Development and validation of a combined selected ion monitoring‐scan <scp>GC</scp>‐<scp>EI</scp>‐<scp>MS</scp> method for nitazene analogs

Novel synthetic opioids (NSOs) emerged during the third wave of the opioid epidemic, which has resulted in countless overdose deaths. Although fentanyl and fentanyl analogs are the most common NSOs, changing legislation in China and the United States led to the emergence of another class of NSOs, known as nitazene analogs. Nitazene analogs are very potent, and given their increasing prevalence in seized drug casework, it is imperative that seized drug laboratories have sufficient methods to identify nitazene analogs. This study developed a combined selected ion monitoring (SIM)-scan method to identify 20 nitazene analogs using gas chromatography-electron ionization-mass spectrometry (GC-EI-MS). The method was evaluated in terms of the limit of detection (LOD), carryover, selectivity, repeatability, reproducibility, and processed sample stability. The LOD for the targeted nitazene analogs ranged from 5 to 10 ppm due to increased sensitivity through SIM acquisition. No carryover was observed, and all nitazene analogs could be differentiated from other analogs and common interferences. The repeatability and reproducibility of the method were demonstrated qualitatively. All nitazene analogs were stable at room temperature for at least 24 hours. When applied to the analysis of blind simulant samples, 33 out of 35 samples were correctly identified, with the two misidentifications being due to ion ratios outside of tolerance. Likewise, two authentic nitazene analog casework samples were also correctly identified. The validated SIM-scan method provides a potential solution to identify low concentration nitazene analogs in seized drug casework that might otherwise be missed using routine GC-MS full scan acquisition.

Authors

Institutions

Publication Details

Journal
Journal of Forensic Sciences
Published
2025-05-19
DOI
https://doi.org/10.1111/1556-4029.70084
Citations
3
Primary Topic
Mass Spectrometry Techniques and Applications
Type
article
Field-Weighted Citation Impact
2.27

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Development and validation of a combined selected ion monitoring‐scan GC‐EI‐MS method for nitazene analogs

Alleigh N. Couch, J. Tyler Davidson, Emma K. Hardwick, C. F. PHELPS
3 citations
Journal of Forensic Sciences
Mass Spectrometry Techniques and Applications
2.27
article

Development and validation of a combined selected ion monitoring‐scan GC‐EI‐MS method for nitazene analogs

Alleigh N. Couch, J. Tyler Davidson, Emma K. Hardwick, C. F. PHELPS
article en
3 citations

Abstract

Novel synthetic opioids (NSOs) emerged during the third wave of the opioid epidemic, which has resulted in countless overdose deaths. Although fentanyl and fentanyl analogs are the most common NSOs, changing legislation in China and the United States led to the emergence of another class of NSOs, known as nitazene analogs. Nitazene analogs are very potent, and given their increasing prevalence in seized drug casework, it is imperative that seized drug laboratories have sufficient methods to identify nitazene analogs. This study developed a combined selected ion monitoring (SIM)-scan method to identify 20 nitazene analogs using gas chromatography-electron ionization-mass spectrometry (GC-EI-MS). The method was evaluated in terms of the limit of detection (LOD), carryover, selectivity, repeatability, reproducibility, and processed sample stability. The LOD for the targeted nitazene analogs ranged from 5 to 10 ppm due to increased sensitivity through SIM acquisition. No carryover was observed, and all nitazene analogs could be differentiated from other analogs and common interferences. The repeatability and reproducibility of the method were demonstrated qualitatively. All nitazene analogs were stable at room temperature for at least 24 hours. When applied to the analysis of blind simulant samples, 33 out of 35 samples were correctly identified, with the two misidentifications being due to ion ratios outside of tolerance. Likewise, two authentic nitazene analog casework samples were also correctly identified. The validated SIM-scan method provides a potential solution to identify low concentration nitazene analogs in seized drug casework that might otherwise be missed using routine GC-MS full scan acquisition.

Journal of Forensic SciencesVol. 70(5)
Sam Houston State University (US)
U.S. Department of Commerce
Good health and well-being
Openalex Percentile: Top 12%
Mass Spectrometry Techniques and Applications
2.27
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.