Evaluation of the Accuracy of a Mechanical Insufflation-Exsufflation Device’s Cough Peak Flow Measurement

<b>Background:</b> Mechanical insufflation-exsufflation (MI-E) is used to augment secretion clearance in neuromuscular patients with weakened cough strength. Cough peak flow (CPF) is a measure of cough function that is used to assess a patient's ability to clear secretions, with thresholds set that categorize cough as effective, ineffective or severely ineffective. MI-E is prescribed according to these thresholds, and CPF is used to assess titration of MI-E settings. The Clearway2 (Breas Medical, Stratford-upon-Avon, United Kingdom) displays a real-time CPF, measured by an internal pneumotachograph. This study sought to assess the agreement and repeatability of this displayed CPF, against the reference CPF measurement by a calibrated pneumotachograph. <b>Methods:</b> This study consisted of two phases (1) lung model (Group A) and (2) acutely unwell individuals with neuromuscular conditions (Group B) and clinically stable individuals with neuromuscular conditions (Group C). Simultaneous CPF measurements were recorded from the MI-E device (CPF<sub>MI-E</sub>) and a calibrated pneumotachograph (CPF), which was inserted into the MI-E circuit. Bland-Altman analysis was used to assess agreement between methods of measurement, and repeatability was assessed using a repeated measures analysis of variance. <b>Results:</b> During phase 1, 805 simulated coughs were evaluated with the Clearway2. The mean bias toward CPF<sub>MI-E</sub> was 33 L/min (95% limits of agreement 6-60 L/min). During phase 2, the mean bias increased to 66 L/min (95% limits of agreement 13-119 L/min). CPF<sub>MI-E</sub> and CPF both had good repeatability in all groups. <b>Conclusions:</b> The Clearway2 MI-E device provided a real-time CPF measurement that was repeatable and systematically higher than CPF. It may therefore be a useful tool to measure change longitudinally, or in response to changes in MI-E settings, in an individual patient. Caution is advised if using the CPF<sub>MI-E</sub> to assess cough efficacy against clinical thresholds.

Authors

Institutions

Publication Details

Journal
Respiratory Care
Published
2025-05-22
DOI
https://doi.org/10.1089/respcare.12514
Citations
2
Primary Topic
Respiratory Support and Mechanisms
Type
article
Field-Weighted Citation Impact
3.04
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Evaluation of the Accuracy of a Mechanical Insufflation-Exsufflation Device’s Cough Peak Flow Measurement

Leyla Osman, Ema L Swingwood, Eui‐Sik Suh, Patrick B. Murphy et al.
2 citations
Respiratory Care
Respiratory Support and Mechanisms
3.04
article

Evaluation of the Accuracy of a Mechanical Insufflation-Exsufflation Device’s Cough Peak Flow Measurement

Leyla Osman, Ema L Swingwood, Eui‐Sik Suh, Patrick B. Murphy, Harriet Shannon, Neeraj Shah, Georgios Kaltsakas, Nicholas Hart, Chloe Apps, Sophie Madden-Scott
article en
2 citations

Abstract

<b>Background:</b> Mechanical insufflation-exsufflation (MI-E) is used to augment secretion clearance in neuromuscular patients with weakened cough strength. Cough peak flow (CPF) is a measure of cough function that is used to assess a patient's ability to clear secretions, with thresholds set that categorize cough as effective, ineffective or severely ineffective. MI-E is prescribed according to these thresholds, and CPF is used to assess titration of MI-E settings. The Clearway2 (Breas Medical, Stratford-upon-Avon, United Kingdom) displays a real-time CPF, measured by an internal pneumotachograph. This study sought to assess the agreement and repeatability of this displayed CPF, against the reference CPF measurement by a calibrated pneumotachograph. <b>Methods:</b> This study consisted of two phases (1) lung model (Group A) and (2) acutely unwell individuals with neuromuscular conditions (Group B) and clinically stable individuals with neuromuscular conditions (Group C). Simultaneous CPF measurements were recorded from the MI-E device (CPF<sub>MI-E</sub>) and a calibrated pneumotachograph (CPF), which was inserted into the MI-E circuit. Bland-Altman analysis was used to assess agreement between methods of measurement, and repeatability was assessed using a repeated measures analysis of variance. <b>Results:</b> During phase 1, 805 simulated coughs were evaluated with the Clearway2. The mean bias toward CPF<sub>MI-E</sub> was 33 L/min (95% limits of agreement 6-60 L/min). During phase 2, the mean bias increased to 66 L/min (95% limits of agreement 13-119 L/min). CPF<sub>MI-E</sub> and CPF both had good repeatability in all groups. <b>Conclusions:</b> The Clearway2 MI-E device provided a real-time CPF measurement that was repeatable and systematically higher than CPF. It may therefore be a useful tool to measure change longitudinally, or in response to changes in MI-E settings, in an individual patient. Caution is advised if using the CPF<sub>MI-E</sub> to assess cough efficacy against clinical thresholds.

Respiratory CareVol. 70(9)
University of the West of England (GB), King's College London (GB), Guy's and St Thomas' NHS Foundation Trust (GB), University Hospitals Bristol NHS Foundation Trust (GB), King's College School (GB), Institute of Infection and Immunity (CA), University College London (GB)
Openalex Percentile: Top 9%
Respiratory Support and Mechanisms
3.04
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.