Advanced Testing Protocols Simulate Failures and Validate Antioxidant Polyethylene in Ankle Implants

Total ankle replacement (TAR) has become an effective treatment for end-stage ankle osteoarthritis. Multiple factors, including patient characteristics, surgical technique, alignment, and bearing surfaces, influence TAR survivorship. Polyethylene (PE) fatigue is a key consideration in improving outcomes. This study establishes a novel, clinically relevant testing protocol incorporating varus-valgus rotation to simulate polyethylene fatigue failures observed in mobile-bearing total ankle replacements. Using this robust methodology, we evaluated the impact of oxidation and antioxidant stabilization on ultrahigh-molecular-weight polyethylene (UHMWPE) performance in a mobile bearing implant application. A six-degree-of-freedom simulator was used to iteratively adjust loading parameters (1500-3000 N, -4° to +8° flexion-extension, ±5° axial rotation, and ±3° or ±8° varus-valgus rotation at 37 ± 3°C in 20 g/L bovine serum) until clinically observed midline fractures were replicated. Oxidation levels were measured by Fourier-transform infrared spectroscopy per ASTM F2102. This validated loading protocol was then applied to conventional (25 kGy GUR 1020) and vitamin E-stabilized (75 kGy GUR 1020-E) UHMWPE inserts and tested to visible fracture or a 3-million-cycle runout. Post-test fractographic analysis identified crack initiation sites. Conventional aged UHMWPE demonstrated fatigue failure under varus-valgus rotation (OI = 2.59 ± 1.11) but no failure without rotation. Vitamin E-stabilized UHMWPE showed no fatigue failure after 3 million cycles, even under varus-valgus rotation (OI = 0.23 ± 0.02). Fractography revealed fractures originating at the trough and propagating with cyclic loading. Oxidation significantly reduces polyethylene fatigue life, and varus-valgus rotation exacerbates this effect in mobile bearing TAR implants. Antioxidant-stabilized UHMWPE showed promising resistance to fatigue and oxidation. These findings support the role of antioxidant stabilization in improving TAR performance, and the protocols developed here provide a framework for assessing the safety of alternative materials.

Authors

Institutions

Publication Details

Journal
Journal of Orthopaedic Research®
Published
2025-05-19
DOI
https://doi.org/10.1002/jor.26103
Primary Topic
Total Knee Arthroplasty Outcomes
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Advanced Testing Protocols Simulate Failures and Validate Antioxidant Polyethylene in Ankle Implants

Patrick T. Hall, James Weldon Johnson, Ryan Siskey, David L. Safranski et al.
Journal of Orthopaedic Research®
Total Knee Arthroplasty Outcomes
article

Advanced Testing Protocols Simulate Failures and Validate Antioxidant Polyethylene in Ankle Implants

Patrick T. Hall, James Weldon Johnson, Ryan Siskey, David L. Safranski, Ariel Palanca, Ruth Heckler
article en

Abstract

Total ankle replacement (TAR) has become an effective treatment for end-stage ankle osteoarthritis. Multiple factors, including patient characteristics, surgical technique, alignment, and bearing surfaces, influence TAR survivorship. Polyethylene (PE) fatigue is a key consideration in improving outcomes. This study establishes a novel, clinically relevant testing protocol incorporating varus-valgus rotation to simulate polyethylene fatigue failures observed in mobile-bearing total ankle replacements. Using this robust methodology, we evaluated the impact of oxidation and antioxidant stabilization on ultrahigh-molecular-weight polyethylene (UHMWPE) performance in a mobile bearing implant application. A six-degree-of-freedom simulator was used to iteratively adjust loading parameters (1500-3000 N, -4° to +8° flexion-extension, ±5° axial rotation, and ±3° or ±8° varus-valgus rotation at 37 ± 3°C in 20 g/L bovine serum) until clinically observed midline fractures were replicated. Oxidation levels were measured by Fourier-transform infrared spectroscopy per ASTM F2102. This validated loading protocol was then applied to conventional (25 kGy GUR 1020) and vitamin E-stabilized (75 kGy GUR 1020-E) UHMWPE inserts and tested to visible fracture or a 3-million-cycle runout. Post-test fractographic analysis identified crack initiation sites. Conventional aged UHMWPE demonstrated fatigue failure under varus-valgus rotation (OI = 2.59 ± 1.11) but no failure without rotation. Vitamin E-stabilized UHMWPE showed no fatigue failure after 3 million cycles, even under varus-valgus rotation (OI = 0.23 ± 0.02). Fractography revealed fractures originating at the trough and propagating with cyclic loading. Oxidation significantly reduces polyethylene fatigue life, and varus-valgus rotation exacerbates this effect in mobile bearing TAR implants. Antioxidant-stabilized UHMWPE showed promising resistance to fatigue and oxidation. These findings support the role of antioxidant stabilization in improving TAR performance, and the protocols developed here provide a framework for assessing the safety of alternative materials.

Journal of Orthopaedic Research®Vol. 44(1)
Exponent (United States) (US), Palomar Health (US), Clovis Oncology (United States) (US), Drexel University (US)
Openalex Percentile: Top 81%
Total Knee Arthroplasty Outcomes
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.