Stress–Strain State Investigation and Ultimate Load on Femoral Implants Based on S-Type Ti6Al4V Titanium Alloy

Hip replacement is a widespread surgical procedure that eliminates pain and restores motor functions of the pathologically altered hip joint. The issue lies in the lack of pre-operative strength calculations for implant shapes. So, they tend to break after surgery or damage the bone due to the complex stress-strain state. In the present paper, we studied the stress-strain state and ultimate load of S-type canine femoral implants based on titanium alloy Ti6Al4V using finite element analysis for static and cyclic loads. X-ray computed micro tomography data were used to construct the models. Re-engineering and restoration of the 3D geometry of the product were conducted. Strength analysis was performed in the finite element analysis software package Ansys Mechanical was used for various types of implant support. Locations with stress concentrators were identified, and ultimate loads on the implant were obtained. The influence of the rigidity of the support on the prosthesis stem was also studied. For the case of rigid support, the stress-strain state of the prosthesis was studied and the ultimate load was found to be 30.1 kg.

Authors

Institutions

Publication Details

Journal
Journal of Functional Biomaterials
Published
2025-05-19
DOI
https://doi.org/10.3390/jfb16050187
Citations
3
Primary Topic
Orthopaedic implants and arthroplasty
Type
article
Field-Weighted Citation Impact
6.02
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Stress–Strain State Investigation and Ultimate Load on Femoral Implants Based on S-Type Ti6Al4V Titanium Alloy

С. М. Айзикович, Alexey Nikolaevich Beskopylny, Evgeniy V. Sadyrin, Ivan Panfilov et al.
3 citations
Journal of Functional Biomaterials
Orthopaedic implants and arthroplasty
6.02
article

Stress–Strain State Investigation and Ultimate Load on Femoral Implants Based on S-Type Ti6Al4V Titanium Alloy

С. М. Айзикович, Alexey Nikolaevich Beskopylny, Evgeniy V. Sadyrin, Ivan Panfilov, Besarion Meskhi, Il'ya Vilkovyskiy
article en
3 citations

Abstract

Hip replacement is a widespread surgical procedure that eliminates pain and restores motor functions of the pathologically altered hip joint. The issue lies in the lack of pre-operative strength calculations for implant shapes. So, they tend to break after surgery or damage the bone due to the complex stress-strain state. In the present paper, we studied the stress-strain state and ultimate load of S-type canine femoral implants based on titanium alloy Ti6Al4V using finite element analysis for static and cyclic loads. X-ray computed micro tomography data were used to construct the models. Re-engineering and restoration of the 3D geometry of the product were conducted. Strength analysis was performed in the finite element analysis software package Ansys Mechanical was used for various types of implant support. Locations with stress concentrators were identified, and ultimate loads on the implant were obtained. The influence of the rigidity of the support on the prosthesis stem was also studied. For the case of rigid support, the stress-strain state of the prosthesis was studied and the ultimate load was found to be 30.1 kg.

Journal of Functional BiomaterialsVol. 16(5)
Peoples' Friendship University of Russia (RU), Institute of Service and Entrepreneurship of DGTU (RU), Don State Technical University (RU)
Openalex Percentile: Top 4%
Orthopaedic implants and arthroplasty
6.02
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Stress–Strain State Investigation and Ultimate Load on Femoral Implants Based on S-Type Ti6Al4V Titanium Alloy — С. М. Айзикович, Alexey Nikolaevich Beskopylny, et al. · Journal of Functional Biomaterials (2025) | TGRS Research Map | TGRS