Feasibility Study of Thallium Bromide Detectors for Boron Neutron Capture Therapy Applications
A thallium bromide (TlBr) detector with four-channel readout electrodes was fabricated from a crystal grown by the traveling molten zone method from a zone-purified material for boron neutron capture therapy applications.The detector crystal dimensions were 12 12 10 mm 3 .The detector had four circular anodes of 2 mm diameter.The detector exhibited energy resolutions sufficient to resolve 478 keV gamma rays from 511 keV gamma rays.A prompt gamma-ray detection system was constructed using the TlBr detector and a tungsten collimator.A mouse phantom with a tumor sphere containing 1475 ppm 10 B was irradiated with thermal neutrons.The 10 B distribution was imaged successfully with the TlBr detector system.The detector system exhibited a linear relationship between the counts for 478 keV and the 10 B concentration in the range of 0-5000 ppm.
Authors
- Hiroki Tanaka (ORCID: https://orcid.org/0000-0002-5178-7777)
- Nishiki Matsubayashi (ORCID: https://orcid.org/0000-0002-0090-7759)
- Keitaro Hitomi (ORCID: https://orcid.org/0000-0002-4844-9136)
- Kenichi Watanabe
- Mitsuhiro Nogami
Publication Details
- Journal
- Sensors and Materials
- Published
- 2025-05-22
- DOI
- https://doi.org/10.18494/sam5613
- Citations
- 3
- Primary Topic
- Boron Compounds in Chemistry
- Type
- article
- Field-Weighted Citation Impact
- 4.08
Funders
- Japan Society for the Promotion of Science