Impact of Patient Size and Positioning on Radiation Dose in Multiphase Liver CT Examinations

Investigating the correlation between patient dimensions (Body Mass Index (BMI), weight, height) and radiation dose, the study also assesses the impact of the isocenter on dosage and X-ray tube. This retrospective study analyzed 258 consecutive three-phase liver CT exams (135 women, 123 men) performed on a Siemens SOMATOM Definition Edge scanner between January 3 and December 15, 2023. BMI, weight, height, maximum abdominal width (from topograms), and vertical isocenter status were extracted using the Dose&Care system and Horos software. BMI strongly correlated with total DLP in abdomen-plus-pelvis scans (r = 0.70) and moderately in abdomen-only scans (r = 0.54). Liver-phase DLP correlations were weaker (r = 0.44 and r = 0.28, respectively). Abdominal width showed similar associations with total DLP (r = 0.67 and r = 0.64) and liver-phase DLP (r = 0.41 and 0.37). Vertical mis-centering did not significantly affect total DLP (p = 0.174 and p = 0.705) or tube load. Patient size—not minor deviations from the isocenter—is the principal driver of radiation dose in multiphase liver CT. Automated, size-adapted protocols and restriction of scan range to clinically essential regions can preserve image quality while minimizing unnecessary radiation in routine practice.

Authors

Institutions

Publication Details

Journal
Applied Sciences
Published
2025-05-22
DOI
https://doi.org/10.3390/app15115815
Primary Topic
Radiation Dose and Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Impact of Patient Size and Positioning on Radiation Dose in Multiphase Liver CT Examinations

Sandra Modlińska, Marcin Rojek, Jakub Kufel, Michał Bielówka et al.
Applied Sciences
Radiation Dose and Imaging
article

Impact of Patient Size and Positioning on Radiation Dose in Multiphase Liver CT Examinations

Sandra Modlińska, Marcin Rojek, Jakub Kufel, Michał Bielówka, Adam Mitręga, Michał Azierski, Michał Janik, Natalia Denisiewicz
article en

Abstract

Investigating the correlation between patient dimensions (Body Mass Index (BMI), weight, height) and radiation dose, the study also assesses the impact of the isocenter on dosage and X-ray tube. This retrospective study analyzed 258 consecutive three-phase liver CT exams (135 women, 123 men) performed on a Siemens SOMATOM Definition Edge scanner between January 3 and December 15, 2023. BMI, weight, height, maximum abdominal width (from topograms), and vertical isocenter status were extracted using the Dose&Care system and Horos software. BMI strongly correlated with total DLP in abdomen-plus-pelvis scans (r = 0.70) and moderately in abdomen-only scans (r = 0.54). Liver-phase DLP correlations were weaker (r = 0.44 and r = 0.28, respectively). Abdominal width showed similar associations with total DLP (r = 0.67 and r = 0.64) and liver-phase DLP (r = 0.41 and 0.37). Vertical mis-centering did not significantly affect total DLP (p = 0.174 and p = 0.705) or tube load. Patient size—not minor deviations from the isocenter—is the principal driver of radiation dose in multiphase liver CT. Automated, size-adapted protocols and restriction of scan range to clinically essential regions can preserve image quality while minimizing unnecessary radiation in routine practice.

Applied SciencesVol. 15(11)
Medical University of Silesia (PL)
Openalex Percentile: Top 82%
Radiation Dose and Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.