A Risk-Based Approach for Pre-Use/Post-Sterilization Integrity Test Simulation During Bacterial Retention Testing as Part of the Process-Specific Filter Validation of Sterilizing Grade Filters

Pre-use post-sterilization integrity testing is implemented in sterile filtration applications by drug manufacturers using a risk-based approach with consideration of the stipulation in Eudralex Volume 4, Annex 1, Manufacture of Sterile Medicinal Products, which states that "The integrity of the sterilizing filter should be verified before use". Within the bacterial retention test design, performed as part of the process-specific sterile filter validation, it is important to simulate the filtration process under evaluation as closely as possible. Worst-case conditions experienced by the process filter during routine use that may impact the process filters' ability to produce a sterile effluent should be accounted for in the study. As performing pre-use post-sterilization integrity tests introduces additional mechanical stress on the process filter and a new potential route for the introduction of bioburden into the process fluid flow path related to the filter wetting procedure, a risk-based bacterial retention test design incorporating a pre-use post-sterilization integrity test simulation phase should be considered after a thorough evaluation of the process-specific conditions. The risk assessment should include evaluation of process pre-use post-sterilization integrity test conditions, including but not limited to the integrity test method and specifications, the wetting fluid type, the maximum allowable number of pre-use post-sterilization integrity test repetitions, as well as the permitted bioburden level of the pre-use post-sterilization integrity test wetting fluid. The outcome of the assessment provides a basis for the process-specific bacterial retention test design. In the following, we present our perspective on the topic as well as detailed insight into various aspects of pre-use post-sterilization integrity test simulation test design for consideration. Finally, we present all three PUPSIT-BCT approaches described (non-bactericidal, partially bactericidal, and bactericidal). Executed studies yielded acceptable results. Target parameters agreed upon were met.

Authors

Institutions

Publication Details

Journal
PDA Journal of Pharmaceutical Science and Technology
Published
2025-01-01
DOI
https://doi.org/10.5731/pdajpst.2024.012990
Primary Topic
Medical Device Sterilization and Disinfection
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A Risk-Based Approach for Pre-Use/Post-Sterilization Integrity Test Simulation During Bacterial Retention Testing as Part of the Process-Specific Filter Validation of Sterilizing Grade Filters

Ashira Bindels, Magnus Stering, T. Vinh, Mariam Salamatian et al.
PDA Journal of Pharmaceutical Science and Technology
Medical Device Sterilization and Disinfection
article

A Risk-Based Approach for Pre-Use/Post-Sterilization Integrity Test Simulation During Bacterial Retention Testing as Part of the Process-Specific Filter Validation of Sterilizing Grade Filters

Ashira Bindels, Magnus Stering, T. Vinh, Mariam Salamatian, Yvonne Gross
article en

Abstract

Pre-use post-sterilization integrity testing is implemented in sterile filtration applications by drug manufacturers using a risk-based approach with consideration of the stipulation in Eudralex Volume 4, Annex 1, Manufacture of Sterile Medicinal Products, which states that "The integrity of the sterilizing filter should be verified before use". Within the bacterial retention test design, performed as part of the process-specific sterile filter validation, it is important to simulate the filtration process under evaluation as closely as possible. Worst-case conditions experienced by the process filter during routine use that may impact the process filters' ability to produce a sterile effluent should be accounted for in the study. As performing pre-use post-sterilization integrity tests introduces additional mechanical stress on the process filter and a new potential route for the introduction of bioburden into the process fluid flow path related to the filter wetting procedure, a risk-based bacterial retention test design incorporating a pre-use post-sterilization integrity test simulation phase should be considered after a thorough evaluation of the process-specific conditions. The risk assessment should include evaluation of process pre-use post-sterilization integrity test conditions, including but not limited to the integrity test method and specifications, the wetting fluid type, the maximum allowable number of pre-use post-sterilization integrity test repetitions, as well as the permitted bioburden level of the pre-use post-sterilization integrity test wetting fluid. The outcome of the assessment provides a basis for the process-specific bacterial retention test design. In the following, we present our perspective on the topic as well as detailed insight into various aspects of pre-use post-sterilization integrity test simulation test design for consideration. Finally, we present all three PUPSIT-BCT approaches described (non-bactericidal, partially bactericidal, and bactericidal). Executed studies yielded acceptable results. Target parameters agreed upon were met.

PDA Journal of Pharmaceutical Science and TechnologyVol. 79(4)
GlaxoSmithKline (Belgium) (BE)
Openalex Percentile: Top 90%
Medical Device Sterilization and Disinfection
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.