Embedded C-Dipole Antenna for the Wireless Monitoring of Abdominal Grafts
Abdominal aortic aneurysms (AAAs) and related vascular conditions often require the implantation of stent grafts to prevent life-threatening complications, such as rupture or internal bleeding, yet the long-term success of these devices can be compromised by issues like endoleaks, thrombosis, and mechanical failure. This paper presents the design and testing of a wireless RFID-based sensing device integrated into an abdominal graft for real-time monitoring. The system, which includes a miniaturized C-dipole antenna embedded between the stent's metallic rings, leverages battery-free RFID technology for remote data transmission. The proposed approach allows non-invasive, continuous monitoring of the stent graft, reducing the need for costly imaging techniques and hospital visits. The numerical simulations and preliminary tests in a liquid human body phantom show that the device allows a reliable communication link at a depth of 7 cm, and so has the potential to detect complications by leveraging on its sensing capabilities.
Authors
- Francesco M. Mancuso (ORCID: https://orcid.org/0000-0001-5499-6630)
- Gaetano Marrocco (ORCID: https://orcid.org/0000-0003-3151-3071)
- F. Nanni (ORCID: https://orcid.org/0000-0002-7282-8507)
- Sergio Biondi
- Camilla Medori
- Eleonora Oddi
Institutions
- University of Rome Tor Vergata (IT)
Publication Details
- Published
- 2025-03-30
- DOI
- https://doi.org/10.23919/eucap63536.2025.10999844
- Primary Topic
- Wireless Body Area Networks
- Type
- article
- Field-Weighted Citation Impact
- 0.00