RELIABILITY OF BLOOD PRESSURE ESTIMATION FOR EXTREME VALUES

Objective: The use of a cuff for continuous blood pressure (BP) monitoring is limited by discomfort, interference with measurements and excessive time between measurements. The new cuffless methods based on pulse wave velocity seem to provide measurements with a correct accuracy. However, this accuracy may not maintained over the whole range of BP values. This study explores the accuracy of BP estimation using photoplethysmography (PPG) signals, in a wide range of intra-individual BP variations. Design and method: The cohort of the clinical trial (NCT05393401, approved by Human Research Review Board) included 18 healthy subjects (9F/9 M, age [19-60] years). Each subject undertook physical / mental activities and relaxation. The subject was monitored with electrocardiogram (ECG) and PPG sensors at the radial artery, at 1 kHz frequency. Signals were collected and synchronised by a custom-built multi-sensing platform. The Pulse Arrival Time was computed from ECG R-peaks and PPG signals for each heartbeat. Other inputs included subject’s age and body mass index, and pulse wave morphological metrics. Cuffless BP was estimated using the Adaboost regressor. In parallel, the subject’s BP was measured at every heartbeat with a medical-grade continuous measurement device. The model performance was quantified by the root mean square error (RMSE). Results: The measured systolic pressure ranged from 100 to 193 mmHg and the diastolic pressure between 49 and 106 mmHg. In this dataset, 84% of systolic values are within the interval [110;160] mmHg and are estimated with a RMSE of 13.4 mmHg. Similarly, 78% of diastolic values are within [60;90] mmHg and are estimated with a RMSE of 8.0 mmHg. For higher BP values, the estimation performances deteriorate with a RMSE of 22.0 mmHg for systolic and 14.5 mmHg for diastolic. For lower values, the RMSE are also worse with 26.8 mmHg for systolic and 11.6 mmHg for diastolic BP.Conclusions: The algorithms used to estimate BP from the pulse wave must be tested not only at small variations but also at extreme values to increase the robustness of the model. The experiments must also provide a significant proportion of high and low pressure values.

Authors

Institutions

Publication Details

Journal
Journal of Hypertension
Published
2025-05-01
DOI
https://doi.org/10.1097/01.hjh.0001116112.00896.d2
Primary Topic
Hemodynamic Monitoring and Therapy
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

RELIABILITY OF BLOOD PRESSURE ESTIMATION FOR EXTREME VALUES

Rémi Gerbelot, Daniel Anglade, Emma Villeneuve, Caroline Sandre-Ballester et al.
1 links
Journal of Hypertension
Hemodynamic Monitoring and Therapy
article

RELIABILITY OF BLOOD PRESSURE ESTIMATION FOR EXTREME VALUES

Rémi Gerbelot, Daniel Anglade, Emma Villeneuve, Caroline Sandre-Ballester, Loïc Auclair, Marine Mastrosimone, Robinson Prada-Mejia, Julie Rindone, Pierre Blandin, Olivier Ormezzano, Marion Coquand-Gandit
article en

Abstract

Objective: The use of a cuff for continuous blood pressure (BP) monitoring is limited by discomfort, interference with measurements and excessive time between measurements. The new cuffless methods based on pulse wave velocity seem to provide measurements with a correct accuracy. However, this accuracy may not maintained over the whole range of BP values. This study explores the accuracy of BP estimation using photoplethysmography (PPG) signals, in a wide range of intra-individual BP variations. Design and method: The cohort of the clinical trial (NCT05393401, approved by Human Research Review Board) included 18 healthy subjects (9F/9 M, age [19-60] years). Each subject undertook physical / mental activities and relaxation. The subject was monitored with electrocardiogram (ECG) and PPG sensors at the radial artery, at 1 kHz frequency. Signals were collected and synchronised by a custom-built multi-sensing platform. The Pulse Arrival Time was computed from ECG R-peaks and PPG signals for each heartbeat. Other inputs included subject’s age and body mass index, and pulse wave morphological metrics. Cuffless BP was estimated using the Adaboost regressor. In parallel, the subject’s BP was measured at every heartbeat with a medical-grade continuous measurement device. The model performance was quantified by the root mean square error (RMSE). Results: The measured systolic pressure ranged from 100 to 193 mmHg and the diastolic pressure between 49 and 106 mmHg. In this dataset, 84% of systolic values are within the interval [110;160] mmHg and are estimated with a RMSE of 13.4 mmHg. Similarly, 78% of diastolic values are within [60;90] mmHg and are estimated with a RMSE of 8.0 mmHg. For higher BP values, the estimation performances deteriorate with a RMSE of 22.0 mmHg for systolic and 14.5 mmHg for diastolic. For lower values, the RMSE are also worse with 26.8 mmHg for systolic and 11.6 mmHg for diastolic BP.Conclusions: The algorithms used to estimate BP from the pulse wave must be tested not only at small variations but also at extreme values to increase the robustness of the model. The experiments must also provide a significant proportion of high and low pressure values.

Journal of HypertensionVol. 43(Suppl 1)
Commissariat à l'Énergie Atomique et aux Énergies Alternatives (FR), Centre Hospitalier Universitaire de Grenoble (FR), CEA Grenoble (FR), Clinatec (FR), Laboratoire d'Électronique des Technologies de l'Information (FR), Université Grenoble Alpes (FR)
Openalex Percentile: Top 80%
Hemodynamic Monitoring and Therapy
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.