Lethal isotherm ratio to standardize the comparison of cryoablation needles and systems

Abstract Purpose To establish the lethal isotherm ratio (LIR), a standardized metric for comparing cryoprobes and technologies based on their performance. Materials and methods Isotherms at the freezing front (0°C), -20°C, and -40°C, measured in ultrasonic gels under standardized conditions and FDA-certified, were obtained from Argon-based Joule-Thompson systems A and B, and liquid nitrogen-based system C ( n = 216). The LIR—defined as the ratio between isotherm widths at lethal temperatures (-20°C/-40°C) and the visible freezing front (0°C)—was calculated to compare systems, single needles, and needle combinations. Results Strong linear correlations between isotherms were observed across all systems (ρ:0.880–0.999, p = .0001). System C showed significantly higher LIRs (80.6% ± 3.9 at -20°C [95%CI: 79.8–81.4%], 60.3% ± 5.6 at -40°C [95%CI: 59.2–61.4%]) than systems A (69.6% ± 4.6 [95%CI: 68.7–70.5%], 44% ± 6.3 [95%CI: 42.8–45.2%]; both p = .0001) and B (69.6% ± 7 [95%CI: 67.2–72.0%], 51.4% ± 7.4 [95%CI: 48.9–53.9%]; p = .0001, p = .0002). Linear regression analysis demonstrated that clustered needles provide higher LIRs for ice balls exceeding 45mm width. Conclusion A single liquid nitrogen needle offers superior LIR for ice balls <45mm, whereas needle clusters are more effective beyond this threshold. LIR might provide clinicians a standardized approach to optimize system selection, potentially improving procedural outcomes and reducing local recurrence rates.

Authors

Institutions

Publication Details

Journal
CVIR Oncology
Published
2025-05-20
DOI
https://doi.org/10.1007/s44343-025-00004-4
Citations
2
Primary Topic
Spacecraft and Cryogenic Technologies
Type
article
Field-Weighted Citation Impact
9.99
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Lethal isotherm ratio to standardize the comparison of cryoablation needles and systems

Sylvain Bodard, F. Cornélis, Arthur A. Cornelis
2 citations
CVIR Oncology
Spacecraft and Cryogenic Technologies
9.99
article

Lethal isotherm ratio to standardize the comparison of cryoablation needles and systems

Sylvain Bodard, F. Cornélis, Arthur A. Cornelis
article en
2 citations

Abstract

Abstract Purpose To establish the lethal isotherm ratio (LIR), a standardized metric for comparing cryoprobes and technologies based on their performance. Materials and methods Isotherms at the freezing front (0°C), -20°C, and -40°C, measured in ultrasonic gels under standardized conditions and FDA-certified, were obtained from Argon-based Joule-Thompson systems A and B, and liquid nitrogen-based system C ( n = 216). The LIR—defined as the ratio between isotherm widths at lethal temperatures (-20°C/-40°C) and the visible freezing front (0°C)—was calculated to compare systems, single needles, and needle combinations. Results Strong linear correlations between isotherms were observed across all systems (ρ:0.880–0.999, p = .0001). System C showed significantly higher LIRs (80.6% ± 3.9 at -20°C [95%CI: 79.8–81.4%], 60.3% ± 5.6 at -40°C [95%CI: 59.2–61.4%]) than systems A (69.6% ± 4.6 [95%CI: 68.7–70.5%], 44% ± 6.3 [95%CI: 42.8–45.2%]; both p = .0001) and B (69.6% ± 7 [95%CI: 67.2–72.0%], 51.4% ± 7.4 [95%CI: 48.9–53.9%]; p = .0001, p = .0002). Linear regression analysis demonstrated that clustered needles provide higher LIRs for ice balls exceeding 45mm width. Conclusion A single liquid nitrogen needle offers superior LIR for ice balls <45mm, whereas needle clusters are more effective beyond this threshold. LIR might provide clinicians a standardized approach to optimize system selection, potentially improving procedural outcomes and reducing local recurrence rates.

CVIR OncologyVol. 1(1)
Memorial Sloan Kettering Cancer Center (US), Cornell University (US), Brooklyn Technical High School (US)
Openalex Percentile: Top 3%
Spacecraft and Cryogenic Technologies
9.99
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.