Analytical Verification and Method Comparison of the Maglumi X8 for Thyroid Function Tests
The precision and trueness verification results for TSH and FT4 demonstrated acceptable performance under the CLSI EP15-A3. TSH results between the two analyzers were consistent and can be transferred; FT4 results from Maglumi X8 require careful interpretation and may need harmonization and standardization due to the observed bias.