A 0.48° Phase Error 91.1dB SNR Bioimpedance Measurement IC for Monitoring Cardiopulmonary Diseases

This article presents a low-power and low phase error bioimpedance (BioZ) measurement IC designed for monitoring cardiopulmonary diseases. To compensate for the phase shift introduced along the signal path by current generator (CG), electrodes and sensor analog front-end (AFE), a novel phase shift calibration logic is proposed. Utilizing this calibration logic, a single-channel in-phase demodulation-based impedance measurement scheme is developed. A noise shaping pseudo-sine wave CG with data-weighted averaging (DWA) is used to minimize modulation harmonics. Fabricated in a 0.18µm CMOS process, the chip occupies 0.73 mm2 and consumes between 52.7 to 97.5 µA current from a 1.8V supply. The CG achieves 74.1 dB SFDR and -70dB THD at 15.5 kHz with a 50µApk stimulation current. The chip achieves 2mΩ/√Hz input-referred impedance noise at 1Hz, 91.1 dB SNR (BW = 4 Hz), 36 kΩ input range and less than 0.48$^{\circ}$ phase error (0 - 90$^{\circ}$, 1 - 20 kHz). On-body BioZ experiments using a 4-electrode configuration demonstrate clear recordings of Impedance Cardiography (ICG) and respiration signals.

Authors

Institutions

Publication Details

Journal
IEEE Transactions on Biomedical Circuits and Systems
Published
2025-01-01
DOI
https://doi.org/10.1109/tbcas.2025.3572374
Primary Topic
Body Composition Measurement Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A 0.48° Phase Error 91.1dB SNR Bioimpedance Measurement IC for Monitoring Cardiopulmonary Diseases

Yanxing Suo, Qiao Cai, Yong Lian, Hui Wang et al.
4 links
IEEE Transactions on Biomedical Circuits and Systems
Body Composition Measurement Techniques
article

A 0.48° Phase Error 91.1dB SNR Bioimpedance Measurement IC for Monitoring Cardiopulmonary Diseases

Yanxing Suo, Qiao Cai, Yong Lian, Hui Wang, Yongfu Li, Yang Zhao, Jiarun Yuan
article en

Abstract

This article presents a low-power and low phase error bioimpedance (BioZ) measurement IC designed for monitoring cardiopulmonary diseases. To compensate for the phase shift introduced along the signal path by current generator (CG), electrodes and sensor analog front-end (AFE), a novel phase shift calibration logic is proposed. Utilizing this calibration logic, a single-channel in-phase demodulation-based impedance measurement scheme is developed. A noise shaping pseudo-sine wave CG with data-weighted averaging (DWA) is used to minimize modulation harmonics. Fabricated in a 0.18µm CMOS process, the chip occupies 0.73 mm2 and consumes between 52.7 to 97.5 µA current from a 1.8V supply. The CG achieves 74.1 dB SFDR and -70dB THD at 15.5 kHz with a 50µApk stimulation current. The chip achieves 2mΩ/√Hz input-referred impedance noise at 1Hz, 91.1 dB SNR (BW = 4 Hz), 36 kΩ input range and less than 0.48$^{\circ}$ phase error (0 - 90$^{\circ}$, 1 - 20 kHz). On-body BioZ experiments using a 4-electrode configuration demonstrate clear recordings of Impedance Cardiography (ICG) and respiration signals.

IEEE Transactions on Biomedical Circuits and SystemsVol. 19(5)
Shanghai Jiao Tong University (CN), York University (CA)
Openalex Percentile: Top 85%
Body Composition Measurement Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.