessed for AVD optimization with SyncAV, selecting the offset (i.e. 10 to 40% AVD) that resulted in the shortest QRS duration (QRSd). Subsequently, patients performed moderate physical activity using a bicycle ergo-meter while a surface ECG was collected. QRSd was assessed once patients achieved a heart rate over 20 bpm to baseline heart rate and during a programmed increase in atrial pacing. Results 20 patients (mean age 70 ± 9.6 years; 50% male) underwent CRT-D implantation, the mean baseline PR interval was 178 ± 26 ms, QRSd 162 ± 21 ms, and left ventricular ejection fraction (LVEF) 29 ± 4%. AVD optimization was conducted at rest using SyncAV, yielding a median optimal offset of 20% (IQR: 15%–22.5%). Optimized SyncAV resulted in a 20% QRSd reduction compared to intrinsic (157.5 ± 24.1 ms to 125.9 ± 15.8 ms, p < 0.05) and an 8% reduction compared to biventricular (BiV) pacing with fixed AVD (p = 0.13). During exercise, heart rate increased by 23% (65.5 ± 9.1 bpm to 84.9 ± 14.4 bpm, p < 0.05), with a mean QRSd reduction of 19% using SyncAV compared to intrinsic (157.8 ± 26.1 ms to 127.7 ± 17.4 ms, p < 0.05). Also, during atrial pacing, SyncAV showed a 13% QRSd reduction compared to intrinsic (150.0 ± 23.0 ms to 130.1 ± 18.8 ms, p < 0.05). No significant differences were found in QRSd across rest, exercise, and atrial pacing conditions with optimized SyncAV, suggesting consistent benefit in QRS narrowing regardless patients heart rate and device operating scenario. Conclusions AVD optimization using SyncAV reduced QRS duration compared to intrinsic conduction and BiV pacing with fixed AVD, both at rest and during exercise and atrial pacing. These findings suggest that device-based algorithms effectively enhance ventricular synchrony and boosts CRT-D efficacy across various physiological conditions. This optimization may contribute to improved clinical outcomes, supporting its use as a valuable tool to personalize CRT-D therapy.

EP EuropaceVol. 27(Supplement_1)
Abbott (Sweden) (SE), Ospedale A. Perrino (IT), Nuovo Ospedale San Giovanni di Dio (IT), Clinica Mediterranea (IT), Ospedale Monaldi (IT)
Openalex Percentile: Top 82%
Advanced MRI Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.