Assessing Array-Type Differences in Cochlear Implant Users Using the Panoramic ECAP Method
These results suggest that for users of the Cochlear device, precurved electrode arrays show narrower current spread within the cochlea than those with lateral-wall electrode arrays, with the strongest effect present at the apex. No corresponding effects of array type were found in the Advanced Bionics device. This could have implications for device selection in clinical settings, although the authors underscore that this is a post-hoc analysis and does not demonstrate a causal link wherein device selection can be expected to give rise to specific neural excitation patterns.